Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature
OBJECTIVE: We describe 3 cases with folding deformation of a PRECISE (Cordis, Miami, FL, USA) stent in carotid artery stenting (CAS). CASE PRESENTATIONS: The 3 cases with cervical carotid stenosis consisted of 3 males around 80 years old and included 2 symptomatic lesions. During CAS, distal embolic...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Japanese Society for Neuroendovascular Therapy
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10370877/ https://www.ncbi.nlm.nih.gov/pubmed/37501740 http://dx.doi.org/10.5797/jnet.cr.2022-0028 |
_version_ | 1785078032307847168 |
---|---|
author | Ichikawa, Tomonori Yasuda, Ryuta Yamamoto, Atsushi Nishikawa, Hirofumi Fujimoto, Masashi Miura, Yoichi Toma, Naoki Suzuki, Hidenori |
author_facet | Ichikawa, Tomonori Yasuda, Ryuta Yamamoto, Atsushi Nishikawa, Hirofumi Fujimoto, Masashi Miura, Yoichi Toma, Naoki Suzuki, Hidenori |
author_sort | Ichikawa, Tomonori |
collection | PubMed |
description | OBJECTIVE: We describe 3 cases with folding deformation of a PRECISE (Cordis, Miami, FL, USA) stent in carotid artery stenting (CAS). CASE PRESENTATIONS: The 3 cases with cervical carotid stenosis consisted of 3 males around 80 years old and included 2 symptomatic lesions. During CAS, distal embolic protection was established using a Mo.Ma (Medtronic, Minneapolis, MN, USA) along with a filter device in 2 cases and an Optimo (Tokai Medical Products, Aichi, Japan) along with a filter device in 1 case. For the filter device, either FilterWire EZ (Boston Scientific, Natick, MA, USA) or Spider FX (Covidien, Irvine, CA, USA) was employed. In all cases, a PRECISE stent was deployed after pre-dilation performed using a percutaneous transluminal angioplasty (PTA) balloon with the diameter of 2.5 to 3 mm. Post-dilation was performed after the stent deployment using a PTA balloon whose diameter was about 80% of that of the normal distal internal carotid artery. In all cases, cone-beam CT taken after the deployment of a stent showed folding deformation of the stent. In 2 cases, heavily calcified plaque hampered self-expansion of the stent, which resulted in the stent deformation. On the other hand, in the remaining 1 case, a distal shaft of the Mo.Ma caused the stent deformation, which was likely accelerated by head rotation and cervical compression that was performed to resolve difficulties for a filter retrieval device to pass through the stent, and post-dilation after the stenting. CONCLUSION: Heavily calcified plaque and a distal shaft of a Mo.Ma would result in stent deformation. |
format | Online Article Text |
id | pubmed-10370877 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | The Japanese Society for Neuroendovascular Therapy |
record_format | MEDLINE/PubMed |
spelling | pubmed-103708772023-07-27 Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature Ichikawa, Tomonori Yasuda, Ryuta Yamamoto, Atsushi Nishikawa, Hirofumi Fujimoto, Masashi Miura, Yoichi Toma, Naoki Suzuki, Hidenori J Neuroendovasc Ther Case Report OBJECTIVE: We describe 3 cases with folding deformation of a PRECISE (Cordis, Miami, FL, USA) stent in carotid artery stenting (CAS). CASE PRESENTATIONS: The 3 cases with cervical carotid stenosis consisted of 3 males around 80 years old and included 2 symptomatic lesions. During CAS, distal embolic protection was established using a Mo.Ma (Medtronic, Minneapolis, MN, USA) along with a filter device in 2 cases and an Optimo (Tokai Medical Products, Aichi, Japan) along with a filter device in 1 case. For the filter device, either FilterWire EZ (Boston Scientific, Natick, MA, USA) or Spider FX (Covidien, Irvine, CA, USA) was employed. In all cases, a PRECISE stent was deployed after pre-dilation performed using a percutaneous transluminal angioplasty (PTA) balloon with the diameter of 2.5 to 3 mm. Post-dilation was performed after the stent deployment using a PTA balloon whose diameter was about 80% of that of the normal distal internal carotid artery. In all cases, cone-beam CT taken after the deployment of a stent showed folding deformation of the stent. In 2 cases, heavily calcified plaque hampered self-expansion of the stent, which resulted in the stent deformation. On the other hand, in the remaining 1 case, a distal shaft of the Mo.Ma caused the stent deformation, which was likely accelerated by head rotation and cervical compression that was performed to resolve difficulties for a filter retrieval device to pass through the stent, and post-dilation after the stenting. CONCLUSION: Heavily calcified plaque and a distal shaft of a Mo.Ma would result in stent deformation. The Japanese Society for Neuroendovascular Therapy 2022-10-21 2022 /pmc/articles/PMC10370877/ /pubmed/37501740 http://dx.doi.org/10.5797/jnet.cr.2022-0028 Text en ©2022 The Japanese Society for Neuroendovascular Therapy https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives International License (https://creativecommons.org/licenses/by-nc-nd/4.0/) |
spellingShingle | Case Report Ichikawa, Tomonori Yasuda, Ryuta Yamamoto, Atsushi Nishikawa, Hirofumi Fujimoto, Masashi Miura, Yoichi Toma, Naoki Suzuki, Hidenori Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title | Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title_full | Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title_fullStr | Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title_full_unstemmed | Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title_short | Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature |
title_sort | folding deformation of open-cell stents in carotid artery stenting: report of three cases and review of literature |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10370877/ https://www.ncbi.nlm.nih.gov/pubmed/37501740 http://dx.doi.org/10.5797/jnet.cr.2022-0028 |
work_keys_str_mv | AT ichikawatomonori foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT yasudaryuta foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT yamamotoatsushi foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT nishikawahirofumi foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT fujimotomasashi foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT miurayoichi foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT tomanaoki foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature AT suzukihidenori foldingdeformationofopencellstentsincarotidarterystentingreportofthreecasesandreviewofliterature |