A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques
Objective: Devices for cuffless blood pressure (BP) measurement have become increasingly widespread in recent years. Non-invasive continuous BP monitor (BPM) devices can diagnose potential hypertensive patients at an early stage; however, these cuffless BPMs require more reliable pulse wave simulati...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10305378/ https://www.ncbi.nlm.nih.gov/pubmed/37374803 http://dx.doi.org/10.3390/mi14061218 |
_version_ | 1785065719445061632 |
---|---|
author | Guo, Cheng-Yan Perng, Jau-Woei Chen, Li-Ching Hsieh, Tung-Li |
author_facet | Guo, Cheng-Yan Perng, Jau-Woei Chen, Li-Ching Hsieh, Tung-Li |
author_sort | Guo, Cheng-Yan |
collection | PubMed |
description | Objective: Devices for cuffless blood pressure (BP) measurement have become increasingly widespread in recent years. Non-invasive continuous BP monitor (BPM) devices can diagnose potential hypertensive patients at an early stage; however, these cuffless BPMs require more reliable pulse wave simulation equipment and verification methods. Therefore, we propose a device to simulate human pulse wave signals that can test the accuracy of cuffless BPM devices using pulse wave velocity (PWV). Methods: We design and develop a simulator capable of simulating human pulse waves comprising an electromechanical system to simulate the circulatory system and an arm model-embedded arterial phantom. These parts form a pulse wave simulator with hemodynamic characteristics. We use a cuffless device for measuring local PWV as the device under test to measure the PWV of the pulse wave simulator. We then use a hemodynamic model to fit the cuffless BPM and pulse wave simulator results; this model can rapidly calibrate the cuffless BPM’s hemodynamic measurement performance. Results: We first used multiple linear regression (MLR) to generate a cuffless BPM calibration model and then investigated differences between the measured PWV with and without MLR model calibration. The mean absolute error of the studied cuffless BPM without the MLR model is 0.77 m/s, which improves to 0.06 m/s when using the model for calibration. The measurement error of the cuffless BPM at BPs of 100–180 mmHg is 1.7–5.99 mmHg before calibration, which decreases to 0.14–0.48 mmHg after calibration. Conclusion: This study proposes a design of a pulse wave simulator based on hemodynamic characteristics and provides a standard performance verification method for cuffless BPMs that requires only MLR modeling on the cuffless BPM and pulse wave simulator. The pulse wave simulator proposed in this study can be used to quantitively assess the performance of cuffless BPMs. The proposed pulse wave simulator is suitable for mass production for the verification of cuffless BPMs. As cuffless BPMs become increasingly widespread, this study can provide performance testing standards for cuffless devices. |
format | Online Article Text |
id | pubmed-10305378 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-103053782023-06-29 A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques Guo, Cheng-Yan Perng, Jau-Woei Chen, Li-Ching Hsieh, Tung-Li Micromachines (Basel) Article Objective: Devices for cuffless blood pressure (BP) measurement have become increasingly widespread in recent years. Non-invasive continuous BP monitor (BPM) devices can diagnose potential hypertensive patients at an early stage; however, these cuffless BPMs require more reliable pulse wave simulation equipment and verification methods. Therefore, we propose a device to simulate human pulse wave signals that can test the accuracy of cuffless BPM devices using pulse wave velocity (PWV). Methods: We design and develop a simulator capable of simulating human pulse waves comprising an electromechanical system to simulate the circulatory system and an arm model-embedded arterial phantom. These parts form a pulse wave simulator with hemodynamic characteristics. We use a cuffless device for measuring local PWV as the device under test to measure the PWV of the pulse wave simulator. We then use a hemodynamic model to fit the cuffless BPM and pulse wave simulator results; this model can rapidly calibrate the cuffless BPM’s hemodynamic measurement performance. Results: We first used multiple linear regression (MLR) to generate a cuffless BPM calibration model and then investigated differences between the measured PWV with and without MLR model calibration. The mean absolute error of the studied cuffless BPM without the MLR model is 0.77 m/s, which improves to 0.06 m/s when using the model for calibration. The measurement error of the cuffless BPM at BPs of 100–180 mmHg is 1.7–5.99 mmHg before calibration, which decreases to 0.14–0.48 mmHg after calibration. Conclusion: This study proposes a design of a pulse wave simulator based on hemodynamic characteristics and provides a standard performance verification method for cuffless BPMs that requires only MLR modeling on the cuffless BPM and pulse wave simulator. The pulse wave simulator proposed in this study can be used to quantitively assess the performance of cuffless BPMs. The proposed pulse wave simulator is suitable for mass production for the verification of cuffless BPMs. As cuffless BPMs become increasingly widespread, this study can provide performance testing standards for cuffless devices. MDPI 2023-06-09 /pmc/articles/PMC10305378/ /pubmed/37374803 http://dx.doi.org/10.3390/mi14061218 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Guo, Cheng-Yan Perng, Jau-Woei Chen, Li-Ching Hsieh, Tung-Li A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title | A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title_full | A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title_fullStr | A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title_full_unstemmed | A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title_short | A Hemodynamic Pulse Wave Simulator Designed for Calibration of Local Pulse Wave Velocities Measurement for Cuffless Techniques |
title_sort | hemodynamic pulse wave simulator designed for calibration of local pulse wave velocities measurement for cuffless techniques |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10305378/ https://www.ncbi.nlm.nih.gov/pubmed/37374803 http://dx.doi.org/10.3390/mi14061218 |
work_keys_str_mv | AT guochengyan ahemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT perngjauwoei ahemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT chenliching ahemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT hsiehtungli ahemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT guochengyan hemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT perngjauwoei hemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT chenliching hemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques AT hsiehtungli hemodynamicpulsewavesimulatordesignedforcalibrationoflocalpulsewavevelocitiesmeasurementforcufflesstechniques |