Cargando…
Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke
OBJECTIVE: The objective of this study is to verify the reliability and the concurrent and discriminant validity of the measurements of spasticity offered by the robotic device, quantifying the (1) test–retest reliability, (2) correlation with the clinical evaluation using the Modified Ashworth Scal...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10356585/ https://www.ncbi.nlm.nih.gov/pubmed/37483539 http://dx.doi.org/10.3389/fnbot.2023.1172770 |
_version_ | 1785075309664534528 |
---|---|
author | Urrutia, Rocío Miren Gutiérrez-Muto, Ane Sanz-Morère, Clara B. Gómez, Arantxa Politi, Angela M. Lunardini, Francesca Baccini, Marco Cecchi, Francesca León, Natacha Oliviero, Antonio Tornero, Jesús |
author_facet | Urrutia, Rocío Miren Gutiérrez-Muto, Ane Sanz-Morère, Clara B. Gómez, Arantxa Politi, Angela M. Lunardini, Francesca Baccini, Marco Cecchi, Francesca León, Natacha Oliviero, Antonio Tornero, Jesús |
author_sort | Urrutia, Rocío |
collection | PubMed |
description | OBJECTIVE: The objective of this study is to verify the reliability and the concurrent and discriminant validity of the measurements of spasticity offered by the robotic device, quantifying the (1) test–retest reliability, (2) correlation with the clinical evaluation using the Modified Ashworth Scale (MAS), (3) inter-rater reliability between the two physiotherapists, and (4) ability to discriminate between healthy and stroke patients. METHODS: A total of 20 stroke patients and 20 healthy volunteers participated in the study. Two physical therapists (PT1 and PT2) independently evaluated the hand spasticity of stroke subjects using the MAS. Spasticity was assessed, both in healthy and stroke patients, with the Amadeo device at three increasing velocities of passive movement for three consecutive repeated assessments, while raw data of force and position were collected through an external program. DATA ANALYSIS: The intraclass correlation coefficient (ICC) and the weighted kappa were computed to estimate the reliability of the Amadeo device measurements, the inter-rater reliability of MAS, and the correlation between the MAS and Amadeo device measurements. The discriminant ability of the Amadeo device was assessed by comparing the stroke and healthy subjects' spasticity measurements with the percentage of agreements with 0 in MAS for healthy subjects. RESULTS: The test–retest reliability of the Amadeo device was high with ICC at all three velocities (ICC = 0.908, 0.958, and 0.964, respectively) but lower if analyzed with weighted kappa correlation (0.584, 0.748, and 0.749, respectively) as mean values for each velocity. The correlation between Amadeo and the clinical scale for stroke patients with weighted kappa correlation was poor (0.280 ± 0.212 for PT1 and 0.290 ± 0.155 for PT2). The inter-rater reliability of the clinical MAS was high (ICC = 0.911). CONCLUSION: Both MAS and Amadeo spasticity scores showed good reliability. The Amadeo scores did not show a strong clinical correlation with the MAS in stroke patients. Hitherto, Amadeo evaluation shows trends that are consistent with the characteristics of spasticity, such as an increase in spasticity as the speed of muscle stretching increases. The ability of the device to discriminate between stroke patients and healthy controls is low. Future studies adopting an instrumental gold standard for spasticity may provide further insight into the validity of these measurements. |
format | Online Article Text |
id | pubmed-10356585 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-103565852023-07-21 Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke Urrutia, Rocío Miren Gutiérrez-Muto, Ane Sanz-Morère, Clara B. Gómez, Arantxa Politi, Angela M. Lunardini, Francesca Baccini, Marco Cecchi, Francesca León, Natacha Oliviero, Antonio Tornero, Jesús Front Neurorobot Neuroscience OBJECTIVE: The objective of this study is to verify the reliability and the concurrent and discriminant validity of the measurements of spasticity offered by the robotic device, quantifying the (1) test–retest reliability, (2) correlation with the clinical evaluation using the Modified Ashworth Scale (MAS), (3) inter-rater reliability between the two physiotherapists, and (4) ability to discriminate between healthy and stroke patients. METHODS: A total of 20 stroke patients and 20 healthy volunteers participated in the study. Two physical therapists (PT1 and PT2) independently evaluated the hand spasticity of stroke subjects using the MAS. Spasticity was assessed, both in healthy and stroke patients, with the Amadeo device at three increasing velocities of passive movement for three consecutive repeated assessments, while raw data of force and position were collected through an external program. DATA ANALYSIS: The intraclass correlation coefficient (ICC) and the weighted kappa were computed to estimate the reliability of the Amadeo device measurements, the inter-rater reliability of MAS, and the correlation between the MAS and Amadeo device measurements. The discriminant ability of the Amadeo device was assessed by comparing the stroke and healthy subjects' spasticity measurements with the percentage of agreements with 0 in MAS for healthy subjects. RESULTS: The test–retest reliability of the Amadeo device was high with ICC at all three velocities (ICC = 0.908, 0.958, and 0.964, respectively) but lower if analyzed with weighted kappa correlation (0.584, 0.748, and 0.749, respectively) as mean values for each velocity. The correlation between Amadeo and the clinical scale for stroke patients with weighted kappa correlation was poor (0.280 ± 0.212 for PT1 and 0.290 ± 0.155 for PT2). The inter-rater reliability of the clinical MAS was high (ICC = 0.911). CONCLUSION: Both MAS and Amadeo spasticity scores showed good reliability. The Amadeo scores did not show a strong clinical correlation with the MAS in stroke patients. Hitherto, Amadeo evaluation shows trends that are consistent with the characteristics of spasticity, such as an increase in spasticity as the speed of muscle stretching increases. The ability of the device to discriminate between stroke patients and healthy controls is low. Future studies adopting an instrumental gold standard for spasticity may provide further insight into the validity of these measurements. Frontiers Media S.A. 2023-07-05 /pmc/articles/PMC10356585/ /pubmed/37483539 http://dx.doi.org/10.3389/fnbot.2023.1172770 Text en Copyright © 2023 Urrutia, Miren Gutiérrez-Muto, Sanz-Morère, Gómez, Politi, Lunardini, Baccini, Cecchi, León, Oliviero and Tornero. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Neuroscience Urrutia, Rocío Miren Gutiérrez-Muto, Ane Sanz-Morère, Clara B. Gómez, Arantxa Politi, Angela M. Lunardini, Francesca Baccini, Marco Cecchi, Francesca León, Natacha Oliviero, Antonio Tornero, Jesús Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title | Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title_full | Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title_fullStr | Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title_full_unstemmed | Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title_short | Spasticity evaluation with the Amadeo Tyromotion device in patients with hemispheric stroke |
title_sort | spasticity evaluation with the amadeo tyromotion device in patients with hemispheric stroke |
topic | Neuroscience |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10356585/ https://www.ncbi.nlm.nih.gov/pubmed/37483539 http://dx.doi.org/10.3389/fnbot.2023.1172770 |
work_keys_str_mv | AT urrutiarocio spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT mirengutierrezmutoane spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT sanzmorereclarab spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT gomezarantxa spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT politiangelam spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT lunardinifrancesca spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT baccinimarco spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT cecchifrancesca spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT leonnatacha spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT olivieroantonio spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke AT tornerojesus spasticityevaluationwiththeamadeotyromotiondeviceinpatientswithhemisphericstroke |