Cargando…

Multichannel Resonant Acoustic Rheometry System for Rapid and Efficient Quantification of Human Plasma Coagulation

Resonant Acoustic Rheometry (RAR), a newly developed ultrasound-based technique for non-contact characterization of soft viscoelastic materials, has shown promise for quantitative assessment of plasma coagulation by monitoring the entire dynamic process in real time. Here, we report the development...

Descripción completa

Detalles Bibliográficos
Autores principales: Hendren, Christina, Li, Weiping, Stegemann, Jan P., Hall, Timothy L., Deng, Cheri. X.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Journal Experts 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10371129/
https://www.ncbi.nlm.nih.gov/pubmed/37503114
http://dx.doi.org/10.21203/rs.3.rs-3132931/v1
Descripción
Sumario:Resonant Acoustic Rheometry (RAR), a newly developed ultrasound-based technique for non-contact characterization of soft viscoelastic materials, has shown promise for quantitative assessment of plasma coagulation by monitoring the entire dynamic process in real time. Here, we report the development of a multichannel RAR (mRAR) system for simultaneous monitoring of the coagulation of multiple small-volume plasma samples, a capability that is critical to efficiently provide improved assessment of coagulation. The mRAR system was constructed using an array of 4 custom-designed ultrasound transducers at 5.0 MHz and an electronic driving system that controlled the generation of synchronized ultrasound pulses for real time monitoring of multiple samples simultaneously. The mRAR system was tested using Coumadin-treated plasma samples with a range of International Normalized Ratio (INR) values, as well as normal pooled plasma samples. Tracking of dynamic changes in clotting of plasma samples triggered by either kaolin or tissue factor was performed for the entire duration of coagulation. The mRAR system captured distinct changes in the samples and identified parameters including clotting time, clotting speed, and the mechanical properties of the clots that were consistent with Coumadin dose and INR levels Data from this study demonstrate the feasibility of the mRAR system for the rapid, efficient, and accurate characterization of plasma coagulation.