Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis

The Amplatzer Amulet (AA) and Watchman devices (WD) are the 2 most frequently used devices for percutaneous LAA closure globally. OBJECTIVE: To evaluate the safety and clinical outcomes associated with these 2 devices in patients undergoing percutaneous LAA closure. METHOD: We systematically searche...

Descripción completa

Detalles Bibliográficos
Autores principales: Jaiswal, Vikash, Ang, Song Peng, Shrestha, Abhigan Babu, Wajid, Zarghoona, Endurance, Evbayekha Osas, Ayoobkhan, Fathima Shehnaz, Khan, Shazia, Garimellla, Vamsi, Huang, Helen, Ghanim, Mohammed, Song, David, Sharma, Prachi, Alraies, M. Chadi, Biswas, Monodeep
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10313297/
https://www.ncbi.nlm.nih.gov/pubmed/37390240
http://dx.doi.org/10.1097/MD.0000000000034185
_version_ 1785067095760830464
author Jaiswal, Vikash
Ang, Song Peng
Shrestha, Abhigan Babu
Wajid, Zarghoona
Endurance, Evbayekha Osas
Ayoobkhan, Fathima Shehnaz
Khan, Shazia
Garimellla, Vamsi
Huang, Helen
Ghanim, Mohammed
Song, David
Sharma, Prachi
Alraies, M. Chadi
Biswas, Monodeep
author_facet Jaiswal, Vikash
Ang, Song Peng
Shrestha, Abhigan Babu
Wajid, Zarghoona
Endurance, Evbayekha Osas
Ayoobkhan, Fathima Shehnaz
Khan, Shazia
Garimellla, Vamsi
Huang, Helen
Ghanim, Mohammed
Song, David
Sharma, Prachi
Alraies, M. Chadi
Biswas, Monodeep
author_sort Jaiswal, Vikash
collection PubMed
description The Amplatzer Amulet (AA) and Watchman devices (WD) are the 2 most frequently used devices for percutaneous LAA closure globally. OBJECTIVE: To evaluate the safety and clinical outcomes associated with these 2 devices in patients undergoing percutaneous LAA closure. METHOD: We systematically searched all electronic databases from inception until February 21, 2023. The primary endpoint was procedure related complications. Secondary endpoints were device related thrombus, stroke, cardiovascular mortality, peri device leak, systemic embolism, and all-cause mortality. RESULTS: A total of 3 randomized clinical trials with 2150 patients were included in this meta-analysis. The mean age was 75 and 76 years in the Amplatzer group and in the Watchman group, respectively. The odds of procedure-related complications (OR, 1.80 [95% CI: 1.21–2.67], P < .001) were significantly higher among patients with AA compared to the WD. However, the odds of all-cause mortality (OR, 0.75 (95% CI: 0.49–1.16), P = .20), stroke (OR, 0.79 [95% CI: 0.47–1.34], P = .39), systemic/pulmonary embolism (OR, 1.34 [95% CI: 0.30–6.04], P = .70), and major bleeding (OR, 1.10 [95% CI: 0.83–1.48], P = .50) were comparable between the two devices. The odds of device related thrombus (OR, 0.72 [95% CI: 0.46–1.14], P = .17) was comparable between both the group of patients, however the incidence of peri device leak was significantly lower in AA group (OR, 0.41 [95% CI: 0.26–0.66], P < .001) compared with WD group of patients. CONCLUSION: The AA was not superior to the Watchman device in terms of safety and efficacy. However, the Amulet occluder was associated with a higher incidence of procedure-related complications, and lower peri device leak.
format Online
Article
Text
id pubmed-10313297
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-103132972023-07-01 Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis Jaiswal, Vikash Ang, Song Peng Shrestha, Abhigan Babu Wajid, Zarghoona Endurance, Evbayekha Osas Ayoobkhan, Fathima Shehnaz Khan, Shazia Garimellla, Vamsi Huang, Helen Ghanim, Mohammed Song, David Sharma, Prachi Alraies, M. Chadi Biswas, Monodeep Medicine (Baltimore) 3400 The Amplatzer Amulet (AA) and Watchman devices (WD) are the 2 most frequently used devices for percutaneous LAA closure globally. OBJECTIVE: To evaluate the safety and clinical outcomes associated with these 2 devices in patients undergoing percutaneous LAA closure. METHOD: We systematically searched all electronic databases from inception until February 21, 2023. The primary endpoint was procedure related complications. Secondary endpoints were device related thrombus, stroke, cardiovascular mortality, peri device leak, systemic embolism, and all-cause mortality. RESULTS: A total of 3 randomized clinical trials with 2150 patients were included in this meta-analysis. The mean age was 75 and 76 years in the Amplatzer group and in the Watchman group, respectively. The odds of procedure-related complications (OR, 1.80 [95% CI: 1.21–2.67], P < .001) were significantly higher among patients with AA compared to the WD. However, the odds of all-cause mortality (OR, 0.75 (95% CI: 0.49–1.16), P = .20), stroke (OR, 0.79 [95% CI: 0.47–1.34], P = .39), systemic/pulmonary embolism (OR, 1.34 [95% CI: 0.30–6.04], P = .70), and major bleeding (OR, 1.10 [95% CI: 0.83–1.48], P = .50) were comparable between the two devices. The odds of device related thrombus (OR, 0.72 [95% CI: 0.46–1.14], P = .17) was comparable between both the group of patients, however the incidence of peri device leak was significantly lower in AA group (OR, 0.41 [95% CI: 0.26–0.66], P < .001) compared with WD group of patients. CONCLUSION: The AA was not superior to the Watchman device in terms of safety and efficacy. However, the Amulet occluder was associated with a higher incidence of procedure-related complications, and lower peri device leak. Lippincott Williams & Wilkins 2023-06-30 /pmc/articles/PMC10313297/ /pubmed/37390240 http://dx.doi.org/10.1097/MD.0000000000034185 Text en Copyright © 2023 the Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial License 4.0 (CCBY-NC) (https://creativecommons.org/licenses/by-nc/4.0/) , where it is permissible to download, share, remix, transform, and buildup the work provided it is properly cited. The work cannot be used commercially without permission from the journal.
spellingShingle 3400
Jaiswal, Vikash
Ang, Song Peng
Shrestha, Abhigan Babu
Wajid, Zarghoona
Endurance, Evbayekha Osas
Ayoobkhan, Fathima Shehnaz
Khan, Shazia
Garimellla, Vamsi
Huang, Helen
Ghanim, Mohammed
Song, David
Sharma, Prachi
Alraies, M. Chadi
Biswas, Monodeep
Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title_full Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title_fullStr Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title_full_unstemmed Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title_short Amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: A systematic review and meta-analysis
title_sort amplatzer amulet versus watchman device for percutaneous left atrial appendage closure: a systematic review and meta-analysis
topic 3400
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10313297/
https://www.ncbi.nlm.nih.gov/pubmed/37390240
http://dx.doi.org/10.1097/MD.0000000000034185
work_keys_str_mv AT jaiswalvikash amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT angsongpeng amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT shresthaabhiganbabu amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT wajidzarghoona amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT enduranceevbayekhaosas amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT ayoobkhanfathimashehnaz amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT khanshazia amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT garimelllavamsi amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT huanghelen amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT ghanimmohammed amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT songdavid amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT sharmaprachi amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT alraiesmchadi amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis
AT biswasmonodeep amplatzeramuletversuswatchmandeviceforpercutaneousleftatrialappendageclosureasystematicreviewandmetaanalysis