Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design

BACKGROUND: Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and o...

Descripción completa

Detalles Bibliográficos
Autores principales: Mödinger, Yvonne, Anttila, Eric D., Baker, Grant M., Gross, David C., Porporati, Alessandro A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374871/
https://www.ncbi.nlm.nih.gov/pubmed/37521740
http://dx.doi.org/10.1016/j.artd.2023.101170
_version_ 1785078872297963520
author Mödinger, Yvonne
Anttila, Eric D.
Baker, Grant M.
Gross, David C.
Porporati, Alessandro A.
author_facet Mödinger, Yvonne
Anttila, Eric D.
Baker, Grant M.
Gross, David C.
Porporati, Alessandro A.
author_sort Mödinger, Yvonne
collection PubMed
description BACKGROUND: Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images. METHODS: A novel knee arthroplasty device composed of BIOLOX delta, an alumina matrix composite (AMC) ceramic, was tested in a magnetic resonance (MR) environment. American Society for Testing and Materials test methods were used for evaluating magnetically induced displacement force, magnetically induced torque, radiofrequency-induced heating, and MR image artifact. RESULTS: Magnetically induced displacement force and magnetically induced torque results of the AMC ceramic knee indicated that these effects do not pose a known risk in a clinical MR environment, as assessed in a 3T magnetic field. Moreover, minimal radiofrequency-induced heating of the device was observed. In addition, the AMC ceramic knee demonstrated minimal MR image artifacts (7 mm) in comparison to a cobalt-chromium knee (88 mm). The extremely low magnetic susceptibility of AMC (2 ppm) underlines that it is a nonmetallic and nonmagnetic material well suited for the manufacturing of MR Safe orthopaedic implants. CONCLUSIONS: The AMC ceramic knee is a novel metal-free total knee arthroplasty device that can be regarded as MR Safe, as suggested by the absence of hazards from the exposure of this implant to a MR environment. The AMC ceramic knee presents the advantage of being scanned with superior imaging results in 3T MRI systems compared to alternative metal implants on the market.
format Online
Article
Text
id pubmed-10374871
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-103748712023-07-29 Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design Mödinger, Yvonne Anttila, Eric D. Baker, Grant M. Gross, David C. Porporati, Alessandro A. Arthroplast Today Original Research BACKGROUND: Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images. METHODS: A novel knee arthroplasty device composed of BIOLOX delta, an alumina matrix composite (AMC) ceramic, was tested in a magnetic resonance (MR) environment. American Society for Testing and Materials test methods were used for evaluating magnetically induced displacement force, magnetically induced torque, radiofrequency-induced heating, and MR image artifact. RESULTS: Magnetically induced displacement force and magnetically induced torque results of the AMC ceramic knee indicated that these effects do not pose a known risk in a clinical MR environment, as assessed in a 3T magnetic field. Moreover, minimal radiofrequency-induced heating of the device was observed. In addition, the AMC ceramic knee demonstrated minimal MR image artifacts (7 mm) in comparison to a cobalt-chromium knee (88 mm). The extremely low magnetic susceptibility of AMC (2 ppm) underlines that it is a nonmetallic and nonmagnetic material well suited for the manufacturing of MR Safe orthopaedic implants. CONCLUSIONS: The AMC ceramic knee is a novel metal-free total knee arthroplasty device that can be regarded as MR Safe, as suggested by the absence of hazards from the exposure of this implant to a MR environment. The AMC ceramic knee presents the advantage of being scanned with superior imaging results in 3T MRI systems compared to alternative metal implants on the market. Elsevier 2023-07-23 /pmc/articles/PMC10374871/ /pubmed/37521740 http://dx.doi.org/10.1016/j.artd.2023.101170 Text en © 2023 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research
Mödinger, Yvonne
Anttila, Eric D.
Baker, Grant M.
Gross, David C.
Porporati, Alessandro A.
Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title_full Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title_fullStr Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title_full_unstemmed Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title_short Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design
title_sort magnetic resonance safety evaluation of a novel alumina matrix composite ceramic knee and image artifact comparison to a metal knee implant of analogous design
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374871/
https://www.ncbi.nlm.nih.gov/pubmed/37521740
http://dx.doi.org/10.1016/j.artd.2023.101170
work_keys_str_mv AT modingeryvonne magneticresonancesafetyevaluationofanovelaluminamatrixcompositeceramickneeandimageartifactcomparisontoametalkneeimplantofanalogousdesign
AT anttilaericd magneticresonancesafetyevaluationofanovelaluminamatrixcompositeceramickneeandimageartifactcomparisontoametalkneeimplantofanalogousdesign
AT bakergrantm magneticresonancesafetyevaluationofanovelaluminamatrixcompositeceramickneeandimageartifactcomparisontoametalkneeimplantofanalogousdesign
AT grossdavidc magneticresonancesafetyevaluationofanovelaluminamatrixcompositeceramickneeandimageartifactcomparisontoametalkneeimplantofanalogousdesign
AT porporatialessandroa magneticresonancesafetyevaluationofanovelaluminamatrixcompositeceramickneeandimageartifactcomparisontoametalkneeimplantofanalogousdesign