A method for selecting reference beam model of VMAT plans with three 6MV beam-matched linear accelerators during radiation oncology

Our objective was to provide a method for selecting reference beam model and evaluating the dosimetric accuracy of volumetric modulated arc therapy (VMAT) plans delivered on three Elekta beam-matched linacs during radiation oncology. Beam data was measured on three beam-matched linacs including Syne...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Yi, Wu, Wenjing, Yuan, Wei, Chai, Linyan, Tang, Fengwen, He, Ruixin, Lu, Yongkai, Zhang, Yuemei, Wang, Long, Wang, Mengfei, Zhang, Xiaozhi, Zhang, Long
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10287655/
https://www.ncbi.nlm.nih.gov/pubmed/37349531
http://dx.doi.org/10.1038/s41598-023-36930-7
Descripción
Sumario:Our objective was to provide a method for selecting reference beam model and evaluating the dosimetric accuracy of volumetric modulated arc therapy (VMAT) plans delivered on three Elekta beam-matched linacs during radiation oncology. Beam data was measured on three beam-matched linacs including Synergy1, Synergy2 and VersaHD. For eighteen lung and esophagus cases, fifty-four plans were generated using VMAT technique with three linac beam models respectively for point dose measurement and three-dimensional dose measurement. Each VMAT plan was executed sequentially on three linacs respectively. Measurement results were compared with treatment planning system (TPS) calculation results for all VMAT plans. Among three beam-matched linacs, discrepancy in beam output factor, percentage depth dose at 5 cm, 10 cm, 20 cm depth and MLC leaf offset are all within 1% except 20 × 20 cm(2) and 30 × 30 cm(2) field sizes, and discrepancy in beam profile is all within 2%. With comparison between measurement result and TPS calculation result, the absolute dose deviations are within the range of ± 3%, and the gamma passing rates are all over 95% for all VMAT plans, which are within the tolerance of clinical acceptability. Compared with all plans delivered on Synegy1 and VersaHD, the point dose discrepancy between measured results and TPS calculated results for plans delivered on Synergy2 is smallest, and the gamma passing rate between measured results and TPS calculated results for plans delivered on Synergy2 is highest. The beam-matched linacs demonstrate good agreement between measurement result and TPS calculation result for VMAT plans. The method can be used for selecting reference beam model for VMAT plans.