Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac

To evaluate the image quality, dosimetric properties, setup reproducibility, and planar cine motion detection of a high‐resolution brain coil and integrated stereotactic brain immobilization system that constitute a new brain treatment package (BTP) on a low‐field magnetic resonance imaging (MRI) li...

Descripción completa

Detalles Bibliográficos
Autores principales: Snyder, Karen Chin, Mao, Weihua, Kim, Joshua P., Cunningham, Justine, Chetty, Indrin J., Siddiqui, Salim M., Parikh, Parag, Dolan, Jennifer
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10243323/
https://www.ncbi.nlm.nih.gov/pubmed/37278646
http://dx.doi.org/10.1002/acm2.13919
_version_ 1785054402867888128
author Snyder, Karen Chin
Mao, Weihua
Kim, Joshua P.
Cunningham, Justine
Chetty, Indrin J.
Siddiqui, Salim M.
Parikh, Parag
Dolan, Jennifer
author_facet Snyder, Karen Chin
Mao, Weihua
Kim, Joshua P.
Cunningham, Justine
Chetty, Indrin J.
Siddiqui, Salim M.
Parikh, Parag
Dolan, Jennifer
author_sort Snyder, Karen Chin
collection PubMed
description To evaluate the image quality, dosimetric properties, setup reproducibility, and planar cine motion detection of a high‐resolution brain coil and integrated stereotactic brain immobilization system that constitute a new brain treatment package (BTP) on a low‐field magnetic resonance imaging (MRI) linear accelerator (MR‐linac). Image quality of the high‐resolution brain coil was evaluated with the 17 cm diameter spherical phantom and the American College of Radiology (ACR) Large MRI Phantom. Patient imaging studies approved by the institutional review board (IRB) assisted in selecting image acquisition parameters. Radiographic and dosimetric evaluation of the high‐resolution brain coil and the associated immobilization devices was performed using dose calculations and ion chamber measurements. End‐to‐end testing was performed simulating a cranial lesion in a phantom. Inter‐fraction setup variability and motion detection tests were evaluated on four healthy volunteers. Inter‐fraction variability was assessed based on three repeat setups for each volunteer. Motion detection was evaluated using three‐plane (axial, coronal, and sagittal) MR‐cine imaging sessions, where volunteers were asked to perform a set of specific motions. The images were post‐processed and evaluated using an in‐house program. Contrast resolution of the high‐resolution brain coil is superior to the head/neck and torso coils. The BTP receiver coils have an average HU value of 525 HU. The most significant radiation attenuation (3.14%) of the BTP, occurs through the lateral portion of the overlay board where the high‐precision lateral‐profile mask clips attach to the overlay. The greatest inter‐fraction setup variability occurred in the pitch (average 1.08 degree) and translationally in the superior/inferior direction (average 4.88 mm). Three plane cine imaging with the BTP was able to detect large and small motions. Small voluntary motions, sub‐millimeter in magnitude (maximum 0.9 mm), from motion of external limbs were detected. Imaging tests, inter‐fraction setup variability, attenuation, and end‐to‐end measurements were quantified and performed for the BTP. Results demonstrate better contrast resolution and low contrast detectability that allows for better visualization of soft tissue anatomical changes relative to head/neck and torso coil systems.
format Online
Article
Text
id pubmed-10243323
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-102433232023-06-07 Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac Snyder, Karen Chin Mao, Weihua Kim, Joshua P. Cunningham, Justine Chetty, Indrin J. Siddiqui, Salim M. Parikh, Parag Dolan, Jennifer J Appl Clin Med Phys Radiation Oncology Physics To evaluate the image quality, dosimetric properties, setup reproducibility, and planar cine motion detection of a high‐resolution brain coil and integrated stereotactic brain immobilization system that constitute a new brain treatment package (BTP) on a low‐field magnetic resonance imaging (MRI) linear accelerator (MR‐linac). Image quality of the high‐resolution brain coil was evaluated with the 17 cm diameter spherical phantom and the American College of Radiology (ACR) Large MRI Phantom. Patient imaging studies approved by the institutional review board (IRB) assisted in selecting image acquisition parameters. Radiographic and dosimetric evaluation of the high‐resolution brain coil and the associated immobilization devices was performed using dose calculations and ion chamber measurements. End‐to‐end testing was performed simulating a cranial lesion in a phantom. Inter‐fraction setup variability and motion detection tests were evaluated on four healthy volunteers. Inter‐fraction variability was assessed based on three repeat setups for each volunteer. Motion detection was evaluated using three‐plane (axial, coronal, and sagittal) MR‐cine imaging sessions, where volunteers were asked to perform a set of specific motions. The images were post‐processed and evaluated using an in‐house program. Contrast resolution of the high‐resolution brain coil is superior to the head/neck and torso coils. The BTP receiver coils have an average HU value of 525 HU. The most significant radiation attenuation (3.14%) of the BTP, occurs through the lateral portion of the overlay board where the high‐precision lateral‐profile mask clips attach to the overlay. The greatest inter‐fraction setup variability occurred in the pitch (average 1.08 degree) and translationally in the superior/inferior direction (average 4.88 mm). Three plane cine imaging with the BTP was able to detect large and small motions. Small voluntary motions, sub‐millimeter in magnitude (maximum 0.9 mm), from motion of external limbs were detected. Imaging tests, inter‐fraction setup variability, attenuation, and end‐to‐end measurements were quantified and performed for the BTP. Results demonstrate better contrast resolution and low contrast detectability that allows for better visualization of soft tissue anatomical changes relative to head/neck and torso coil systems. John Wiley and Sons Inc. 2023-02-18 /pmc/articles/PMC10243323/ /pubmed/37278646 http://dx.doi.org/10.1002/acm2.13919 Text en © 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Snyder, Karen Chin
Mao, Weihua
Kim, Joshua P.
Cunningham, Justine
Chetty, Indrin J.
Siddiqui, Salim M.
Parikh, Parag
Dolan, Jennifer
Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title_full Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title_fullStr Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title_full_unstemmed Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title_short Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field MR‐linac
title_sort commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low‐field mr‐linac
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10243323/
https://www.ncbi.nlm.nih.gov/pubmed/37278646
http://dx.doi.org/10.1002/acm2.13919
work_keys_str_mv AT snyderkarenchin commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT maoweihua commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT kimjoshuap commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT cunninghamjustine commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT chettyindrinj commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT siddiquisalimm commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT parikhparag commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac
AT dolanjennifer commissioningclinicalimplementationandinitialexperiencewithanewbraintumortreatmentpackageonalowfieldmrlinac