Cargando…
In-stent stenosis after stenting the PulseRider for a middle cerebral artery aneurysm: A case report
PulseRider is a self-expanding stent implant used to treat wide-necked aneurysms. As this device has a lower metal mass than conventional stents, it is expected to have a lower rate of ischemic complications and a shorter period with antiplatelet drugs. We experienced in-stent stenosis after endovas...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10369375/ https://www.ncbi.nlm.nih.gov/pubmed/37502481 http://dx.doi.org/10.1016/j.radcr.2023.07.010 |
Sumario: | PulseRider is a self-expanding stent implant used to treat wide-necked aneurysms. As this device has a lower metal mass than conventional stents, it is expected to have a lower rate of ischemic complications and a shorter period with antiplatelet drugs. We experienced in-stent stenosis after endovascular treatment with the PulseRider for a middle cerebral artery aneurysm. A 40-year-old woman with an unruptured aneurysm in the middle cerebral artery bifurcation underwent coil embolization using a PulseRider. The periprocedural course was not problematic, and postoperative angiography showed contrast filling in the aneurysm. The patient was discharged from our hospital on postembolization day 3. However, follow-up angiography after 6 months showed in-stent stenosis. Fortunately, no ischemic complications occurred after treatment. Although the PulseRider is characterized by a low metal mass, it should be noted that this device has some metal markers that can induce in-stent stenosis. |
---|