Cargando…
Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram
OBJECTIVE: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy. CASE PRESENTATION: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Japanese Society for Neuroendovascular Therapy
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10370995/ https://www.ncbi.nlm.nih.gov/pubmed/37502448 http://dx.doi.org/10.5797/jnet.cr.2021-0062 |
_version_ | 1785078058609278976 |
---|---|
author | Sano, Takanori Hakozaki, Koichi Kobayashi, Kazuto Ichikawa, Tomonori Tanemura, Hiroshi Ishigaki, Tomoki Miya, Fumitaka |
author_facet | Sano, Takanori Hakozaki, Koichi Kobayashi, Kazuto Ichikawa, Tomonori Tanemura, Hiroshi Ishigaki, Tomoki Miya, Fumitaka |
author_sort | Sano, Takanori |
collection | PubMed |
description | OBJECTIVE: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy. CASE PRESENTATION: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left middle cerebral artery (M1) was occluded, and emergency mechanical thrombectomy was performed. After partial recanalization was achieved, angiography through a JET 7 Xtra Flex was attempted. After manual injection of contrast media via a 10-mL syringe through the JET 7 Xtra Flex, the catheter moved, jumping forward, and the distal tip of the catheter expanded and ruptured. This resulted in intracranial vessel damage and subsequent patient death. CONCLUSION: Contrast media must not be injected through the JET 7 Xtra Flex. If contrast media needs to be injected for angiography during mechanical thrombectomy with a reperfusion catheter, it should always be through the guide catheter. |
format | Online Article Text |
id | pubmed-10370995 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | The Japanese Society for Neuroendovascular Therapy |
record_format | MEDLINE/PubMed |
spelling | pubmed-103709952023-07-27 Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram Sano, Takanori Hakozaki, Koichi Kobayashi, Kazuto Ichikawa, Tomonori Tanemura, Hiroshi Ishigaki, Tomoki Miya, Fumitaka J Neuroendovasc Ther Case Report OBJECTIVE: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy. CASE PRESENTATION: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left middle cerebral artery (M1) was occluded, and emergency mechanical thrombectomy was performed. After partial recanalization was achieved, angiography through a JET 7 Xtra Flex was attempted. After manual injection of contrast media via a 10-mL syringe through the JET 7 Xtra Flex, the catheter moved, jumping forward, and the distal tip of the catheter expanded and ruptured. This resulted in intracranial vessel damage and subsequent patient death. CONCLUSION: Contrast media must not be injected through the JET 7 Xtra Flex. If contrast media needs to be injected for angiography during mechanical thrombectomy with a reperfusion catheter, it should always be through the guide catheter. The Japanese Society for Neuroendovascular Therapy 2021-09-09 2022 /pmc/articles/PMC10370995/ /pubmed/37502448 http://dx.doi.org/10.5797/jnet.cr.2021-0062 Text en ©2022 The Japanese Society for Neuroendovascular Therapy https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives International License (https://creativecommons.org/licenses/by-nc-nd/4.0/) |
spellingShingle | Case Report Sano, Takanori Hakozaki, Koichi Kobayashi, Kazuto Ichikawa, Tomonori Tanemura, Hiroshi Ishigaki, Tomoki Miya, Fumitaka Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title | Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title_full | Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title_fullStr | Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title_full_unstemmed | Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title_short | Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram |
title_sort | vessel rupture complication due to ballooning of the distal segment of a jet 7 xtra flex reperfusion catheter during angiogram |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10370995/ https://www.ncbi.nlm.nih.gov/pubmed/37502448 http://dx.doi.org/10.5797/jnet.cr.2021-0062 |
work_keys_str_mv | AT sanotakanori vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT hakozakikoichi vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT kobayashikazuto vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT ichikawatomonori vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT tanemurahiroshi vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT ishigakitomoki vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram AT miyafumitaka vesselrupturecomplicationduetoballooningofthedistalsegmentofajet7xtraflexreperfusioncatheterduringangiogram |