Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement

AIMS: After implantation of a subcutaneous implantable cardioverter-defibrillator (S-ICD), a defibrillation test (DFT) is performed to ensure that the device can effectively detect and terminate the induced ventricular arrhythmia. Data on DFT efficacy at generator replacement are scarce with a limit...

Descripción completa

Detalles Bibliográficos
Autores principales: de Veld, Jolien A, Pepplinkhuizen, Shari, van der Stuijt, Willeke, Quast, Anne-Floor B E, Olde Nordkamp, Louise R A, Kooiman, Kirsten M, Wilde, Arthur A M, Smeding, Lonneke, Knops, Reinoud E
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10325005/
https://www.ncbi.nlm.nih.gov/pubmed/37379530
http://dx.doi.org/10.1093/europace/euad184
_version_ 1785069218521153536
author de Veld, Jolien A
Pepplinkhuizen, Shari
van der Stuijt, Willeke
Quast, Anne-Floor B E
Olde Nordkamp, Louise R A
Kooiman, Kirsten M
Wilde, Arthur A M
Smeding, Lonneke
Knops, Reinoud E
author_facet de Veld, Jolien A
Pepplinkhuizen, Shari
van der Stuijt, Willeke
Quast, Anne-Floor B E
Olde Nordkamp, Louise R A
Kooiman, Kirsten M
Wilde, Arthur A M
Smeding, Lonneke
Knops, Reinoud E
author_sort de Veld, Jolien A
collection PubMed
description AIMS: After implantation of a subcutaneous implantable cardioverter-defibrillator (S-ICD), a defibrillation test (DFT) is performed to ensure that the device can effectively detect and terminate the induced ventricular arrhythmia. Data on DFT efficacy at generator replacement are scarce with a limited number of patients and conflicting results. This study evaluates conversion efficacy during DFT at elective S-ICD generator replacement in a large cohort from our tertiary centre. METHODS AND RESULTS: Retrospective data of patients who underwent an S-ICD generator replacement for battery depletion with subsequent DFT between February 2015 and June 2022 were collected. Defibrillation test data were collected from both implant and replacement procedures. PRAETORIAN scores at implant were calculated. Defibrillation test was defined unsuccessful when two conversions at 65 J failed. A total of 121 patients were included. The defibrillation test was successful in 95% after the first and 98% after two consecutive tests. This was comparable with success rates at implant, despite a significant rise in shock impedance (73 ± 23 vs. 83 ± 24 Ω, P < 0.001). Both patients with an unsuccessful DFT at 65 J successfully converted with 80 J. CONCLUSION: This study shows a high DFT conversion rate at elective S-ICD generator replacement, which is comparable to conversion rates at implant, despite a rise in shock impedance. Evaluating device position before generator replacement may be recommended to optimize defibrillation success at generator replacement.
format Online
Article
Text
id pubmed-10325005
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-103250052023-07-07 Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement de Veld, Jolien A Pepplinkhuizen, Shari van der Stuijt, Willeke Quast, Anne-Floor B E Olde Nordkamp, Louise R A Kooiman, Kirsten M Wilde, Arthur A M Smeding, Lonneke Knops, Reinoud E Europace Clinical Research AIMS: After implantation of a subcutaneous implantable cardioverter-defibrillator (S-ICD), a defibrillation test (DFT) is performed to ensure that the device can effectively detect and terminate the induced ventricular arrhythmia. Data on DFT efficacy at generator replacement are scarce with a limited number of patients and conflicting results. This study evaluates conversion efficacy during DFT at elective S-ICD generator replacement in a large cohort from our tertiary centre. METHODS AND RESULTS: Retrospective data of patients who underwent an S-ICD generator replacement for battery depletion with subsequent DFT between February 2015 and June 2022 were collected. Defibrillation test data were collected from both implant and replacement procedures. PRAETORIAN scores at implant were calculated. Defibrillation test was defined unsuccessful when two conversions at 65 J failed. A total of 121 patients were included. The defibrillation test was successful in 95% after the first and 98% after two consecutive tests. This was comparable with success rates at implant, despite a significant rise in shock impedance (73 ± 23 vs. 83 ± 24 Ω, P < 0.001). Both patients with an unsuccessful DFT at 65 J successfully converted with 80 J. CONCLUSION: This study shows a high DFT conversion rate at elective S-ICD generator replacement, which is comparable to conversion rates at implant, despite a rise in shock impedance. Evaluating device position before generator replacement may be recommended to optimize defibrillation success at generator replacement. Oxford University Press 2023-06-28 /pmc/articles/PMC10325005/ /pubmed/37379530 http://dx.doi.org/10.1093/europace/euad184 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Clinical Research
de Veld, Jolien A
Pepplinkhuizen, Shari
van der Stuijt, Willeke
Quast, Anne-Floor B E
Olde Nordkamp, Louise R A
Kooiman, Kirsten M
Wilde, Arthur A M
Smeding, Lonneke
Knops, Reinoud E
Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title_full Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title_fullStr Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title_full_unstemmed Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title_short Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
title_sort successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
topic Clinical Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10325005/
https://www.ncbi.nlm.nih.gov/pubmed/37379530
http://dx.doi.org/10.1093/europace/euad184
work_keys_str_mv AT develdjoliena successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT pepplinkhuizenshari successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT vanderstuijtwilleke successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT quastannefloorbe successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT oldenordkamplouisera successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT kooimankirstenm successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT wildearthuram successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT smedinglonneke successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement
AT knopsreinoude successfuldefibrillationtestinginpatientsundergoingelectivesubcutaneousimplantablecardioverterdefibrillatorgeneratorreplacement