Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions

Intracardiac electrograms (iEGMs) are time traces of the electrical potential recorded close to the heart muscle. We calculate unipolar and bipolar iEGMs analytically for a myocardial slab with parallel myofibers and validate them against numerical bidomain simulations. The analytical solution obtai...

Descripción completa

Detalles Bibliográficos
Autores principales: Leenknegt, Lore, Panfilov, Alexander V., Dierckx, Hans
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10364610/
https://www.ncbi.nlm.nih.gov/pubmed/37492643
http://dx.doi.org/10.3389/fphys.2023.1213218
_version_ 1785076878902558720
author Leenknegt, Lore
Panfilov, Alexander V.
Dierckx, Hans
author_facet Leenknegt, Lore
Panfilov, Alexander V.
Dierckx, Hans
author_sort Leenknegt, Lore
collection PubMed
description Intracardiac electrograms (iEGMs) are time traces of the electrical potential recorded close to the heart muscle. We calculate unipolar and bipolar iEGMs analytically for a myocardial slab with parallel myofibers and validate them against numerical bidomain simulations. The analytical solution obtained via the method of mirrors is an infinite series of arctangents. It goes beyond the solid angle theory and is in good agreement with the simulations, even though bath loading effects were not accounted for in the analytical calculation. At a large distance from the myocardium, iEGMs decay as 1/R (unipolar), 1/R (2) (bipolar and parallel), and 1/R (3) (bipolar and perpendicular to the endocardium). At the endocardial surface, there is a mathematical branch cut. Here, we show how a thicker myocardium generates iEGMs with larger amplitudes and how anisotropy affects the iEGM width and amplitude. If only the leading-order term of our expansion is retained, it can be determined how the conductivities of the bath, torso, myocardium, and myofiber direction together determine the iEGM amplitude. Our results will be useful in the quantitative interpretation of iEGMs, the selection of thresholds to characterize viable tissues, and for future inferences of tissue parameters.
format Online
Article
Text
id pubmed-10364610
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-103646102023-07-25 Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions Leenknegt, Lore Panfilov, Alexander V. Dierckx, Hans Front Physiol Physiology Intracardiac electrograms (iEGMs) are time traces of the electrical potential recorded close to the heart muscle. We calculate unipolar and bipolar iEGMs analytically for a myocardial slab with parallel myofibers and validate them against numerical bidomain simulations. The analytical solution obtained via the method of mirrors is an infinite series of arctangents. It goes beyond the solid angle theory and is in good agreement with the simulations, even though bath loading effects were not accounted for in the analytical calculation. At a large distance from the myocardium, iEGMs decay as 1/R (unipolar), 1/R (2) (bipolar and parallel), and 1/R (3) (bipolar and perpendicular to the endocardium). At the endocardial surface, there is a mathematical branch cut. Here, we show how a thicker myocardium generates iEGMs with larger amplitudes and how anisotropy affects the iEGM width and amplitude. If only the leading-order term of our expansion is retained, it can be determined how the conductivities of the bath, torso, myocardium, and myofiber direction together determine the iEGM amplitude. Our results will be useful in the quantitative interpretation of iEGMs, the selection of thresholds to characterize viable tissues, and for future inferences of tissue parameters. Frontiers Media S.A. 2023-07-10 /pmc/articles/PMC10364610/ /pubmed/37492643 http://dx.doi.org/10.3389/fphys.2023.1213218 Text en Copyright © 2023 Leenknegt, Panfilov and Dierckx. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Physiology
Leenknegt, Lore
Panfilov, Alexander V.
Dierckx, Hans
Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title_full Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title_fullStr Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title_full_unstemmed Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title_short Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
title_sort impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions
topic Physiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10364610/
https://www.ncbi.nlm.nih.gov/pubmed/37492643
http://dx.doi.org/10.3389/fphys.2023.1213218
work_keys_str_mv AT leenknegtlore impactofelectrodeorientationmyocardialwallthicknessandmyofiberdirectiononintracardiacelectrogramsnumericalmodelingandanalyticalsolutions
AT panfilovalexanderv impactofelectrodeorientationmyocardialwallthicknessandmyofiberdirectiononintracardiacelectrogramsnumericalmodelingandanalyticalsolutions
AT dierckxhans impactofelectrodeorientationmyocardialwallthicknessandmyofiberdirectiononintracardiacelectrogramsnumericalmodelingandanalyticalsolutions