Single-use flexible bronchoscopes vs traditional reusable flexible bronchoscopes: a prospective controlled study

BACKGROUND: Single-use flexible bronchoscopes(SFB) eliminate the risk of bronchoscopy-related infection compared with traditional reusable flexible bronchoscopes(RFB). At present, there is no comparative study between SFB and RFB in the aspects of biopsy and interventional therapy. This study aims t...

Descripción completa

Detalles Bibliográficos
Autores principales: He, Shuzhen, Xie, Lihua, Liu, Jianming, Zou, Lijun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10252171/
https://www.ncbi.nlm.nih.gov/pubmed/37296389
http://dx.doi.org/10.1186/s12890-023-02478-5
Descripción
Sumario:BACKGROUND: Single-use flexible bronchoscopes(SFB) eliminate the risk of bronchoscopy-related infection compared with traditional reusable flexible bronchoscopes(RFB). At present, there is no comparative study between SFB and RFB in the aspects of biopsy and interventional therapy. This study aims to explore whether SFB can perform complex bronchoscopic procedures such as transbronchial biopsies just like RFB. METHODS: We conducted a prospective controlled study. A total of 45 patients who required bronchoscopic biopsy in our hospital from June 2022 to December 2022 were enrolled. The patients were divided into the SFB group and the RFB group, and routine bronchoscopy, bronchoalveolar lavage, and biopsy were performed respectively. Data on the time of routine bronchoscopy, the recovery rate of bronchoalveolar lavage fluid(BALF), biopsy time, and bleeding volume were collected. Then we used the two-sample t-test and the χ(2) test to assess the performance differences between SFB and RFB. We also designed a questionnaire to compare the performance between SFB and RFB by different bronchoscope operators. RESULTS: The routine examination time of SFB and RFB was 3.40 ± 0.50 min and 3.55 ± 0.42 min, respectively. There was no significant difference between the two groups (P = 0.308). The recovery rate of BALF was (46.56 ± 8.22) % in the SFB group and (47.00 ± 8.07) in the RFB group, without a significant difference between the two groups(P = 0.863). The biopsy time was similar(4.67 ± 0.51 min VS 4.57 ± 0.45 min) in both groups, with no significant difference(P = 0.512). The positive biopsy rate was 100% in both groups, with no significant difference. Overall, the bronchoscope operators were generally satisfied with SFB. CONCLUSION: SFBs are non-inferior to RFBs in routine bronchoscopy, bronchoalveolar lavage, and biopsy. It is suggested that SFBs have a wider clinical application.