Cargando…

Central and peripheral refraction measured by a novel double-pass instrument

A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye’s central and peripheral poin...

Descripción completa

Detalles Bibliográficos
Autores principales: Christaras, Dimitrios, Tsoukalas, Spyridon, Papadogiannis, Petros, Börjeson, Charlie, Volny, Moa, Lundström, Linda, Artal, Pablo, Ginis, Harilaos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Optica Publishing Group 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10278616/
https://www.ncbi.nlm.nih.gov/pubmed/37342694
http://dx.doi.org/10.1364/BOE.489881
Descripción
Sumario:A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye’s central and peripheral point-spread function (PSF) using an infrared laser source, a tunable lens and a CMOS camera. The through-focus images were analyzed to determine defocus and astigmatism at 0° and 30° visual field. These values were compared to those obtained with a lab-based Hartmann-Shack wavefront sensor. The two instruments provided data showing good correlation at both eccentricities, particularly in the estimation of defocus.