In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex

BACKGROUND: Instrumentation technology for transepidermal water loss measurements has not been substantially modified since its introduction by Nilsson in 1977. Recent progress in sensor development allowed a new sensor arrangement using a matrix of 30 sensors. Raw measurement values are processed w...

Descripción completa

Detalles Bibliográficos
Autores principales: Fluhr, Joachim W., Wiora, Georg, Nikolaeva, Dessyslava G., Miséry, Laurent, Darlenski, Razvigor
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10234168/
https://www.ncbi.nlm.nih.gov/pubmed/37113085
http://dx.doi.org/10.1111/srt.13307
_version_ 1785052425326952448
author Fluhr, Joachim W.
Wiora, Georg
Nikolaeva, Dessyslava G.
Miséry, Laurent
Darlenski, Razvigor
author_facet Fluhr, Joachim W.
Wiora, Georg
Nikolaeva, Dessyslava G.
Miséry, Laurent
Darlenski, Razvigor
author_sort Fluhr, Joachim W.
collection PubMed
description BACKGROUND: Instrumentation technology for transepidermal water loss measurements has not been substantially modified since its introduction by Nilsson in 1977. Recent progress in sensor development allowed a new sensor arrangement using a matrix of 30 sensors. Raw measurement values are processed with spatial statistical analysis. We aimed to compare the new, multi‐sensor probe (Tewameter TM Hex) with the established Tewameter TM 300 probe and to gain reference data for the new parameters of transepidermal energy loss and water vapor concentration on skin. MATERIAL AND METHODS: Baseline measurements and repeated measurements on the volar forearm and assessment on eight different anatomical locations were performed on 24 healthy volunteers (both gender) with the TM Hex and the TM 300. RESULTS: A significant correlation (p < 0.001; R‐coefficient = 0.9) between TM Hex and the TM 300 with a low coefficient of variance (CV) 11% for TM Hex and 19% for TM 300, could be assessed. The CV ranged between 7% (right inner upper arm) and 14% (palms). Average transepidermal heat loss ranged from 12 W/m(2) on the lower leg to 38.8 W/m(2) on the palm. CONCLUSION: The correlation between TM Hex and TM 300 along with the robustness of the measurements with the TM Hex shows that the new probe for assessment of epidermal barrier function is comparable to the TM 300. In most conditions, TM Hex provides more accurate measurements than TM 300. New parameters open the field to studying skin's water and energy balance.
format Online
Article
Text
id pubmed-10234168
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-102341682023-08-11 In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex Fluhr, Joachim W. Wiora, Georg Nikolaeva, Dessyslava G. Miséry, Laurent Darlenski, Razvigor Skin Res Technol Original Articles BACKGROUND: Instrumentation technology for transepidermal water loss measurements has not been substantially modified since its introduction by Nilsson in 1977. Recent progress in sensor development allowed a new sensor arrangement using a matrix of 30 sensors. Raw measurement values are processed with spatial statistical analysis. We aimed to compare the new, multi‐sensor probe (Tewameter TM Hex) with the established Tewameter TM 300 probe and to gain reference data for the new parameters of transepidermal energy loss and water vapor concentration on skin. MATERIAL AND METHODS: Baseline measurements and repeated measurements on the volar forearm and assessment on eight different anatomical locations were performed on 24 healthy volunteers (both gender) with the TM Hex and the TM 300. RESULTS: A significant correlation (p < 0.001; R‐coefficient = 0.9) between TM Hex and the TM 300 with a low coefficient of variance (CV) 11% for TM Hex and 19% for TM 300, could be assessed. The CV ranged between 7% (right inner upper arm) and 14% (palms). Average transepidermal heat loss ranged from 12 W/m(2) on the lower leg to 38.8 W/m(2) on the palm. CONCLUSION: The correlation between TM Hex and TM 300 along with the robustness of the measurements with the TM Hex shows that the new probe for assessment of epidermal barrier function is comparable to the TM 300. In most conditions, TM Hex provides more accurate measurements than TM 300. New parameters open the field to studying skin's water and energy balance. John Wiley and Sons Inc. 2023-03-31 /pmc/articles/PMC10234168/ /pubmed/37113085 http://dx.doi.org/10.1111/srt.13307 Text en © 2023 The Authors. Skin Research and Technology published by John Wiley & Sons Ltd. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Original Articles
Fluhr, Joachim W.
Wiora, Georg
Nikolaeva, Dessyslava G.
Miséry, Laurent
Darlenski, Razvigor
In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title_full In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title_fullStr In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title_full_unstemmed In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title_short In vivo transepidermal water loss: Validation of a new multi‐sensor open chamber water evaporation system Tewameter TM Hex
title_sort in vivo transepidermal water loss: validation of a new multi‐sensor open chamber water evaporation system tewameter tm hex
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10234168/
https://www.ncbi.nlm.nih.gov/pubmed/37113085
http://dx.doi.org/10.1111/srt.13307
work_keys_str_mv AT fluhrjoachimw invivotransepidermalwaterlossvalidationofanewmultisensoropenchamberwaterevaporationsystemtewametertmhex
AT wiorageorg invivotransepidermalwaterlossvalidationofanewmultisensoropenchamberwaterevaporationsystemtewametertmhex
AT nikolaevadessyslavag invivotransepidermalwaterlossvalidationofanewmultisensoropenchamberwaterevaporationsystemtewametertmhex
AT miserylaurent invivotransepidermalwaterlossvalidationofanewmultisensoropenchamberwaterevaporationsystemtewametertmhex
AT darlenskirazvigor invivotransepidermalwaterlossvalidationofanewmultisensoropenchamberwaterevaporationsystemtewametertmhex