Flow rate accuracy of infusion devices within healthcare settings: a systematic review

BACKGROUND: One in five patients admitted to the hospital treated with intravenous (IV) fluid therapy suffer complications due to inappropriate administration. Errors have been reported in 13–84% of the preparation and administration of IV medications. The safe delivery of IV fluids requires precise...

Descripción completa

Detalles Bibliográficos
Autores principales: Atanda, Opeyemi, West, Jonathan, Stables, Tom, Johnson, Chris, Merrifield, Robert, Kinross, James
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10363896/
https://www.ncbi.nlm.nih.gov/pubmed/37492690
http://dx.doi.org/10.1177/20420986231188602
_version_ 1785076737085800448
author Atanda, Opeyemi
West, Jonathan
Stables, Tom
Johnson, Chris
Merrifield, Robert
Kinross, James
author_facet Atanda, Opeyemi
West, Jonathan
Stables, Tom
Johnson, Chris
Merrifield, Robert
Kinross, James
author_sort Atanda, Opeyemi
collection PubMed
description BACKGROUND: One in five patients admitted to the hospital treated with intravenous (IV) fluid therapy suffer complications due to inappropriate administration. Errors have been reported in 13–84% of the preparation and administration of IV medications. The safe delivery of IV fluids requires precise rate administration. OBJECTIVES: This systematic review aims to determine the accuracy of infusion sets and devices and examine the factors that affect the flow rate accuracy of devices. DATA SOURCES AND METHODS: Six databases (CINAHL, MEDLINE PubMed, EMBASE, Web of Science and Cochrane Database of systematic reviews) were systematically searched. Search terms included infusion pumps, infusion devices, flow rate accuracy, fluid administration rate, gravity-led infusion set and fluid balance. Studies were included if they examined infusion devices’ flow rate accuracy and drop rates for fluids or non-oncological drugs. Findings were tabulated and synthesised qualitatively. The quality of the studies was examined based on the design of the studies due to their heterogeneity. RESULTS: Eight studies were included: Four studies were conducted on human subjects in the hospital environment; studies recruited 182 participants between the ages of 18 and 94 years. Two studies examined flow rate accuracy in recruited patients across 509 observations and 2387 drip hours. No trials prospectively assessed the accuracy of infusion pumps in the clinical domain, and no studies were reported on patient safety outcomes. Four studies examined the impact of mechanical and physiological factors on the flow rate accuracies of infusion devices. Height and back pressure simulated vibrating conditions, the viscosity of IV fluid and the positions of patients were reported to have a significant impact on infusion volume and flow rates of infusion devices. Additionally, giving sets that vary from the manufacturer’s specifications are reported to increase error percent by 10–20%. CONCLUSION: Infusion devices are an important source of error in administering IV fluids. Yet, there needs to be more prospective trial data to support their clinical accuracy and the impact on patient outcomes. Future flow variability and accuracy studies should capture their impact on patient safety and clinical outcomes.
format Online
Article
Text
id pubmed-10363896
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-103638962023-07-25 Flow rate accuracy of infusion devices within healthcare settings: a systematic review Atanda, Opeyemi West, Jonathan Stables, Tom Johnson, Chris Merrifield, Robert Kinross, James Ther Adv Drug Saf Systematic Review BACKGROUND: One in five patients admitted to the hospital treated with intravenous (IV) fluid therapy suffer complications due to inappropriate administration. Errors have been reported in 13–84% of the preparation and administration of IV medications. The safe delivery of IV fluids requires precise rate administration. OBJECTIVES: This systematic review aims to determine the accuracy of infusion sets and devices and examine the factors that affect the flow rate accuracy of devices. DATA SOURCES AND METHODS: Six databases (CINAHL, MEDLINE PubMed, EMBASE, Web of Science and Cochrane Database of systematic reviews) were systematically searched. Search terms included infusion pumps, infusion devices, flow rate accuracy, fluid administration rate, gravity-led infusion set and fluid balance. Studies were included if they examined infusion devices’ flow rate accuracy and drop rates for fluids or non-oncological drugs. Findings were tabulated and synthesised qualitatively. The quality of the studies was examined based on the design of the studies due to their heterogeneity. RESULTS: Eight studies were included: Four studies were conducted on human subjects in the hospital environment; studies recruited 182 participants between the ages of 18 and 94 years. Two studies examined flow rate accuracy in recruited patients across 509 observations and 2387 drip hours. No trials prospectively assessed the accuracy of infusion pumps in the clinical domain, and no studies were reported on patient safety outcomes. Four studies examined the impact of mechanical and physiological factors on the flow rate accuracies of infusion devices. Height and back pressure simulated vibrating conditions, the viscosity of IV fluid and the positions of patients were reported to have a significant impact on infusion volume and flow rates of infusion devices. Additionally, giving sets that vary from the manufacturer’s specifications are reported to increase error percent by 10–20%. CONCLUSION: Infusion devices are an important source of error in administering IV fluids. Yet, there needs to be more prospective trial data to support their clinical accuracy and the impact on patient outcomes. Future flow variability and accuracy studies should capture their impact on patient safety and clinical outcomes. SAGE Publications 2023-07-21 /pmc/articles/PMC10363896/ /pubmed/37492690 http://dx.doi.org/10.1177/20420986231188602 Text en © The Author(s), 2023 https://creativecommons.org/licenses/by/4.0/This article is distributed under the terms of the Creative Commons Attribution 4.0 License (https://creativecommons.org/licenses/by/4.0/) which permits any use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Systematic Review
Atanda, Opeyemi
West, Jonathan
Stables, Tom
Johnson, Chris
Merrifield, Robert
Kinross, James
Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title_full Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title_fullStr Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title_full_unstemmed Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title_short Flow rate accuracy of infusion devices within healthcare settings: a systematic review
title_sort flow rate accuracy of infusion devices within healthcare settings: a systematic review
topic Systematic Review
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10363896/
https://www.ncbi.nlm.nih.gov/pubmed/37492690
http://dx.doi.org/10.1177/20420986231188602
work_keys_str_mv AT atandaopeyemi flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview
AT westjonathan flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview
AT stablestom flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview
AT johnsonchris flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview
AT merrifieldrobert flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview
AT kinrossjames flowrateaccuracyofinfusiondeviceswithinhealthcaresettingsasystematicreview