Cargando…
Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system
PURPOSE: To implement an in-house developed position monitoring software, SeedTracker, for conventional fractionation prostate radiotherapy, and study the effect on dosimetric impact and intrafraction motion. METHODS: Thirty definitive prostate radiotherapy patients with implanted fiducial markers w...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10375704/ https://www.ncbi.nlm.nih.gov/pubmed/37519787 http://dx.doi.org/10.3389/fonc.2023.1082391 |
_version_ | 1785079092829224960 |
---|---|
author | Arumugam, Sankar Young, Tony Do, Viet Chlap, Phillip Tawfik, Christine Udovitch, Mark Wong, Karen Sidhom, Mark |
author_facet | Arumugam, Sankar Young, Tony Do, Viet Chlap, Phillip Tawfik, Christine Udovitch, Mark Wong, Karen Sidhom, Mark |
author_sort | Arumugam, Sankar |
collection | PubMed |
description | PURPOSE: To implement an in-house developed position monitoring software, SeedTracker, for conventional fractionation prostate radiotherapy, and study the effect on dosimetric impact and intrafraction motion. METHODS: Thirty definitive prostate radiotherapy patients with implanted fiducial markers were included in the study. All patients were treated with VMAT technique and plans were generated using the Pinnacle planning system using the 6MV beam model for Elekta linear accelerator. The target dose of 60 Gy in 20 fractions was prescribed for 29 of 30 patients, and one patient was treated with the target dose of 78 Gy in 39 fractions. The SeedTracker position monitoring system, which uses the x-ray images acquired during treatment delivery in the Elekta linear accelerator and associated XVI system, was used for online prostate position monitoring. The position tolerance for online verification was progressively reduced from 5 mm, 4 mm, and to 3 mm in 10 patient cohorts to effectively manage the treatment interruptions resulting from intrafraction motion in routine clinical practice. The delivered dose to target volumes and organs at risk in each of the treatment fractions was assessed by incorporating the observed target positions into the original treatment plan. RESULTS: In 27 of 30 patients, at least one gating event was observed, with a total of 177 occurrences of position deviation detected in 146 of 619 treatment fractions. In 5 mm, 4 mm, and 3 mm position tolerance cohorts, the position deviations were observed in 13%, 24%, and 33% of treatment fractions, respectively. Overall, the mean (range) deviation of -0.4 (-7.2 to 5.3) mm, -0.9 (-6.1 to 15.6) mm, and -1.7 (-7.0 to 6.1) mm was observed in Left-Right, Anterior-Posterior, and Superior-Inferior directions, respectively. The prostate CTV D99 would have been reduced by a maximum value of 1.3 Gy compared to the planned dose if position deviations were uncorrected, but with corrections, it was 0.3 Gy. Similarly, PTV D98 would have been reduced by a maximum value of 7.6 Gy uncorrected, with this difference reduced to 2.2 Gy with correction. The V60 to the rectum increased by a maximum of 1.0% uncorrected, which was reduced to 0.5%. CONCLUSION: Online target position monitoring for conventional fractionation prostate radiotherapy was successfully implemented on a standard Linear accelerator using an in-house developed position monitoring software, with an improvement in resultant dose to prostate target volume. |
format | Online Article Text |
id | pubmed-10375704 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-103757042023-07-29 Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system Arumugam, Sankar Young, Tony Do, Viet Chlap, Phillip Tawfik, Christine Udovitch, Mark Wong, Karen Sidhom, Mark Front Oncol Oncology PURPOSE: To implement an in-house developed position monitoring software, SeedTracker, for conventional fractionation prostate radiotherapy, and study the effect on dosimetric impact and intrafraction motion. METHODS: Thirty definitive prostate radiotherapy patients with implanted fiducial markers were included in the study. All patients were treated with VMAT technique and plans were generated using the Pinnacle planning system using the 6MV beam model for Elekta linear accelerator. The target dose of 60 Gy in 20 fractions was prescribed for 29 of 30 patients, and one patient was treated with the target dose of 78 Gy in 39 fractions. The SeedTracker position monitoring system, which uses the x-ray images acquired during treatment delivery in the Elekta linear accelerator and associated XVI system, was used for online prostate position monitoring. The position tolerance for online verification was progressively reduced from 5 mm, 4 mm, and to 3 mm in 10 patient cohorts to effectively manage the treatment interruptions resulting from intrafraction motion in routine clinical practice. The delivered dose to target volumes and organs at risk in each of the treatment fractions was assessed by incorporating the observed target positions into the original treatment plan. RESULTS: In 27 of 30 patients, at least one gating event was observed, with a total of 177 occurrences of position deviation detected in 146 of 619 treatment fractions. In 5 mm, 4 mm, and 3 mm position tolerance cohorts, the position deviations were observed in 13%, 24%, and 33% of treatment fractions, respectively. Overall, the mean (range) deviation of -0.4 (-7.2 to 5.3) mm, -0.9 (-6.1 to 15.6) mm, and -1.7 (-7.0 to 6.1) mm was observed in Left-Right, Anterior-Posterior, and Superior-Inferior directions, respectively. The prostate CTV D99 would have been reduced by a maximum value of 1.3 Gy compared to the planned dose if position deviations were uncorrected, but with corrections, it was 0.3 Gy. Similarly, PTV D98 would have been reduced by a maximum value of 7.6 Gy uncorrected, with this difference reduced to 2.2 Gy with correction. The V60 to the rectum increased by a maximum of 1.0% uncorrected, which was reduced to 0.5%. CONCLUSION: Online target position monitoring for conventional fractionation prostate radiotherapy was successfully implemented on a standard Linear accelerator using an in-house developed position monitoring software, with an improvement in resultant dose to prostate target volume. Frontiers Media S.A. 2023-07-14 /pmc/articles/PMC10375704/ /pubmed/37519787 http://dx.doi.org/10.3389/fonc.2023.1082391 Text en Copyright © 2023 Arumugam, Young, Do, Chlap, Tawfik, Udovitch, Wong and Sidhom https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Oncology Arumugam, Sankar Young, Tony Do, Viet Chlap, Phillip Tawfik, Christine Udovitch, Mark Wong, Karen Sidhom, Mark Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title | Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title_full | Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title_fullStr | Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title_full_unstemmed | Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title_short | Assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
title_sort | assessment of intrafraction motion and its dosimetric impact on prostate radiotherapy using an in-house developed position monitoring system |
topic | Oncology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10375704/ https://www.ncbi.nlm.nih.gov/pubmed/37519787 http://dx.doi.org/10.3389/fonc.2023.1082391 |
work_keys_str_mv | AT arumugamsankar assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT youngtony assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT doviet assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT chlapphillip assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT tawfikchristine assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT udovitchmark assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT wongkaren assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem AT sidhommark assessmentofintrafractionmotionanditsdosimetricimpactonprostateradiotherapyusinganinhousedevelopedpositionmonitoringsystem |