Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage

BACKGROUND: Despite improvements in closure techniques by using a vital nasoseptal flap, the use of sealing materials, and improved neurosurgical techniques, cerebrospinal fluid (CSF) leak after transsphenoidal surgery still is a clinically relevant problem. Liqoseal® (Polyganics bv, Groningen, The...

Descripción completa

Detalles Bibliográficos
Autores principales: Slot, Emma M. H., Colmer, Nadia, Serra, Carlo, Holzmann, David, Regli, Luca, van Doormaal, Tristan P. C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Vienna 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10227155/
https://www.ncbi.nlm.nih.gov/pubmed/36624231
http://dx.doi.org/10.1007/s00701-022-05477-3
_version_ 1785050708849983488
author Slot, Emma M. H.
Colmer, Nadia
Serra, Carlo
Holzmann, David
Regli, Luca
van Doormaal, Tristan P. C.
author_facet Slot, Emma M. H.
Colmer, Nadia
Serra, Carlo
Holzmann, David
Regli, Luca
van Doormaal, Tristan P. C.
author_sort Slot, Emma M. H.
collection PubMed
description BACKGROUND: Despite improvements in closure techniques by using a vital nasoseptal flap, the use of sealing materials, and improved neurosurgical techniques, cerebrospinal fluid (CSF) leak after transsphenoidal surgery still is a clinically relevant problem. Liqoseal® (Polyganics bv, Groningen, The Netherlands) is a CE-approved bioresorbable sealant patch for use as an adjunct to standard methods of cranial dural closure to prevent CSF leakage. This study aims to evaluate the application of Liqoseal in transsphenoidal surgery ex vivo and in vivo. METHODS: 1. We created an ex vivo setup simulating the sphenoidal anatomy, using a fluid pump and porcine dura positioned on a conus with the anatomical dimensions of the sella to evaluate whether the burst pressure of Liqoseal applied to a bulging surface was above physiological intracranial pressure. Burst pressure was measured with a probe connected to dedicated computer software. Because of the challenging transsphenoidal environment, we tested in 4 groups with varying compression weight and time for the application of Liqoseal. 2. We subsequently describe the application of Liqoseal® in 3 patients during transsphenoidal procedures with intraoperative CSF leakage to prevent postoperative CSF leakage. RESULTS: 1. Ex vivo: The overall mean burst pressure in the transsphenoidal setup was 231 (± 103) mmHg. There was no significant difference in mean burst pressure between groups based on application weight and time (p = 0.227). 2. In Vivo: None of the patients had a postoperative CSF leak. No nose passage problems were observed. One patient had a postoperative meningitis and ventriculitis, most likely related to preoperative extensive CSF leakage. Postoperative imaging did not show any local infection, swelling, or other device-related adverse effects. CONCLUSIONS: We assess the use of Liqoseal® to seal a dural defect during an endoscopic transsphenoidal procedure as to be likely safe and potentially effective. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00701-022-05477-3.
format Online
Article
Text
id pubmed-10227155
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer Vienna
record_format MEDLINE/PubMed
spelling pubmed-102271552023-05-31 Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage Slot, Emma M. H. Colmer, Nadia Serra, Carlo Holzmann, David Regli, Luca van Doormaal, Tristan P. C. Acta Neurochir (Wien) Original Article - Neurosurgical Technique Evaluation BACKGROUND: Despite improvements in closure techniques by using a vital nasoseptal flap, the use of sealing materials, and improved neurosurgical techniques, cerebrospinal fluid (CSF) leak after transsphenoidal surgery still is a clinically relevant problem. Liqoseal® (Polyganics bv, Groningen, The Netherlands) is a CE-approved bioresorbable sealant patch for use as an adjunct to standard methods of cranial dural closure to prevent CSF leakage. This study aims to evaluate the application of Liqoseal in transsphenoidal surgery ex vivo and in vivo. METHODS: 1. We created an ex vivo setup simulating the sphenoidal anatomy, using a fluid pump and porcine dura positioned on a conus with the anatomical dimensions of the sella to evaluate whether the burst pressure of Liqoseal applied to a bulging surface was above physiological intracranial pressure. Burst pressure was measured with a probe connected to dedicated computer software. Because of the challenging transsphenoidal environment, we tested in 4 groups with varying compression weight and time for the application of Liqoseal. 2. We subsequently describe the application of Liqoseal® in 3 patients during transsphenoidal procedures with intraoperative CSF leakage to prevent postoperative CSF leakage. RESULTS: 1. Ex vivo: The overall mean burst pressure in the transsphenoidal setup was 231 (± 103) mmHg. There was no significant difference in mean burst pressure between groups based on application weight and time (p = 0.227). 2. In Vivo: None of the patients had a postoperative CSF leak. No nose passage problems were observed. One patient had a postoperative meningitis and ventriculitis, most likely related to preoperative extensive CSF leakage. Postoperative imaging did not show any local infection, swelling, or other device-related adverse effects. CONCLUSIONS: We assess the use of Liqoseal® to seal a dural defect during an endoscopic transsphenoidal procedure as to be likely safe and potentially effective. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00701-022-05477-3. Springer Vienna 2023-01-09 2023 /pmc/articles/PMC10227155/ /pubmed/36624231 http://dx.doi.org/10.1007/s00701-022-05477-3 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article - Neurosurgical Technique Evaluation
Slot, Emma M. H.
Colmer, Nadia
Serra, Carlo
Holzmann, David
Regli, Luca
van Doormaal, Tristan P. C.
Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title_full Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title_fullStr Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title_full_unstemmed Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title_short Ex vivo and in vivo evaluation of transsphenoidal Liqoseal application to prevent cerebrospinal fluid leakage
title_sort ex vivo and in vivo evaluation of transsphenoidal liqoseal application to prevent cerebrospinal fluid leakage
topic Original Article - Neurosurgical Technique Evaluation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10227155/
https://www.ncbi.nlm.nih.gov/pubmed/36624231
http://dx.doi.org/10.1007/s00701-022-05477-3
work_keys_str_mv AT slotemmamh exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage
AT colmernadia exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage
AT serracarlo exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage
AT holzmanndavid exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage
AT regliluca exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage
AT vandoormaaltristanpc exvivoandinvivoevaluationoftranssphenoidalliqosealapplicationtopreventcerebrospinalfluidleakage