Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas

Background This study aims to evaluate dosimetric changes that happened during the first month after GammaTile surgically targeted radiation therapy (STaRT) for gliomas due to Cesium-131 (Cs-131) seed displacement caused by cavity shrinkage in brain brachytherapy. Methodology In this study, 10 gliom...

Descripción completa

Detalles Bibliográficos
Autores principales: Ng, Sook Kien, Yue, Yong, Shiue, Kevin, Shah, Mitesh V, Le, Yi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10234842/
https://www.ncbi.nlm.nih.gov/pubmed/37273347
http://dx.doi.org/10.7759/cureus.38463
_version_ 1785052588111036416
author Ng, Sook Kien
Yue, Yong
Shiue, Kevin
Shah, Mitesh V
Le, Yi
author_facet Ng, Sook Kien
Yue, Yong
Shiue, Kevin
Shah, Mitesh V
Le, Yi
author_sort Ng, Sook Kien
collection PubMed
description Background This study aims to evaluate dosimetric changes that happened during the first month after GammaTile surgically targeted radiation therapy (STaRT) for gliomas due to Cesium-131 (Cs-131) seed displacement caused by cavity shrinkage in brain brachytherapy. Methodology In this study, 10 glioma patients had 4-11 GammaTiles placed along the resection bed after maximal safe resection during craniotomy. Each GammaTile is composed of four Cs-131 seeds embedded in a biodegradable collagen sponge to minimize seed movement and maintain seed-to-cavity surface distance. The Cs-131 seed positions were identified using VariSeed on day one. On day 30, post-implant computed tomography (CT) images and dosimetry parameters were calculated. An iterative closest point (ICP) algorithm was used to compute rigid transformation between the day one and day 30 seed clouds. The seed displacement was calculated after registration. The volume receiving 100% of the prescription dose (V100), the dose received by 90% of the planning target volume (D90_PTV), the planning target volume receiving 100% of the prescription dose (V100_PTV), and the dose to organs at risk (OARs) were calculated for both CT images to determine the dosimetric changes from any seed displacement. Results The mean seed displacement of 1.8 ± 1.0 mm for all patients was observed between day one and day 30. The maximum seed displacement for each patient ranged from 2.3 mm to 7.3 mm. The mean V100 difference between day one and day 30 was 2.5 cc (range = 0.5-6.5 cc). The mean D90_PTVs were 95.5% (range = 69.0%-131.0%) and 98.1% (range = 19.9%-149.0%) on day one and day 30, respectively. The mean V100_PTVs were 88.4% (range = 81.3%-99.1%) and 87.9% (range = 47.0%-99.7%) on day one and day 30, respectively. On day one, the brainstem dose was 63.5 Gy for one case and 28.1 Gy for another case; while on day 30, the brainstem dose was 55.8 Gy and 20.6 Gy for the same patients, contributing to 7.7 Gy (12.8%) and 7.5 Gy (12.5%) dose reductions to brainstem for these patients, respectively. Only two patients received a dose to the optic nerves (34.1 Gy and 5.2 Gy). There were small changes (1.8 Gy and 0.5 Gy, respectively) in the dose to optic nerves when comparing the dose calculated on day one and the dose calculated on day 30 CT images. The same two patients received 30.4 Gy and 6.8 Gy to the chiasm, respectively. Small changes in the dose to the chiasm (≤1.1 Gy) were noted between day one and day 30. Conclusions A maximum seed displacement of up to 7.3 mm and a mean seed displacement of 1.8 mm caused by cavity shrinkage were observed during the first month after GammaTile STaRT for gliomas. There were noticeable changes in dosimetry parameters. Changes in the doses to OARs, particularly the brainstem, were large (up to 12.8% of the prescription dose). These changes in dosimetry should be considered when evaluating treatment outcomes and planning future GammaTile treatments.
format Online
Article
Text
id pubmed-10234842
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-102348422023-06-03 Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas Ng, Sook Kien Yue, Yong Shiue, Kevin Shah, Mitesh V Le, Yi Cureus Radiation Oncology Background This study aims to evaluate dosimetric changes that happened during the first month after GammaTile surgically targeted radiation therapy (STaRT) for gliomas due to Cesium-131 (Cs-131) seed displacement caused by cavity shrinkage in brain brachytherapy. Methodology In this study, 10 glioma patients had 4-11 GammaTiles placed along the resection bed after maximal safe resection during craniotomy. Each GammaTile is composed of four Cs-131 seeds embedded in a biodegradable collagen sponge to minimize seed movement and maintain seed-to-cavity surface distance. The Cs-131 seed positions were identified using VariSeed on day one. On day 30, post-implant computed tomography (CT) images and dosimetry parameters were calculated. An iterative closest point (ICP) algorithm was used to compute rigid transformation between the day one and day 30 seed clouds. The seed displacement was calculated after registration. The volume receiving 100% of the prescription dose (V100), the dose received by 90% of the planning target volume (D90_PTV), the planning target volume receiving 100% of the prescription dose (V100_PTV), and the dose to organs at risk (OARs) were calculated for both CT images to determine the dosimetric changes from any seed displacement. Results The mean seed displacement of 1.8 ± 1.0 mm for all patients was observed between day one and day 30. The maximum seed displacement for each patient ranged from 2.3 mm to 7.3 mm. The mean V100 difference between day one and day 30 was 2.5 cc (range = 0.5-6.5 cc). The mean D90_PTVs were 95.5% (range = 69.0%-131.0%) and 98.1% (range = 19.9%-149.0%) on day one and day 30, respectively. The mean V100_PTVs were 88.4% (range = 81.3%-99.1%) and 87.9% (range = 47.0%-99.7%) on day one and day 30, respectively. On day one, the brainstem dose was 63.5 Gy for one case and 28.1 Gy for another case; while on day 30, the brainstem dose was 55.8 Gy and 20.6 Gy for the same patients, contributing to 7.7 Gy (12.8%) and 7.5 Gy (12.5%) dose reductions to brainstem for these patients, respectively. Only two patients received a dose to the optic nerves (34.1 Gy and 5.2 Gy). There were small changes (1.8 Gy and 0.5 Gy, respectively) in the dose to optic nerves when comparing the dose calculated on day one and the dose calculated on day 30 CT images. The same two patients received 30.4 Gy and 6.8 Gy to the chiasm, respectively. Small changes in the dose to the chiasm (≤1.1 Gy) were noted between day one and day 30. Conclusions A maximum seed displacement of up to 7.3 mm and a mean seed displacement of 1.8 mm caused by cavity shrinkage were observed during the first month after GammaTile STaRT for gliomas. There were noticeable changes in dosimetry parameters. Changes in the doses to OARs, particularly the brainstem, were large (up to 12.8% of the prescription dose). These changes in dosimetry should be considered when evaluating treatment outcomes and planning future GammaTile treatments. Cureus 2023-05-02 /pmc/articles/PMC10234842/ /pubmed/37273347 http://dx.doi.org/10.7759/cureus.38463 Text en Copyright © 2023, Ng et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Radiation Oncology
Ng, Sook Kien
Yue, Yong
Shiue, Kevin
Shah, Mitesh V
Le, Yi
Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title_full Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title_fullStr Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title_full_unstemmed Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title_short Dosimetric Impact of Source Displacement in GammaTile Surgically Targeted Radiation Therapy for Gliomas
title_sort dosimetric impact of source displacement in gammatile surgically targeted radiation therapy for gliomas
topic Radiation Oncology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10234842/
https://www.ncbi.nlm.nih.gov/pubmed/37273347
http://dx.doi.org/10.7759/cureus.38463
work_keys_str_mv AT ngsookkien dosimetricimpactofsourcedisplacementingammatilesurgicallytargetedradiationtherapyforgliomas
AT yueyong dosimetricimpactofsourcedisplacementingammatilesurgicallytargetedradiationtherapyforgliomas
AT shiuekevin dosimetricimpactofsourcedisplacementingammatilesurgicallytargetedradiationtherapyforgliomas
AT shahmiteshv dosimetricimpactofsourcedisplacementingammatilesurgicallytargetedradiationtherapyforgliomas
AT leyi dosimetricimpactofsourcedisplacementingammatilesurgicallytargetedradiationtherapyforgliomas