Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study

BACKGROUND: This study evaluated, as a snapshot, the variability in quantification and image quality (IQ) of the clinically utilized PET [(18)F]FDG whole-body protocols in Finland using a NEMA/IEC IQ phantom permanently filled with (68)Ge. METHODS: The phantom was imaged on 14 PET-CT scanners, inclu...

Descripción completa

Detalles Bibliográficos
Autores principales: Sipilä, O., Liukkonen, J., Halme, H.-L., Tolvanen, T., Sohlberg, A., Hakulinen, M., Manninen, A.-L., Tahvanainen, K., Tunninen, V., Ollikainen, T., Kangasmaa, T., Kangasmäki, A., Vuorela, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10271973/
https://www.ncbi.nlm.nih.gov/pubmed/37322376
http://dx.doi.org/10.1186/s40658-023-00551-w
_version_ 1785059413479915520
author Sipilä, O.
Liukkonen, J.
Halme, H.-L.
Tolvanen, T.
Sohlberg, A.
Hakulinen, M.
Manninen, A.-L.
Tahvanainen, K.
Tunninen, V.
Ollikainen, T.
Kangasmaa, T.
Kangasmäki, A.
Vuorela, J.
author_facet Sipilä, O.
Liukkonen, J.
Halme, H.-L.
Tolvanen, T.
Sohlberg, A.
Hakulinen, M.
Manninen, A.-L.
Tahvanainen, K.
Tunninen, V.
Ollikainen, T.
Kangasmaa, T.
Kangasmäki, A.
Vuorela, J.
author_sort Sipilä, O.
collection PubMed
description BACKGROUND: This study evaluated, as a snapshot, the variability in quantification and image quality (IQ) of the clinically utilized PET [(18)F]FDG whole-body protocols in Finland using a NEMA/IEC IQ phantom permanently filled with (68)Ge. METHODS: The phantom was imaged on 14 PET-CT scanners, including a variety of models from two major vendors. The variability of the recovery coefficients (RC(max), RC(mean) and RC(peak)) of the hot spheres as well as percent background variability (PBV), coefficient of variation of the background (COV(BG)) and accuracy of corrections (AOC) were studied using images from clinical and standardized protocols with 20 repeated measurements. The ranges of the RCs were also compared to the limits of the EARL (18)F standards 2 accreditation (EARL2). The impact of image noise on these parameters was studied using averaged images (AVIs). RESULTS: The largest variability in RC values of the routine protocols was found for the RC(max) with a range of 68% and with 10% intra-scanner variability, decreasing to 36% when excluding protocols with suspected cross-calibration failure or without point-spread-function (PSF) correction. The RC ranges of individual hot spheres in routine or standardized protocols or AVIs fulfilled the EARL2 ranges with two minor exceptions, but fulfilling the exact EARL2 limits for all hot spheres was variable. RC(peak) was less dependent on averaging and reconstruction parameters than RC(max) and RC(mean). The PBV, COV(BG) and AOC varied between 2.3–11.8%, 9.6–17.8% and 4.8–32.0%, respectively, for the routine protocols. The RC ranges, PBV and COV(BG) were decreased when using AVIs. With AOC, when excluding routine protocols without PSF correction, the maximum value dropped to 15.5%. CONCLUSION: The maximum variability of the RC values for the [(18)F]FDG whole-body protocols was about 60%. The RC ranges of properly cross-calibrated scanners with PSF correction fitted to the EARL2 RC ranges for individual sphere sizes, but fulfilling the exact RC limits would have needed further optimization. RC(peak) was the most robust RC measure. Besides COV(BG), also RCs and PVB were sensitive to image noise.
format Online
Article
Text
id pubmed-10271973
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-102719732023-06-17 Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study Sipilä, O. Liukkonen, J. Halme, H.-L. Tolvanen, T. Sohlberg, A. Hakulinen, M. Manninen, A.-L. Tahvanainen, K. Tunninen, V. Ollikainen, T. Kangasmaa, T. Kangasmäki, A. Vuorela, J. EJNMMI Phys Original Research BACKGROUND: This study evaluated, as a snapshot, the variability in quantification and image quality (IQ) of the clinically utilized PET [(18)F]FDG whole-body protocols in Finland using a NEMA/IEC IQ phantom permanently filled with (68)Ge. METHODS: The phantom was imaged on 14 PET-CT scanners, including a variety of models from two major vendors. The variability of the recovery coefficients (RC(max), RC(mean) and RC(peak)) of the hot spheres as well as percent background variability (PBV), coefficient of variation of the background (COV(BG)) and accuracy of corrections (AOC) were studied using images from clinical and standardized protocols with 20 repeated measurements. The ranges of the RCs were also compared to the limits of the EARL (18)F standards 2 accreditation (EARL2). The impact of image noise on these parameters was studied using averaged images (AVIs). RESULTS: The largest variability in RC values of the routine protocols was found for the RC(max) with a range of 68% and with 10% intra-scanner variability, decreasing to 36% when excluding protocols with suspected cross-calibration failure or without point-spread-function (PSF) correction. The RC ranges of individual hot spheres in routine or standardized protocols or AVIs fulfilled the EARL2 ranges with two minor exceptions, but fulfilling the exact EARL2 limits for all hot spheres was variable. RC(peak) was less dependent on averaging and reconstruction parameters than RC(max) and RC(mean). The PBV, COV(BG) and AOC varied between 2.3–11.8%, 9.6–17.8% and 4.8–32.0%, respectively, for the routine protocols. The RC ranges, PBV and COV(BG) were decreased when using AVIs. With AOC, when excluding routine protocols without PSF correction, the maximum value dropped to 15.5%. CONCLUSION: The maximum variability of the RC values for the [(18)F]FDG whole-body protocols was about 60%. The RC ranges of properly cross-calibrated scanners with PSF correction fitted to the EARL2 RC ranges for individual sphere sizes, but fulfilling the exact RC limits would have needed further optimization. RC(peak) was the most robust RC measure. Besides COV(BG), also RCs and PVB were sensitive to image noise. Springer International Publishing 2023-06-16 /pmc/articles/PMC10271973/ /pubmed/37322376 http://dx.doi.org/10.1186/s40658-023-00551-w Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Sipilä, O.
Liukkonen, J.
Halme, H.-L.
Tolvanen, T.
Sohlberg, A.
Hakulinen, M.
Manninen, A.-L.
Tahvanainen, K.
Tunninen, V.
Ollikainen, T.
Kangasmaa, T.
Kangasmäki, A.
Vuorela, J.
Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title_full Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title_fullStr Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title_full_unstemmed Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title_short Variability in PET image quality and quantification measured with a permanently filled (68)Ge-phantom: a multi-center study
title_sort variability in pet image quality and quantification measured with a permanently filled (68)ge-phantom: a multi-center study
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10271973/
https://www.ncbi.nlm.nih.gov/pubmed/37322376
http://dx.doi.org/10.1186/s40658-023-00551-w
work_keys_str_mv AT sipilao variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT liukkonenj variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT halmehl variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT tolvanent variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT sohlberga variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT hakulinenm variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT manninenal variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT tahvanainenk variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT tunninenv variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT ollikainent variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT kangasmaat variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT kangasmakia variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy
AT vuorelaj variabilityinpetimagequalityandquantificationmeasuredwithapermanentlyfilled68gephantomamulticenterstudy