Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees
The objective of this study is to examine the mechanical behavior of two-piece abutments (Morse taper with 16° internal angulation and Morse taper with 11.5° internal angulation) before and after cyclic fatigue testing, following ISO 14801:2016 guidelines. The specimens were divided into three group...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10298607/ https://www.ncbi.nlm.nih.gov/pubmed/37372893 http://dx.doi.org/10.3390/healthcare11121775 |
_version_ | 1785064158509662208 |
---|---|
author | Tolesano Loureiro, Karine Regina Boaventura de Moura, Marcos de Oliveira, Guilherme José Pimentel Lopes Resende Novais, Veridiana Zancopé, Karla Simamoto Júnior, Paulo Cézar |
author_facet | Tolesano Loureiro, Karine Regina Boaventura de Moura, Marcos de Oliveira, Guilherme José Pimentel Lopes Resende Novais, Veridiana Zancopé, Karla Simamoto Júnior, Paulo Cézar |
author_sort | Tolesano Loureiro, Karine Regina |
collection | PubMed |
description | The objective of this study is to examine the mechanical behavior of two-piece abutments (Morse taper with 16° internal angulation and Morse taper with 11.5° internal angulation) before and after cyclic fatigue testing, following ISO 14801:2016 guidelines. The specimens were divided into three groups: a modified Morse taper with a taper angle of 16° (GM group), a conventional Morse taper (taper angle of 11.5° deg) with a two-piece (CMt group), and one-piece abutments (CMo group). Each experimental group was formed by ten implants and ten abutments (n = 10) for a total of 30 specimens (n = 30). The abutments were tightened and loosened, and a fatigue test was applied with 15 Hz and 5 × 10(6) cycles. Subsequently, the abutments were loosened, and a pull-out test was performed on the CMt group. Finite element analysis (FEA) was conducted on stress concentration regions. The statistical analysis of the loosening test was performed using two-way ANOVA and Tukey’s tests (p < 0.05) to compare screw loosening within each group and between the groups with and without mechanical fatigue. Significant differences were found among the three groups in the loosening test when analyzing the values with and without fatigue (p < 0.001) within each group. When the groups were compared with each other, there was also a significant difference between them (p < 0.001), except between groups GM and CMt without fatigue (p = 0.840). In the pull-out test of the CMt group, the sample exhibited frictional locking only after fatigue (mean = 94.2 N). The FEA demonstrated a varied stress distribution in all groups. The stress was found to be more concentrated in the upper third and middle third regions of the implant, as well as in the opposite region of the load application for all three groups. Although the CMo group showed lower rates of loosening, it displayed a poorer stress distribution in comparison to the GM and CMt groups. On the other hand, the CMt group exhibited a satisfactory frictional lock after undergoing the fatigue tests. |
format | Online Article Text |
id | pubmed-10298607 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-102986072023-06-28 Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees Tolesano Loureiro, Karine Regina Boaventura de Moura, Marcos de Oliveira, Guilherme José Pimentel Lopes Resende Novais, Veridiana Zancopé, Karla Simamoto Júnior, Paulo Cézar Healthcare (Basel) Article The objective of this study is to examine the mechanical behavior of two-piece abutments (Morse taper with 16° internal angulation and Morse taper with 11.5° internal angulation) before and after cyclic fatigue testing, following ISO 14801:2016 guidelines. The specimens were divided into three groups: a modified Morse taper with a taper angle of 16° (GM group), a conventional Morse taper (taper angle of 11.5° deg) with a two-piece (CMt group), and one-piece abutments (CMo group). Each experimental group was formed by ten implants and ten abutments (n = 10) for a total of 30 specimens (n = 30). The abutments were tightened and loosened, and a fatigue test was applied with 15 Hz and 5 × 10(6) cycles. Subsequently, the abutments were loosened, and a pull-out test was performed on the CMt group. Finite element analysis (FEA) was conducted on stress concentration regions. The statistical analysis of the loosening test was performed using two-way ANOVA and Tukey’s tests (p < 0.05) to compare screw loosening within each group and between the groups with and without mechanical fatigue. Significant differences were found among the three groups in the loosening test when analyzing the values with and without fatigue (p < 0.001) within each group. When the groups were compared with each other, there was also a significant difference between them (p < 0.001), except between groups GM and CMt without fatigue (p = 0.840). In the pull-out test of the CMt group, the sample exhibited frictional locking only after fatigue (mean = 94.2 N). The FEA demonstrated a varied stress distribution in all groups. The stress was found to be more concentrated in the upper third and middle third regions of the implant, as well as in the opposite region of the load application for all three groups. Although the CMo group showed lower rates of loosening, it displayed a poorer stress distribution in comparison to the GM and CMt groups. On the other hand, the CMt group exhibited a satisfactory frictional lock after undergoing the fatigue tests. MDPI 2023-06-16 /pmc/articles/PMC10298607/ /pubmed/37372893 http://dx.doi.org/10.3390/healthcare11121775 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Tolesano Loureiro, Karine Regina Boaventura de Moura, Marcos de Oliveira, Guilherme José Pimentel Lopes Resende Novais, Veridiana Zancopé, Karla Simamoto Júnior, Paulo Cézar Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title | Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title_full | Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title_fullStr | Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title_full_unstemmed | Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title_short | Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees |
title_sort | mechanical behavior of one-piece and two-piece tapered prosthetic abutments with 11.5 and 16 degrees |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10298607/ https://www.ncbi.nlm.nih.gov/pubmed/37372893 http://dx.doi.org/10.3390/healthcare11121775 |
work_keys_str_mv | AT tolesanoloureirokarineregina mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees AT boaventurademouramarcos mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees AT deoliveiraguilhermejosepimentellopes mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees AT resendenovaisveridiana mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees AT zancopekarla mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees AT simamotojuniorpaulocezar mechanicalbehaviorofonepieceandtwopiecetaperedprostheticabutmentswith115and16degrees |