Analytical verification of the Dymind D7-CRP automated analyser
INTRODUCTION: The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers. MATERIALS AND METHODS: Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in co...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10231769/ https://www.ncbi.nlm.nih.gov/pubmed/37324112 http://dx.doi.org/10.11613/BM.2023.020703 |
_version_ | 1785051806360928256 |
---|---|
author | Čolić, Merima Magdić, Bojana Kolundžić, Monika Leniček Krleža, Jasna |
author_facet | Čolić, Merima Magdić, Bojana Kolundžić, Monika Leniček Krleža, Jasna |
author_sort | Čolić, Merima |
collection | PubMed |
description | INTRODUCTION: The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers. MATERIALS AND METHODS: Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database. Method comparison between the Dymind D7-CRP and the Sysmex XN1000 for haematological parameters and the Dymind D7-CRP and the Beckman Coulter AU680 for CRP values was performed on 40 patient samples. RESULTS: Analytical verification criteria were adequately met with the exception of monocyte count for repeatability and within-laboratory precision (13.4% and 11.5%, respectively, acceptance criteria 10.1%) and measurement uncertainty (23.0, acceptance criteria 20.0%) at low level, eosinophil count for BIAS at the low level (37.7%, acceptance criteria 25.2%), basophil count (BAS) for BIAS at the high level (14.2%, acceptance criteria 10.9%), and mean platelet volume (MPV) for repeatability (4.2% and 6.8%), between run precision (2.2% and 4.7%), within-laboratory precision (4.0% and 7.3%) (acceptance criteria 1.7%), and measurement uncertainty (8.0 and 14.6%, acceptance criteria 3.4%) at both the low and high concentrations. Method comparison showed no clinically significant constant or proportional differences for all parameters except BAS and MPV. CONCLUSION: The analytical verification of the Dymind D7-CRP showed adequate analytical characteristics. The Dymind D7-CRP can be used interchangeably with the Sysmex XN-1000 for all parameters tested, except BAS and MPV, and with the Beckman Coulter AU-680 for the determination of CRP. |
format | Online Article Text |
id | pubmed-10231769 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-102317692023-06-15 Analytical verification of the Dymind D7-CRP automated analyser Čolić, Merima Magdić, Bojana Kolundžić, Monika Leniček Krleža, Jasna Biochem Med (Zagreb) Original Articles INTRODUCTION: The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers. MATERIALS AND METHODS: Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database. Method comparison between the Dymind D7-CRP and the Sysmex XN1000 for haematological parameters and the Dymind D7-CRP and the Beckman Coulter AU680 for CRP values was performed on 40 patient samples. RESULTS: Analytical verification criteria were adequately met with the exception of monocyte count for repeatability and within-laboratory precision (13.4% and 11.5%, respectively, acceptance criteria 10.1%) and measurement uncertainty (23.0, acceptance criteria 20.0%) at low level, eosinophil count for BIAS at the low level (37.7%, acceptance criteria 25.2%), basophil count (BAS) for BIAS at the high level (14.2%, acceptance criteria 10.9%), and mean platelet volume (MPV) for repeatability (4.2% and 6.8%), between run precision (2.2% and 4.7%), within-laboratory precision (4.0% and 7.3%) (acceptance criteria 1.7%), and measurement uncertainty (8.0 and 14.6%, acceptance criteria 3.4%) at both the low and high concentrations. Method comparison showed no clinically significant constant or proportional differences for all parameters except BAS and MPV. CONCLUSION: The analytical verification of the Dymind D7-CRP showed adequate analytical characteristics. The Dymind D7-CRP can be used interchangeably with the Sysmex XN-1000 for all parameters tested, except BAS and MPV, and with the Beckman Coulter AU-680 for the determination of CRP. Croatian Society of Medical Biochemistry and Laboratory Medicine 2023-06-15 2023-06-15 /pmc/articles/PMC10231769/ /pubmed/37324112 http://dx.doi.org/10.11613/BM.2023.020703 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles Čolić, Merima Magdić, Bojana Kolundžić, Monika Leniček Krleža, Jasna Analytical verification of the Dymind D7-CRP automated analyser |
title | Analytical verification of the Dymind D7-CRP automated analyser |
title_full | Analytical verification of the Dymind D7-CRP automated analyser |
title_fullStr | Analytical verification of the Dymind D7-CRP automated analyser |
title_full_unstemmed | Analytical verification of the Dymind D7-CRP automated analyser |
title_short | Analytical verification of the Dymind D7-CRP automated analyser |
title_sort | analytical verification of the dymind d7-crp automated analyser |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10231769/ https://www.ncbi.nlm.nih.gov/pubmed/37324112 http://dx.doi.org/10.11613/BM.2023.020703 |
work_keys_str_mv | AT colicmerima analyticalverificationofthedymindd7crpautomatedanalyser AT magdicbojana analyticalverificationofthedymindd7crpautomatedanalyser AT kolundzicmonika analyticalverificationofthedymindd7crpautomatedanalyser AT lenicekkrlezajasna analyticalverificationofthedymindd7crpautomatedanalyser |