Cargando…
Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma
PURPOSE: Measuring the spatial extent of defects may be advantageous in advanced glaucoma where conventional perimetric sensitivity measurements are unreliable. We test whether suprathreshold tests on a higher density grid can more efficiently map advanced visual field loss. METHODS: Data from 97 pa...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10297785/ https://www.ncbi.nlm.nih.gov/pubmed/37358492 http://dx.doi.org/10.1167/tvst.12.6.19 |
_version_ | 1785063959161733120 |
---|---|
author | Denniss, Jonathan McKendrick, Allison M. Turpin, Andrew |
author_facet | Denniss, Jonathan McKendrick, Allison M. Turpin, Andrew |
author_sort | Denniss, Jonathan |
collection | PubMed |
description | PURPOSE: Measuring the spatial extent of defects may be advantageous in advanced glaucoma where conventional perimetric sensitivity measurements are unreliable. We test whether suprathreshold tests on a higher density grid can more efficiently map advanced visual field loss. METHODS: Data from 97 patients with mean deviation < –10 dB were used in simulations comparing two suprathreshold procedures (on a high-density 1.5° grid) to interpolated Full Threshold 24-2. Spatial binary search (SpaBS) presented 20-dB stimuli at locations bisecting seen/unseen points until the seen status of all neighbors matched or until tested points were adjacent. The SupraThreshold Adaptive Mapping Procedure (STAMP) presented 20-dB stimuli where entropy was maximal and modified the status of all points after each presentation, stopping after a fixed number of presentations (estimated as 50%–100% of the presentation number of a current procedure). RESULTS: With typical response errors, SpaBS had worse mean accuracy and repeatability than Full Threshold (both P < 0.0001). Compared to Full Threshold, mean accuracy (Full Threshold: median, 91%; interquartile range [IQR], 87%–94%) was slightly better with STAMP for all stopping criteria, although this was not statistically significant until 100% of conventional test presentations were used. Mean repeatability for STAMP was similar for all stopping criteria (P ≥ 0.02) compared to Full Threshold (Full Threshold: median, 89%; IQR, 82%–93%). CONCLUSIONS: STAMP accurately and repeatably maps the spatial extent of advanced visual field defects in as few as 50% of conventional perimetric test presentations. Further work is needed to test STAMP in human observers and in progressive loss. TRANSLATIONAL RELEVANCE: New perimetric approaches may improve information available for advanced glaucoma management and may potentially be more acceptable to patients. |
format | Online Article Text |
id | pubmed-10297785 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | The Association for Research in Vision and Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-102977852023-06-28 Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma Denniss, Jonathan McKendrick, Allison M. Turpin, Andrew Transl Vis Sci Technol Glaucoma PURPOSE: Measuring the spatial extent of defects may be advantageous in advanced glaucoma where conventional perimetric sensitivity measurements are unreliable. We test whether suprathreshold tests on a higher density grid can more efficiently map advanced visual field loss. METHODS: Data from 97 patients with mean deviation < –10 dB were used in simulations comparing two suprathreshold procedures (on a high-density 1.5° grid) to interpolated Full Threshold 24-2. Spatial binary search (SpaBS) presented 20-dB stimuli at locations bisecting seen/unseen points until the seen status of all neighbors matched or until tested points were adjacent. The SupraThreshold Adaptive Mapping Procedure (STAMP) presented 20-dB stimuli where entropy was maximal and modified the status of all points after each presentation, stopping after a fixed number of presentations (estimated as 50%–100% of the presentation number of a current procedure). RESULTS: With typical response errors, SpaBS had worse mean accuracy and repeatability than Full Threshold (both P < 0.0001). Compared to Full Threshold, mean accuracy (Full Threshold: median, 91%; interquartile range [IQR], 87%–94%) was slightly better with STAMP for all stopping criteria, although this was not statistically significant until 100% of conventional test presentations were used. Mean repeatability for STAMP was similar for all stopping criteria (P ≥ 0.02) compared to Full Threshold (Full Threshold: median, 89%; IQR, 82%–93%). CONCLUSIONS: STAMP accurately and repeatably maps the spatial extent of advanced visual field defects in as few as 50% of conventional perimetric test presentations. Further work is needed to test STAMP in human observers and in progressive loss. TRANSLATIONAL RELEVANCE: New perimetric approaches may improve information available for advanced glaucoma management and may potentially be more acceptable to patients. The Association for Research in Vision and Ophthalmology 2023-06-26 /pmc/articles/PMC10297785/ /pubmed/37358492 http://dx.doi.org/10.1167/tvst.12.6.19 Text en Copyright 2023 The Authors https://creativecommons.org/licenses/by/4.0/This work is licensed under a Creative Commons Attribution 4.0 International License. |
spellingShingle | Glaucoma Denniss, Jonathan McKendrick, Allison M. Turpin, Andrew Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title | Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title_full | Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title_fullStr | Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title_full_unstemmed | Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title_short | Suprathreshold Approaches to Mapping the Visual Field in Advanced Glaucoma |
title_sort | suprathreshold approaches to mapping the visual field in advanced glaucoma |
topic | Glaucoma |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10297785/ https://www.ncbi.nlm.nih.gov/pubmed/37358492 http://dx.doi.org/10.1167/tvst.12.6.19 |
work_keys_str_mv | AT dennissjonathan suprathresholdapproachestomappingthevisualfieldinadvancedglaucoma AT mckendrickallisonm suprathresholdapproachestomappingthevisualfieldinadvancedglaucoma AT turpinandrew suprathresholdapproachestomappingthevisualfieldinadvancedglaucoma |