Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer

This study presents a methodology that combines experimental tests and the finite element method, which is able to analyse the influence of the geometry on the mechanical behaviour of stents made of bioabsorbable polymer PLA (PolyLactic Acid) during their expansion in the treatment of coarctation of...

Descripción completa

Detalles Bibliográficos
Autores principales: dos Santos, Flávio José, Hernandez, Bruno Agostinho, Santos, Rosana, Machado, Marcel, Souza, Mateus, Capello Sousa, Edson A., Andrade, Aron
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10301593/
https://www.ncbi.nlm.nih.gov/pubmed/37374585
http://dx.doi.org/10.3390/ma16124403
_version_ 1785064848941383680
author dos Santos, Flávio José
Hernandez, Bruno Agostinho
Santos, Rosana
Machado, Marcel
Souza, Mateus
Capello Sousa, Edson A.
Andrade, Aron
author_facet dos Santos, Flávio José
Hernandez, Bruno Agostinho
Santos, Rosana
Machado, Marcel
Souza, Mateus
Capello Sousa, Edson A.
Andrade, Aron
author_sort dos Santos, Flávio José
collection PubMed
description This study presents a methodology that combines experimental tests and the finite element method, which is able to analyse the influence of the geometry on the mechanical behaviour of stents made of bioabsorbable polymer PLA (PolyLactic Acid) during their expansion in the treatment of coarctation of the aorta (CoA). Tensile tests with standardized specimen samples were conducted to determine the properties of a 3D-printed PLA. A finite element model of a new stent prototype was generated from CAD files. A rigid cylinder simulating the expansion balloon was also created to simulate the stent opening performance. A tensile test with 3D-printed customized stent specimens was performed to validate the FE stent model. Stent performance was evaluated in terms of elastic return, recoil, and stress levels. The 3D-printed PLA presented an elastic modulus of 1.5 GPa and a yield strength of 30.6 MPa, lower than non-3D-printed PLA. It can also be inferred that crimping had little effect on stent circular recoil performance, as the difference between the two scenarios was on average 1.81%. For an expansion of diameters ranging from 12 mm to 15 mm, as the maximum opening diameter increases, the recoil levels decrease, ranging from 10 to 16.75% within the reported range. These results point out the importance of testing the 3D-printed PLA under the conditions of using it to access its material properties; the results also indicate that the crimping process could be disregarded in simulations to obtain fast results with lower computational cost and that new proposed stent geometry made of PLA might be suitable for use in CoA treatments—the approach that has not been applied before. The next steps will be to simulate the opening of an aorta vessel using this geometry.
format Online
Article
Text
id pubmed-10301593
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-103015932023-06-29 Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer dos Santos, Flávio José Hernandez, Bruno Agostinho Santos, Rosana Machado, Marcel Souza, Mateus Capello Sousa, Edson A. Andrade, Aron Materials (Basel) Article This study presents a methodology that combines experimental tests and the finite element method, which is able to analyse the influence of the geometry on the mechanical behaviour of stents made of bioabsorbable polymer PLA (PolyLactic Acid) during their expansion in the treatment of coarctation of the aorta (CoA). Tensile tests with standardized specimen samples were conducted to determine the properties of a 3D-printed PLA. A finite element model of a new stent prototype was generated from CAD files. A rigid cylinder simulating the expansion balloon was also created to simulate the stent opening performance. A tensile test with 3D-printed customized stent specimens was performed to validate the FE stent model. Stent performance was evaluated in terms of elastic return, recoil, and stress levels. The 3D-printed PLA presented an elastic modulus of 1.5 GPa and a yield strength of 30.6 MPa, lower than non-3D-printed PLA. It can also be inferred that crimping had little effect on stent circular recoil performance, as the difference between the two scenarios was on average 1.81%. For an expansion of diameters ranging from 12 mm to 15 mm, as the maximum opening diameter increases, the recoil levels decrease, ranging from 10 to 16.75% within the reported range. These results point out the importance of testing the 3D-printed PLA under the conditions of using it to access its material properties; the results also indicate that the crimping process could be disregarded in simulations to obtain fast results with lower computational cost and that new proposed stent geometry made of PLA might be suitable for use in CoA treatments—the approach that has not been applied before. The next steps will be to simulate the opening of an aorta vessel using this geometry. MDPI 2023-06-15 /pmc/articles/PMC10301593/ /pubmed/37374585 http://dx.doi.org/10.3390/ma16124403 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
dos Santos, Flávio José
Hernandez, Bruno Agostinho
Santos, Rosana
Machado, Marcel
Souza, Mateus
Capello Sousa, Edson A.
Andrade, Aron
Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title_full Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title_fullStr Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title_full_unstemmed Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title_short Bioabsorbable Polymeric Stent for the Treatment of Coarctation of the Aorta (CoA) in Children: A Methodology to Evaluate the Design and Mechanical Properties of PLA Polymer
title_sort bioabsorbable polymeric stent for the treatment of coarctation of the aorta (coa) in children: a methodology to evaluate the design and mechanical properties of pla polymer
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10301593/
https://www.ncbi.nlm.nih.gov/pubmed/37374585
http://dx.doi.org/10.3390/ma16124403
work_keys_str_mv AT dossantosflaviojose bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT hernandezbrunoagostinho bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT santosrosana bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT machadomarcel bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT souzamateus bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT capellosousaedsona bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer
AT andradearon bioabsorbablepolymericstentforthetreatmentofcoarctationoftheaortacoainchildrenamethodologytoevaluatethedesignandmechanicalpropertiesofplapolymer