Assessing Enhanced Acoustic Absorption From Sonosensitive Perfluorocarbon Emulsion With Magnetic Resonance–Guided High-Intensity Focused Ultrasound and a Percolated Tissue-Mimicking Flow Phantom
OBJECTIVE: Sonosensitive high-boiling point perfluorocarbon F(8)TAC(18)-PFOB emulsions previously exhibited thermal enhancement during focused ultrasound heating in ex vivo pig livers, kidneys and a laminar flow phantom. The main objectives of this study were to evaluate heating under turbulent cond...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Pergamon Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10229106/ https://www.ncbi.nlm.nih.gov/pubmed/37117139 http://dx.doi.org/10.1016/j.ultrasmedbio.2023.01.022 |
Sumario: | OBJECTIVE: Sonosensitive high-boiling point perfluorocarbon F(8)TAC(18)-PFOB emulsions previously exhibited thermal enhancement during focused ultrasound heating in ex vivo pig livers, kidneys and a laminar flow phantom. The main objectives of this study were to evaluate heating under turbulent conditions, observe perfusion effects, quantify heating in terms of acoustic absorption and model the experimental data. METHODS: In this study, similar perfluorocarbon emulsions were circulated at incremental concentrations of 0.07, 0.13, 0.19 and 0.25% v:v through a percolated turbulent flow phantom, more representative of the biological tissue than a laminar flow phantom. The concentrations represent the droplet content in only the perfused fluid, rather than the droplet concentration throughout the entire cross-section. The temperature was measured with magnetic resonance thermometry, during focused ultrasound sonications of 67 W, 95% duty cycle and 33 s duration. These were used in Bioheat equation simulations to investigate in silico the thermal phenomena. The temperature change was compared with the control condition by circulating de-gassed and de-ionized water through the flow phantom without droplets. RESULTS: With these 1.24 µm diameter droplets at 0.25% v:v, the acoustic absorption coefficient increased from 0.93 ± 0.05 at 0.0% v:v to 1.82 ± 0.22 m(–1) at 0.25% v:v using a 0.1 mL s(–1) flow rate. Without perfusion at 0.25% v:v, an increase was observed from 1.23 ± 0.07 m(–1) at 0.0% v:v to 1.65 ± 0.17 m(–1). CONCLUSION: The results further support previously reported thermal enhancement with F(8)TAC(18)-PFOB emulsion, quantified the increased absorption at small concentration intervals, illustrated that the effects can be observed in a variety of visceral tissue models and provided a method to simulate untested scenarios. |
---|