Cargando…
Implementation of the Calypso system: a commissioning experience
BACKGROUND: The aim of this study was to describe the clinical implementation of the Calypso system with its potential impact on the treatment delivery. MATERIALS AND METHODS: The influence of the electromagnetic array was investigated on the kilovoltage cone beam computed tomography (kV-CBCT) image...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Via Medica
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10348330/ https://www.ncbi.nlm.nih.gov/pubmed/37456696 http://dx.doi.org/10.5603/RPOR.a2023.0029 |
Sumario: | BACKGROUND: The aim of this study was to describe the clinical implementation of the Calypso system with its potential impact on the treatment delivery. MATERIALS AND METHODS: The influence of the electromagnetic array was investigated on the kilovoltage cone beam computed tomography (kV-CBCT) image quality using the CATPHAN 504 CBCT images. Then, the QFix kVue Calypso couch top and the array attenuation, and their dosimetric influence on the Volumetric modulated arc therapy (VMAT) treatments of prostate was evaluated. RESULTS: Regarding the image quality, a significant increase of noise (p < 0.01) was detected with the array in place, resulting in a significant decrease in signal noise ratio (SNR) (p < 0.01). No difference in absolute contrast was observed. Finally, there was a significant decrease in contrast noise ratio (CNR) (p < 0.01) even if the deviation was only of 2.5%. For the dosimetric evaluation, the maximum attenuation of the couch was 12.02% and 13.19% for X6 and X6 flattening filter free (FFF), respectively (configuration of rails out). Besides, the mean attenuation of the array was 1.15% and 1.67% for X6 and X6 FFF, respectively. For the VMAT treatment plans, the mean dose was reduced by 0.61% for X6 and by 0.31% for X6 FFF beams when using the electromagnetic array. CONCLUSIONS: The Calypso system does not affect significantly the kV-CBCT image quality and the VMAT plan dose distribution. |
---|