Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram
OBJECTIVE: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy. CASE PRESENTATION: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Japanese Society for Neuroendovascular Therapy
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10370995/ https://www.ncbi.nlm.nih.gov/pubmed/37502448 http://dx.doi.org/10.5797/jnet.cr.2021-0062 |
Sumario: | OBJECTIVE: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy. CASE PRESENTATION: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left middle cerebral artery (M1) was occluded, and emergency mechanical thrombectomy was performed. After partial recanalization was achieved, angiography through a JET 7 Xtra Flex was attempted. After manual injection of contrast media via a 10-mL syringe through the JET 7 Xtra Flex, the catheter moved, jumping forward, and the distal tip of the catheter expanded and ruptured. This resulted in intracranial vessel damage and subsequent patient death. CONCLUSION: Contrast media must not be injected through the JET 7 Xtra Flex. If contrast media needs to be injected for angiography during mechanical thrombectomy with a reperfusion catheter, it should always be through the guide catheter. |
---|