Cargando…

Standardising the biochemical confirmation of adult male hypogonadism; a joint position statement by the Society for Endocrinology and Association of Clinical Biochemistry and Laboratory Medicine*

Background: Inter-assay variation between different immunoassays and different mass spectrometry methods hampers the biochemical confirmation of male hypogonadism. Furthermore, some laboratories utilis eassay manufacturer reference ranges that do not necessarily mirror assay performance characterist...

Descripción completa

Detalles Bibliográficos
Autores principales: Jayasena, Channa N, de Silva, Nipun Lakshitha, O’Reilly, Michael W, MacKenzie, Finlay, Marrington, Rachel, Jones, Hugh, Livingston, Mark, Downie, Paul, Hackett, Geoff, Ramachandran, Sud, Tomlinson, Jeremy, David, Janine, Boot, Christopher, Patel, Mayur, Tarling, Julie, Wu, Fredrick, Quinton, Richard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10357891/
https://www.ncbi.nlm.nih.gov/pubmed/37391940
http://dx.doi.org/10.1177/00045632231179022
Descripción
Sumario:Background: Inter-assay variation between different immunoassays and different mass spectrometry methods hampers the biochemical confirmation of male hypogonadism. Furthermore, some laboratories utilis eassay manufacturer reference ranges that do not necessarily mirror assay performance characteristics, with the lower limit of normality ranging from 4.9 nmol/L to 11 nmol/L. The quality of the normative data underlying commercial immunoassay reference ranges is uncertain. Design: A working group reviewed published evidence and agreed upon standardised reporting guidance to augment total testosterone reports. Results: Evidence-based guidance on appropriate blood sampling, clinical action limits, and other major factors likely to affect the interpretation of results are provided. Conclusions: This article aims to improve the quality of the interpretation of testosterone results by non-specialist clinicians. It also discusses approaches for assay harmonisation which have been successful in some but not all healthcare systems.