Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes

Background: Air-pouch balloon-assisted probes have proven to be both simple and reliable tools for intracranial pressure (ICP) monitoring. However, we experienced reproducible falsely high ICP measurements when the ICP probe was inserted into the intracerebral hematoma cavity. Thus, the aim of the e...

Descripción completa

Detalles Bibliográficos
Autores principales: Won, Sae-Yeon, Herrmann, Sascha, Dubinski, Daniel, Behmanesh, Bedjan, Trnovec, Svorad, Dinc, Nazife, Bernstock, Joshua D., Freiman, Thomas M., Gessler, Florian A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10253537/
https://www.ncbi.nlm.nih.gov/pubmed/37297856
http://dx.doi.org/10.3390/jcm12113661
_version_ 1785056429683507200
author Won, Sae-Yeon
Herrmann, Sascha
Dubinski, Daniel
Behmanesh, Bedjan
Trnovec, Svorad
Dinc, Nazife
Bernstock, Joshua D.
Freiman, Thomas M.
Gessler, Florian A.
author_facet Won, Sae-Yeon
Herrmann, Sascha
Dubinski, Daniel
Behmanesh, Bedjan
Trnovec, Svorad
Dinc, Nazife
Bernstock, Joshua D.
Freiman, Thomas M.
Gessler, Florian A.
author_sort Won, Sae-Yeon
collection PubMed
description Background: Air-pouch balloon-assisted probes have proven to be both simple and reliable tools for intracranial pressure (ICP) monitoring. However, we experienced reproducible falsely high ICP measurements when the ICP probe was inserted into the intracerebral hematoma cavity. Thus, the aim of the experimental and translational study was to analyze the influence of ICP probe placement with regard to measured ICP values. Methods: Two Spiegelberg 3PN sensors were simultaneously inserted into a closed drain system and were connected to two separate ICP monitors thereby allowing for simultaneous ICP measurements. This closed system was also engineered to allow for pressure to be gradually increased in a controlled fashion. Once the pressure was verified using two identical ICP probes, one of the probes was coated with blood in an effort to replicate placement within an intraparenchymal hematoma. Pressures recorded using the coated probe and control probe were then recorded and compared across a range of 0–60 mmHg. In an effort to further the translational relevance of our results, two ICP probes were inserted in a patient that presented with a large basal ganglia hemorrhage that met criteria for ICP monitoring. One probe was inserted into the hematoma and the other into brain parenchyma; ICP values were recorded from both probes and the results compared. Results: The experimental set-up demonstrated a reliable correlation between both control ICP probes. Interestingly, the ICP probe covered with clot displayed a significantly higher average ICP value when compared to the control probe between 0 mmHg and 50 mmHg (p < 0.001); at 60 mmHg, there was no significant difference noted. Critically, this trend in discordance was even more pronounced in the clinical setting with the ICP probe placed within the hematoma cavity having reported significantly higher ICP values as compared to the probe within brain parenchyma. Conclusions: Our experimental study and clinical pilot highlight a potential pitfall in ICP measurement that may result secondary to probe placement within hematoma. Such aberrant results may lead to inappropriate interventions in an effort to address falsely elevated ICPs.
format Online
Article
Text
id pubmed-10253537
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-102535372023-06-10 Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes Won, Sae-Yeon Herrmann, Sascha Dubinski, Daniel Behmanesh, Bedjan Trnovec, Svorad Dinc, Nazife Bernstock, Joshua D. Freiman, Thomas M. Gessler, Florian A. J Clin Med Article Background: Air-pouch balloon-assisted probes have proven to be both simple and reliable tools for intracranial pressure (ICP) monitoring. However, we experienced reproducible falsely high ICP measurements when the ICP probe was inserted into the intracerebral hematoma cavity. Thus, the aim of the experimental and translational study was to analyze the influence of ICP probe placement with regard to measured ICP values. Methods: Two Spiegelberg 3PN sensors were simultaneously inserted into a closed drain system and were connected to two separate ICP monitors thereby allowing for simultaneous ICP measurements. This closed system was also engineered to allow for pressure to be gradually increased in a controlled fashion. Once the pressure was verified using two identical ICP probes, one of the probes was coated with blood in an effort to replicate placement within an intraparenchymal hematoma. Pressures recorded using the coated probe and control probe were then recorded and compared across a range of 0–60 mmHg. In an effort to further the translational relevance of our results, two ICP probes were inserted in a patient that presented with a large basal ganglia hemorrhage that met criteria for ICP monitoring. One probe was inserted into the hematoma and the other into brain parenchyma; ICP values were recorded from both probes and the results compared. Results: The experimental set-up demonstrated a reliable correlation between both control ICP probes. Interestingly, the ICP probe covered with clot displayed a significantly higher average ICP value when compared to the control probe between 0 mmHg and 50 mmHg (p < 0.001); at 60 mmHg, there was no significant difference noted. Critically, this trend in discordance was even more pronounced in the clinical setting with the ICP probe placed within the hematoma cavity having reported significantly higher ICP values as compared to the probe within brain parenchyma. Conclusions: Our experimental study and clinical pilot highlight a potential pitfall in ICP measurement that may result secondary to probe placement within hematoma. Such aberrant results may lead to inappropriate interventions in an effort to address falsely elevated ICPs. MDPI 2023-05-25 /pmc/articles/PMC10253537/ /pubmed/37297856 http://dx.doi.org/10.3390/jcm12113661 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Won, Sae-Yeon
Herrmann, Sascha
Dubinski, Daniel
Behmanesh, Bedjan
Trnovec, Svorad
Dinc, Nazife
Bernstock, Joshua D.
Freiman, Thomas M.
Gessler, Florian A.
Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title_full Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title_fullStr Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title_full_unstemmed Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title_short Blood Clots May Compromise Intracranial Pressure Measurement Using Air-Pouch Intracranial Pressure Probes
title_sort blood clots may compromise intracranial pressure measurement using air-pouch intracranial pressure probes
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10253537/
https://www.ncbi.nlm.nih.gov/pubmed/37297856
http://dx.doi.org/10.3390/jcm12113661
work_keys_str_mv AT wonsaeyeon bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT herrmannsascha bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT dubinskidaniel bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT behmaneshbedjan bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT trnovecsvorad bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT dincnazife bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT bernstockjoshuad bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT freimanthomasm bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes
AT gesslerfloriana bloodclotsmaycompromiseintracranialpressuremeasurementusingairpouchintracranialpressureprobes