Cargando…

Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series

BACKGROUND: In 2010, a recall was issued for a specific monoblock large head metal-on-metal (MoM) hip prosthesis due to short-term revision rates of 12%-13% (articular surface replacement, DePuy Orthopaedics, Inc., Warsaw, IN). High complication, infection, and rerevision rates for revised MoM impla...

Descripción completa

Detalles Bibliográficos
Autores principales: Daher, Jimmy, Desai, Bhumit, Nammour, Michael, Warren, Michael, Chimento, George
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374868/
https://www.ncbi.nlm.nih.gov/pubmed/37521732
http://dx.doi.org/10.1016/j.artd.2023.101163
Descripción
Sumario:BACKGROUND: In 2010, a recall was issued for a specific monoblock large head metal-on-metal (MoM) hip prosthesis due to short-term revision rates of 12%-13% (articular surface replacement, DePuy Orthopaedics, Inc., Warsaw, IN). High complication, infection, and rerevision rates for revised MoM implants have been reported. The purpose of the study is to report long-term outcomes and trend metal ion levels of this recalled MoM prosthesis from a single surgeon series. METHODS: Retrospective chart review was performed on all patients that underwent revision of large MoM hip replacements between 2010 and 2015. Pre- and post-revision Harris Hip Score (HHS), cup abduction angles, anteversion angles, and cup sizes were compared. Survivorship and HHS were the primary outcomes measured; serum cobalt and chromium levels were secondary outcomes. Multivariate linear regression was used to examine the correlation between prerevision serum metal ion levels and HHS. RESULTS: A total of 24 hips (21 patients) met inclusion criteria. Mean time to revision was 4.12 years ± 1.1. Mean follow-up was 10.0 years (7-11.9 years). Mean HHS increased significantly after revision from 48.5 to 89.5 (P < .001). Higher prerevision cobalt levels were correlated with lower prerevision HHS (cobalt R = 0.25; chromium R = 0.3160). There was no correlation with prerevision cobalt (P = .2671) or chromium (P = .3160) with postrevision HHS. Most recent metal ion testing revealed a significant decrease in both cobalt (P = .0084) and chromium (P = .0115). Survival rate is 100%. CONCLUSIONS: Our study showed excellent survivorship and outcomes at 10 years. There were no failures for any reason including infection. This differs from previous studies and confirms excellent long-term results are possible with revision of this recalled MoM implant.