Cargando…

Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation

PURPOSE: This study investigated the functional outcomes and complication rates of cuff downsizing for the treatment of recurrent or persistent stress urinary incontinence (SUI) in men after the implantation of an artificial urinary sphincter (AUS). METHODS: Data from our institutional AUS database...

Descripción completa

Detalles Bibliográficos
Autores principales: Weis, Sylvia, Ludwig, Tim A., Bahassan, Omar, Gild, Philipp, Vetterlein, Malte W., Fisch, Margit, Dahlem, Roland, Maurer, Valentin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Continence Society 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10325935/
https://www.ncbi.nlm.nih.gov/pubmed/37401025
http://dx.doi.org/10.5213/inj.2346030.015
_version_ 1785069321771286528
author Weis, Sylvia
Ludwig, Tim A.
Bahassan, Omar
Gild, Philipp
Vetterlein, Malte W.
Fisch, Margit
Dahlem, Roland
Maurer, Valentin
author_facet Weis, Sylvia
Ludwig, Tim A.
Bahassan, Omar
Gild, Philipp
Vetterlein, Malte W.
Fisch, Margit
Dahlem, Roland
Maurer, Valentin
author_sort Weis, Sylvia
collection PubMed
description PURPOSE: This study investigated the functional outcomes and complication rates of cuff downsizing for the treatment of recurrent or persistent stress urinary incontinence (SUI) in men after the implantation of an artificial urinary sphincter (AUS). METHODS: Data from our institutional AUS database spanning the period from 2009 to 2020 were retrospectively analyzed. The number of pads per day was determined, a standardized quality of life (QoL) questionnaire and the International Consultation on Incontinence Questionnaire (ICIQ) were administered, and postoperative complications according to the Clavien-Dindo classification were analyzed. RESULTS: Out of 477 patients who received AUS implantation during the study period, 25 (5.2%) underwent cuff downsizing (median age, 77 years; interquartile range [IQR], 74–81 years; median follow-up, 4.4 years; IQR, 3–6.9 years). Before downsizing, SUI was very severe (ICIQ score 19–21) or severe (ICQ score 13–18) in 80% of patients, moderate (ICIQ score 6–12) in 12%, and slight (ICIQ score 1–5) in 8%. After downsizing, 52% showed an improvement of >5 out of 21 points. However, 28% still had very severe or severe SUI, 48% had moderate SUI, and 20% had slight SUI. One patient no longer had SUI. In 52% of patients, the use of pads per day was reduced by ≥50%. QoL improved by >2 out of 6 points in 56% of patients. Complications (infections/urethral erosions) requiring device explantation occurred in 36% of patients, with a median time to event of 14.5 months. CONCLUSIONS: Although cuff downsizing carries a risk of AUS explantation, it can be a valuable treatment option for selected patients with persistent or recurrent SUI after AUS implantation. Over half of patients experienced improvements in symptoms, satisfaction, ICIQ scores, and pad use. It is important to inform patients about the potential risks and benefits of AUS to manage their expectations and assess individual risks.
format Online
Article
Text
id pubmed-10325935
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Korean Continence Society
record_format MEDLINE/PubMed
spelling pubmed-103259352023-07-08 Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation Weis, Sylvia Ludwig, Tim A. Bahassan, Omar Gild, Philipp Vetterlein, Malte W. Fisch, Margit Dahlem, Roland Maurer, Valentin Int Neurourol J Original Article PURPOSE: This study investigated the functional outcomes and complication rates of cuff downsizing for the treatment of recurrent or persistent stress urinary incontinence (SUI) in men after the implantation of an artificial urinary sphincter (AUS). METHODS: Data from our institutional AUS database spanning the period from 2009 to 2020 were retrospectively analyzed. The number of pads per day was determined, a standardized quality of life (QoL) questionnaire and the International Consultation on Incontinence Questionnaire (ICIQ) were administered, and postoperative complications according to the Clavien-Dindo classification were analyzed. RESULTS: Out of 477 patients who received AUS implantation during the study period, 25 (5.2%) underwent cuff downsizing (median age, 77 years; interquartile range [IQR], 74–81 years; median follow-up, 4.4 years; IQR, 3–6.9 years). Before downsizing, SUI was very severe (ICIQ score 19–21) or severe (ICQ score 13–18) in 80% of patients, moderate (ICIQ score 6–12) in 12%, and slight (ICIQ score 1–5) in 8%. After downsizing, 52% showed an improvement of >5 out of 21 points. However, 28% still had very severe or severe SUI, 48% had moderate SUI, and 20% had slight SUI. One patient no longer had SUI. In 52% of patients, the use of pads per day was reduced by ≥50%. QoL improved by >2 out of 6 points in 56% of patients. Complications (infections/urethral erosions) requiring device explantation occurred in 36% of patients, with a median time to event of 14.5 months. CONCLUSIONS: Although cuff downsizing carries a risk of AUS explantation, it can be a valuable treatment option for selected patients with persistent or recurrent SUI after AUS implantation. Over half of patients experienced improvements in symptoms, satisfaction, ICIQ scores, and pad use. It is important to inform patients about the potential risks and benefits of AUS to manage their expectations and assess individual risks. Korean Continence Society 2023-06 2023-06-30 /pmc/articles/PMC10325935/ /pubmed/37401025 http://dx.doi.org/10.5213/inj.2346030.015 Text en Copyright © 2023 Korean Continence Society https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Weis, Sylvia
Ludwig, Tim A.
Bahassan, Omar
Gild, Philipp
Vetterlein, Malte W.
Fisch, Margit
Dahlem, Roland
Maurer, Valentin
Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title_full Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title_fullStr Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title_full_unstemmed Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title_short Outcomes and Complication Rates of Cuff Downsizing in the Treatment of Worsening or Persistent Incontinence After Artificial Urinary Sphincter Implantation
title_sort outcomes and complication rates of cuff downsizing in the treatment of worsening or persistent incontinence after artificial urinary sphincter implantation
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10325935/
https://www.ncbi.nlm.nih.gov/pubmed/37401025
http://dx.doi.org/10.5213/inj.2346030.015
work_keys_str_mv AT weissylvia outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT ludwigtima outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT bahassanomar outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT gildphilipp outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT vetterleinmaltew outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT fischmargit outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT dahlemroland outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation
AT maurervalentin outcomesandcomplicationratesofcuffdownsizinginthetreatmentofworseningorpersistentincontinenceafterartificialurinarysphincterimplantation