A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation
BACKGROUND: The energy spectrum is the property of the X-ray tube that describes the energy fluence per unit interval of photon energy. The existing indirect methods for estimating the spectrum ignore the influence caused by the voltage fluctuation of the X-ray tube. METHODS: In this work, we propos...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
AME Publishing Company
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10239992/ https://www.ncbi.nlm.nih.gov/pubmed/37284079 http://dx.doi.org/10.21037/qims-22-1055 |
_version_ | 1785053635409870848 |
---|---|
author | Pan, Zengxiang Li, Xu Lin, Guoqin Qin, Peishan Piao, Zun Huang, Shuang Wu, Wangjiang Qi, Mengke Zhou, Linghong Xu, Yuan |
author_facet | Pan, Zengxiang Li, Xu Lin, Guoqin Qin, Peishan Piao, Zun Huang, Shuang Wu, Wangjiang Qi, Mengke Zhou, Linghong Xu, Yuan |
author_sort | Pan, Zengxiang |
collection | PubMed |
description | BACKGROUND: The energy spectrum is the property of the X-ray tube that describes the energy fluence per unit interval of photon energy. The existing indirect methods for estimating the spectrum ignore the influence caused by the voltage fluctuation of the X-ray tube. METHODS: In this work, we propose a method for estimating the X-ray energy spectrum more accurately by including the voltage fluctuation of the X-ray tube. It expresses the spectrum as the weighted summation of a set of model spectra within a certain voltage fluctuation range. The difference between the raw projection and the estimated projection is considered as the objective function for obtaining the corresponding weight of each model spectrum. The equilibrium optimizer (EO) algorithm is used to find the weight combination that minimizes the objective function. Finally, the estimated spectrum is obtained. We refer to the proposed method as the poly-voltage method. The method is mainly aimed at the cone-beam computed tomography (CBCT) system. RESULTS: The model spectra mixture evaluation and projection evaluation showed that the reference spectrum can be combined by multiple model spectra. They also showed that it is appropriate to choose about 10% of the preset voltage as the voltage range of the model spectra, which can match the reference spectrum and projection quite well. The phantom evaluation showed that the beam-hardening artifact can be corrected using the estimated spectrum via the poly-voltage method, and the poly-voltage method provides not only the accurate reprojection but also an accurate spectrum. The normalized root mean square error (NRMSE) index between the spectrum generated via the poly-voltage method and the reference spectrum could be kept within 3% according to above evaluations. There existed a 1.77% percentage error between the estimated scatter of polymethyl methacrylate (PMMA) phantom using the two spectra generated via the poly-voltage method and the single-voltage method, and it could be considered for scatter simulation. CONCLUSIONS: Our proposed poly-voltage method could estimate the spectrum more accurately for both ideal and more realistic voltage spectra, and it is robust to the different modes of voltage pulse. |
format | Online Article Text |
id | pubmed-10239992 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | AME Publishing Company |
record_format | MEDLINE/PubMed |
spelling | pubmed-102399922023-06-06 A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation Pan, Zengxiang Li, Xu Lin, Guoqin Qin, Peishan Piao, Zun Huang, Shuang Wu, Wangjiang Qi, Mengke Zhou, Linghong Xu, Yuan Quant Imaging Med Surg Original Article BACKGROUND: The energy spectrum is the property of the X-ray tube that describes the energy fluence per unit interval of photon energy. The existing indirect methods for estimating the spectrum ignore the influence caused by the voltage fluctuation of the X-ray tube. METHODS: In this work, we propose a method for estimating the X-ray energy spectrum more accurately by including the voltage fluctuation of the X-ray tube. It expresses the spectrum as the weighted summation of a set of model spectra within a certain voltage fluctuation range. The difference between the raw projection and the estimated projection is considered as the objective function for obtaining the corresponding weight of each model spectrum. The equilibrium optimizer (EO) algorithm is used to find the weight combination that minimizes the objective function. Finally, the estimated spectrum is obtained. We refer to the proposed method as the poly-voltage method. The method is mainly aimed at the cone-beam computed tomography (CBCT) system. RESULTS: The model spectra mixture evaluation and projection evaluation showed that the reference spectrum can be combined by multiple model spectra. They also showed that it is appropriate to choose about 10% of the preset voltage as the voltage range of the model spectra, which can match the reference spectrum and projection quite well. The phantom evaluation showed that the beam-hardening artifact can be corrected using the estimated spectrum via the poly-voltage method, and the poly-voltage method provides not only the accurate reprojection but also an accurate spectrum. The normalized root mean square error (NRMSE) index between the spectrum generated via the poly-voltage method and the reference spectrum could be kept within 3% according to above evaluations. There existed a 1.77% percentage error between the estimated scatter of polymethyl methacrylate (PMMA) phantom using the two spectra generated via the poly-voltage method and the single-voltage method, and it could be considered for scatter simulation. CONCLUSIONS: Our proposed poly-voltage method could estimate the spectrum more accurately for both ideal and more realistic voltage spectra, and it is robust to the different modes of voltage pulse. AME Publishing Company 2023-04-07 2023-06-01 /pmc/articles/PMC10239992/ /pubmed/37284079 http://dx.doi.org/10.21037/qims-22-1055 Text en 2023 Quantitative Imaging in Medicine and Surgery. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) . |
spellingShingle | Original Article Pan, Zengxiang Li, Xu Lin, Guoqin Qin, Peishan Piao, Zun Huang, Shuang Wu, Wangjiang Qi, Mengke Zhou, Linghong Xu, Yuan A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title | A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title_full | A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title_fullStr | A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title_full_unstemmed | A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title_short | A transmission measurement-based spectrum estimation method incorporating X-ray tube voltage fluctuation |
title_sort | transmission measurement-based spectrum estimation method incorporating x-ray tube voltage fluctuation |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10239992/ https://www.ncbi.nlm.nih.gov/pubmed/37284079 http://dx.doi.org/10.21037/qims-22-1055 |
work_keys_str_mv | AT panzengxiang atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT lixu atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT linguoqin atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT qinpeishan atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT piaozun atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT huangshuang atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT wuwangjiang atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT qimengke atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT zhoulinghong atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT xuyuan atransmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT panzengxiang transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT lixu transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT linguoqin transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT qinpeishan transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT piaozun transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT huangshuang transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT wuwangjiang transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT qimengke transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT zhoulinghong transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation AT xuyuan transmissionmeasurementbasedspectrumestimationmethodincorporatingxraytubevoltagefluctuation |