The Accuracy of Ultrasound Controlled Attenuation Parameter in Diagnosing Hepatic Fat Content
PURPOSE: The Controlled Attenuation Parameter (CAP score) is based on ultrasonic properties of retropropagated radiofrequency signals acquired by Fibroscan(TM) (Echosens, Paris, France). Since ultrasound propagation is influenced by the presence of fat, CAP score was developed to quantify steatosis....
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10263157/ https://www.ncbi.nlm.nih.gov/pubmed/37325088 http://dx.doi.org/10.2147/HMER.S411619 |
Sumario: | PURPOSE: The Controlled Attenuation Parameter (CAP score) is based on ultrasonic properties of retropropagated radiofrequency signals acquired by Fibroscan(TM) (Echosens, Paris, France). Since ultrasound propagation is influenced by the presence of fat, CAP score was developed to quantify steatosis. The aim of this study was to delineate the accuracy of CAP in diagnosing hepatic steatosis, compared to the gold standard of liver biopsy. PATIENTS AND METHODS: A total of 150 patients underwent same-day liver biopsy and measurement of hepatic steatosis with Fibroscan. Only examinations with 10 satisfactory measurements, and an inter-quartile range of less than 30% of the median liver stiffness values were included for data analysis. Histological staging was then correlated with median values and Spearman correlation calculated. P values of <0.05 were considered statistically significant. RESULTS: For diagnosis of hepatic steatosis (HS), CAP could predict the steatosis S2 with AUROC 0.815 (95% CI 0.741–0.889), sensitivity (0.81) and specificity (0.73) when the optimal cut-off value was set at 288 dB/m. CAP detected histological grade S3 with AUROC 0.735 (95% CI 0.618–0.851), sensitivity (0.71) and specificity (0.74), with a cut-off value of 330 dB/m. The AUROC for steatosis grade S1 was 0.741 (95% CI 0.650–0.824), with a cut-off value of 263 dB/m with sensitivity 0.75 and specificity 0.70. Univariate analysis showed a correlation between CAP and diabetes (p 0.048). CONCLUSION: The performance of CAP to diagnose steatosis severity decreases as steatosis progresses. CAP is associated with diabetes but not other clinical factors and parameters of the metabolic syndrome. |
---|