Cargando…

Finite element analysis of the mechanical strength of a new hip Spacer

BACKGROUND AND OBJECTIVE: At present, the influence of the internal metallic endoskeleton of Spacer on the biomechanical strength of Spacer remains unclear. The aim of this study was to analyze the mechanical stability of a novel Spacer applying a annular skeleton that mimics the structure of trabec...

Descripción completa

Detalles Bibliográficos
Autores principales: Ge, Hao, Yan, Hongsong, Liu, Xianwang, Huang, Yiwei, Zeng, Jianchun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10228057/
https://www.ncbi.nlm.nih.gov/pubmed/37254116
http://dx.doi.org/10.1186/s12891-023-06562-z
_version_ 1785050892705202176
author Ge, Hao
Yan, Hongsong
Liu, Xianwang
Huang, Yiwei
Zeng, Jianchun
author_facet Ge, Hao
Yan, Hongsong
Liu, Xianwang
Huang, Yiwei
Zeng, Jianchun
author_sort Ge, Hao
collection PubMed
description BACKGROUND AND OBJECTIVE: At present, the influence of the internal metallic endoskeleton of Spacer on the biomechanical strength of Spacer remains unclear. The aim of this study was to analyze the mechanical stability of a novel Spacer applying a annular skeleton that mimics the structure of trabecular bone using finite element methods. METHEDS: The femur models of three healthy individuals and skeletonless Spacer, K-Spacer, and AD-Spacer were assembled to create 15 3D models. Finite element analysis was performed in an Ansys Bench2022R1. Biomechanical parameters such as stress and strain of the Spacer, internal skeleton and femur were evaluated under three loads, which were applied with the maximum force borne by the hip joint (2100 N), standing on one leg (700 N), and standing on two legs (350 N). The mechanical properties of the new hip Spacer were evaluated. RESULT: The stresses on the medial and lateral surfaces of the AD-Spacer and K-Spacer were smaller than the stresses in the state without skeletal support. The maximum stresses on the medial and lateral surfaces of the AD-Spacer were smaller than those of the inserted K-Spacer, and the difference gradually increased with the increase of force intensity. When the skeleton diameter was increased from 3 to 4 mm, the stresses in the medial and lateral sides of the AD-Spacer and K-Spacer necks decreased. The stress of both skeletons was concentrated at the neck, but the stress of the annular skeleton was evenly distributed on the medial and lateral sides of the skeleton. The mean stress in the proximal femur was higher in femurs with K-Spacer than in femurs with AD-Spacer. CONCLUSIONS: AD-Spacer has lower stress and higher load-bearing capacity than K-Spacer, and the advantages of AD-Spacer are more obvious under the maximum load state of hip joint.
format Online
Article
Text
id pubmed-10228057
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-102280572023-05-31 Finite element analysis of the mechanical strength of a new hip Spacer Ge, Hao Yan, Hongsong Liu, Xianwang Huang, Yiwei Zeng, Jianchun BMC Musculoskelet Disord Research BACKGROUND AND OBJECTIVE: At present, the influence of the internal metallic endoskeleton of Spacer on the biomechanical strength of Spacer remains unclear. The aim of this study was to analyze the mechanical stability of a novel Spacer applying a annular skeleton that mimics the structure of trabecular bone using finite element methods. METHEDS: The femur models of three healthy individuals and skeletonless Spacer, K-Spacer, and AD-Spacer were assembled to create 15 3D models. Finite element analysis was performed in an Ansys Bench2022R1. Biomechanical parameters such as stress and strain of the Spacer, internal skeleton and femur were evaluated under three loads, which were applied with the maximum force borne by the hip joint (2100 N), standing on one leg (700 N), and standing on two legs (350 N). The mechanical properties of the new hip Spacer were evaluated. RESULT: The stresses on the medial and lateral surfaces of the AD-Spacer and K-Spacer were smaller than the stresses in the state without skeletal support. The maximum stresses on the medial and lateral surfaces of the AD-Spacer were smaller than those of the inserted K-Spacer, and the difference gradually increased with the increase of force intensity. When the skeleton diameter was increased from 3 to 4 mm, the stresses in the medial and lateral sides of the AD-Spacer and K-Spacer necks decreased. The stress of both skeletons was concentrated at the neck, but the stress of the annular skeleton was evenly distributed on the medial and lateral sides of the skeleton. The mean stress in the proximal femur was higher in femurs with K-Spacer than in femurs with AD-Spacer. CONCLUSIONS: AD-Spacer has lower stress and higher load-bearing capacity than K-Spacer, and the advantages of AD-Spacer are more obvious under the maximum load state of hip joint. BioMed Central 2023-05-30 /pmc/articles/PMC10228057/ /pubmed/37254116 http://dx.doi.org/10.1186/s12891-023-06562-z Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Ge, Hao
Yan, Hongsong
Liu, Xianwang
Huang, Yiwei
Zeng, Jianchun
Finite element analysis of the mechanical strength of a new hip Spacer
title Finite element analysis of the mechanical strength of a new hip Spacer
title_full Finite element analysis of the mechanical strength of a new hip Spacer
title_fullStr Finite element analysis of the mechanical strength of a new hip Spacer
title_full_unstemmed Finite element analysis of the mechanical strength of a new hip Spacer
title_short Finite element analysis of the mechanical strength of a new hip Spacer
title_sort finite element analysis of the mechanical strength of a new hip spacer
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10228057/
https://www.ncbi.nlm.nih.gov/pubmed/37254116
http://dx.doi.org/10.1186/s12891-023-06562-z
work_keys_str_mv AT gehao finiteelementanalysisofthemechanicalstrengthofanewhipspacer
AT yanhongsong finiteelementanalysisofthemechanicalstrengthofanewhipspacer
AT liuxianwang finiteelementanalysisofthemechanicalstrengthofanewhipspacer
AT huangyiwei finiteelementanalysisofthemechanicalstrengthofanewhipspacer
AT zengjianchun finiteelementanalysisofthemechanicalstrengthofanewhipspacer