Cargando…
Kinematically Aligned Oxford Unicompartmental Knee Arthroplasty Using the Microplasty Instrumentation System
This technical note demonstrates kinematically aligned Oxford unicompartmental knee arthroplasty using the Microplasty instrumentation system with custom-made devices. The medial joint line is evaluated preoperatively; if it is aligned and parallel with the lateral joint line, they are considered to...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Orthopaedic Association
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10375823/ https://www.ncbi.nlm.nih.gov/pubmed/37529198 http://dx.doi.org/10.4055/cios22205 |
Sumario: | This technical note demonstrates kinematically aligned Oxford unicompartmental knee arthroplasty using the Microplasty instrumentation system with custom-made devices. The medial joint line is evaluated preoperatively; if it is aligned and parallel with the lateral joint line, they are considered to comprise the coronal knee joint line (CJL). In this case, the coronal inclination of a spoon gauge inserted into the medial joint space indicates the CJL. Otherwise, an accessory spoon is inserted and connected to the medial spoon to refer to the posterior condylar line, which is considered the CJL. The tibial cutting block is then connected without changing the inclination of the spoon and the coronal tilt of the tibial extramedullary rod is adjusted, which is implemented with a custom-made ankle yoke. The remainder of the steps is then identical to the conventional Microplasty procedure. This technique can imitate the cutting line to the CJL, which might be considered ideal from mechanical and kinematic perspectives. |
---|