Cargando…
Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series
BACKGROUND: In 2010, a recall was issued for a specific monoblock large head metal-on-metal (MoM) hip prosthesis due to short-term revision rates of 12%-13% (articular surface replacement, DePuy Orthopaedics, Inc., Warsaw, IN). High complication, infection, and rerevision rates for revised MoM impla...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374868/ https://www.ncbi.nlm.nih.gov/pubmed/37521732 http://dx.doi.org/10.1016/j.artd.2023.101163 |
_version_ | 1785078871591223296 |
---|---|
author | Daher, Jimmy Desai, Bhumit Nammour, Michael Warren, Michael Chimento, George |
author_facet | Daher, Jimmy Desai, Bhumit Nammour, Michael Warren, Michael Chimento, George |
author_sort | Daher, Jimmy |
collection | PubMed |
description | BACKGROUND: In 2010, a recall was issued for a specific monoblock large head metal-on-metal (MoM) hip prosthesis due to short-term revision rates of 12%-13% (articular surface replacement, DePuy Orthopaedics, Inc., Warsaw, IN). High complication, infection, and rerevision rates for revised MoM implants have been reported. The purpose of the study is to report long-term outcomes and trend metal ion levels of this recalled MoM prosthesis from a single surgeon series. METHODS: Retrospective chart review was performed on all patients that underwent revision of large MoM hip replacements between 2010 and 2015. Pre- and post-revision Harris Hip Score (HHS), cup abduction angles, anteversion angles, and cup sizes were compared. Survivorship and HHS were the primary outcomes measured; serum cobalt and chromium levels were secondary outcomes. Multivariate linear regression was used to examine the correlation between prerevision serum metal ion levels and HHS. RESULTS: A total of 24 hips (21 patients) met inclusion criteria. Mean time to revision was 4.12 years ± 1.1. Mean follow-up was 10.0 years (7-11.9 years). Mean HHS increased significantly after revision from 48.5 to 89.5 (P < .001). Higher prerevision cobalt levels were correlated with lower prerevision HHS (cobalt R = 0.25; chromium R = 0.3160). There was no correlation with prerevision cobalt (P = .2671) or chromium (P = .3160) with postrevision HHS. Most recent metal ion testing revealed a significant decrease in both cobalt (P = .0084) and chromium (P = .0115). Survival rate is 100%. CONCLUSIONS: Our study showed excellent survivorship and outcomes at 10 years. There were no failures for any reason including infection. This differs from previous studies and confirms excellent long-term results are possible with revision of this recalled MoM implant. |
format | Online Article Text |
id | pubmed-10374868 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-103748682023-07-29 Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series Daher, Jimmy Desai, Bhumit Nammour, Michael Warren, Michael Chimento, George Arthroplast Today Original Research BACKGROUND: In 2010, a recall was issued for a specific monoblock large head metal-on-metal (MoM) hip prosthesis due to short-term revision rates of 12%-13% (articular surface replacement, DePuy Orthopaedics, Inc., Warsaw, IN). High complication, infection, and rerevision rates for revised MoM implants have been reported. The purpose of the study is to report long-term outcomes and trend metal ion levels of this recalled MoM prosthesis from a single surgeon series. METHODS: Retrospective chart review was performed on all patients that underwent revision of large MoM hip replacements between 2010 and 2015. Pre- and post-revision Harris Hip Score (HHS), cup abduction angles, anteversion angles, and cup sizes were compared. Survivorship and HHS were the primary outcomes measured; serum cobalt and chromium levels were secondary outcomes. Multivariate linear regression was used to examine the correlation between prerevision serum metal ion levels and HHS. RESULTS: A total of 24 hips (21 patients) met inclusion criteria. Mean time to revision was 4.12 years ± 1.1. Mean follow-up was 10.0 years (7-11.9 years). Mean HHS increased significantly after revision from 48.5 to 89.5 (P < .001). Higher prerevision cobalt levels were correlated with lower prerevision HHS (cobalt R = 0.25; chromium R = 0.3160). There was no correlation with prerevision cobalt (P = .2671) or chromium (P = .3160) with postrevision HHS. Most recent metal ion testing revealed a significant decrease in both cobalt (P = .0084) and chromium (P = .0115). Survival rate is 100%. CONCLUSIONS: Our study showed excellent survivorship and outcomes at 10 years. There were no failures for any reason including infection. This differs from previous studies and confirms excellent long-term results are possible with revision of this recalled MoM implant. Elsevier 2023-07-23 /pmc/articles/PMC10374868/ /pubmed/37521732 http://dx.doi.org/10.1016/j.artd.2023.101163 Text en © 2023 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Research Daher, Jimmy Desai, Bhumit Nammour, Michael Warren, Michael Chimento, George Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title | Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title_full | Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title_fullStr | Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title_full_unstemmed | Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title_short | Long-term Follow-up on Revisions of a Recalled Large Head Metal-on-metal Hip Prosthesis: A Single Surgeon Series |
title_sort | long-term follow-up on revisions of a recalled large head metal-on-metal hip prosthesis: a single surgeon series |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374868/ https://www.ncbi.nlm.nih.gov/pubmed/37521732 http://dx.doi.org/10.1016/j.artd.2023.101163 |
work_keys_str_mv | AT daherjimmy longtermfollowuponrevisionsofarecalledlargeheadmetalonmetalhipprosthesisasinglesurgeonseries AT desaibhumit longtermfollowuponrevisionsofarecalledlargeheadmetalonmetalhipprosthesisasinglesurgeonseries AT nammourmichael longtermfollowuponrevisionsofarecalledlargeheadmetalonmetalhipprosthesisasinglesurgeonseries AT warrenmichael longtermfollowuponrevisionsofarecalledlargeheadmetalonmetalhipprosthesisasinglesurgeonseries AT chimentogeorge longtermfollowuponrevisionsofarecalledlargeheadmetalonmetalhipprosthesisasinglesurgeonseries |