Electron-counting in MicroED

The combination of high sensitivity and rapid readout makes it possible for electron-counting detectors to record cryogenic electron microscopy data faster and more accurately without increasing the exposure. This is especially useful for MicroED of macromolecular crystals where the strength of the...

Descripción completa

Detalles Bibliográficos
Autores principales: Hattne, Johan, Clabbers, Max T. B., Martynowycz, Michael W., Gonen, Tamir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cold Spring Harbor Laboratory 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10327187/
https://www.ncbi.nlm.nih.gov/pubmed/37425889
http://dx.doi.org/10.1101/2023.06.29.547123
Descripción
Sumario:The combination of high sensitivity and rapid readout makes it possible for electron-counting detectors to record cryogenic electron microscopy data faster and more accurately without increasing the exposure. This is especially useful for MicroED of macromolecular crystals where the strength of the diffracted signal at high resolution is comparable to the surrounding background. The ability to decrease the exposure also alleviates concerns about radiation damage which limits the information that can be recovered from a diffraction measurement. However, the dynamic range of electron-counting detectors requires careful data collection to avoid errors from coincidence loss. Nevertheless, these detectors are increasingly deployed in cryo-EM facilities, and several have been successfully used for MicroED. Provided coincidence loss can be minimized, electron-counting detectors bring high potential rewards.