All surgical supra-annular aortic valvar tissue prostheses are labelled too large

OBJECTIVES: Surgical supra-annular aortic valvar tissue prostheses are labelled in an inconsistent and confusing manner. If the replaced valve is too small for a given patient, the risk of patient-prosthesis mismatch is increased, which is associated with the risk of morbidity and mortality. The lab...

Descripción completa

Detalles Bibliográficos
Autores principales: van Boxtel, Astrid Gerritje Maria, Mariani, Massimo Alessandro, Ebels, Tjark
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10301699/
https://www.ncbi.nlm.nih.gov/pubmed/37184926
http://dx.doi.org/10.1093/icvts/ivad076
_version_ 1785064874373545984
author van Boxtel, Astrid Gerritje Maria
Mariani, Massimo Alessandro
Ebels, Tjark
author_facet van Boxtel, Astrid Gerritje Maria
Mariani, Massimo Alessandro
Ebels, Tjark
author_sort van Boxtel, Astrid Gerritje Maria
collection PubMed
description OBJECTIVES: Surgical supra-annular aortic valvar tissue prostheses are labelled in an inconsistent and confusing manner. If the replaced valve is too small for a given patient, the risk of patient-prosthesis mismatch is increased, which is associated with the risk of morbidity and mortality. The labelled diameter (LD) of these valves should coincide with the inflow orifice diameter (IOD). Therefore, our goal was to measure all relevant IODs. METHODS: Valvar design was assessed in terms of the intended position of the valve in relation to the patient’s annulus. The IODs of all available supra-annular aortic valvar prostheses were measured using a conical gauge. The IODs were compared to the LDs. We searched for instructions for use, websites, packing boxes and regulatory institutions involved in the process. RESULTS: Eight valve models from 4 manufacturers were included. None of these valves were clearly labelled as supra-annular on the packing box, although for 3, the supra-annular label could be found in the written specifications. All valves had an IOD smaller than their LD, with a median difference of 15% (range: 4%–25%). The departure from LD differed per valve model and valve size. CONCLUSIONS: Valve packages should be labelled accurately and clearly so that surgeons can make a well-informed choice. Currently essential information is missing because the intended position in relation to the annulus is not consistently marked on the packing boxes, and valve sizes are labelled incorrectly. We propose a change for the better: relabel all valves according to their true IOD in a structured manner.
format Online
Article
Text
id pubmed-10301699
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-103016992023-06-29 All surgical supra-annular aortic valvar tissue prostheses are labelled too large van Boxtel, Astrid Gerritje Maria Mariani, Massimo Alessandro Ebels, Tjark Interdiscip Cardiovasc Thorac Surg Valvular Heart Disease OBJECTIVES: Surgical supra-annular aortic valvar tissue prostheses are labelled in an inconsistent and confusing manner. If the replaced valve is too small for a given patient, the risk of patient-prosthesis mismatch is increased, which is associated with the risk of morbidity and mortality. The labelled diameter (LD) of these valves should coincide with the inflow orifice diameter (IOD). Therefore, our goal was to measure all relevant IODs. METHODS: Valvar design was assessed in terms of the intended position of the valve in relation to the patient’s annulus. The IODs of all available supra-annular aortic valvar prostheses were measured using a conical gauge. The IODs were compared to the LDs. We searched for instructions for use, websites, packing boxes and regulatory institutions involved in the process. RESULTS: Eight valve models from 4 manufacturers were included. None of these valves were clearly labelled as supra-annular on the packing box, although for 3, the supra-annular label could be found in the written specifications. All valves had an IOD smaller than their LD, with a median difference of 15% (range: 4%–25%). The departure from LD differed per valve model and valve size. CONCLUSIONS: Valve packages should be labelled accurately and clearly so that surgeons can make a well-informed choice. Currently essential information is missing because the intended position in relation to the annulus is not consistently marked on the packing boxes, and valve sizes are labelled incorrectly. We propose a change for the better: relabel all valves according to their true IOD in a structured manner. Oxford University Press 2023-05-15 /pmc/articles/PMC10301699/ /pubmed/37184926 http://dx.doi.org/10.1093/icvts/ivad076 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Valvular Heart Disease
van Boxtel, Astrid Gerritje Maria
Mariani, Massimo Alessandro
Ebels, Tjark
All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title_full All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title_fullStr All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title_full_unstemmed All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title_short All surgical supra-annular aortic valvar tissue prostheses are labelled too large
title_sort all surgical supra-annular aortic valvar tissue prostheses are labelled too large
topic Valvular Heart Disease
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10301699/
https://www.ncbi.nlm.nih.gov/pubmed/37184926
http://dx.doi.org/10.1093/icvts/ivad076
work_keys_str_mv AT vanboxtelastridgerritjemaria allsurgicalsupraannularaorticvalvartissueprosthesesarelabelledtoolarge
AT marianimassimoalessandro allsurgicalsupraannularaorticvalvartissueprosthesesarelabelledtoolarge
AT ebelstjark allsurgicalsupraannularaorticvalvartissueprosthesesarelabelledtoolarge