Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study
In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless—and continual—BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Nature Singapore
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10239726/ https://www.ncbi.nlm.nih.gov/pubmed/37012424 http://dx.doi.org/10.1038/s41440-023-01258-2 |
Sumario: | In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless—and continual—BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am–9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [−1.5, 4.6] mmHg, P = 0.306; correlation [R(2)]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was found for DBP (−2.2 ± 8.0 [−4.5, 0.1] mmHg, P = 0.058; R(2): 0.66; ±10/±15 mmHg agreements: 78%, 96%). These intermediate results show that daytime BP measurements using the Aktiia monitor generate data comparable to that of an ABPM monitor. |
---|