Evaluation of Behavior of Castable versus Machined Solid Abutments for Morse Tapper Implant Connection: A Clinical Retrospective Study
Objective: The primary objective of the present retrospective clinical study was to evaluate and compare the clinical performance presented by castable abutments developed for the MT system versus intermediate machined abutments, specifically regarding prosthetic or implant fractures/loss; the secon...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10385154/ https://www.ncbi.nlm.nih.gov/pubmed/37512062 http://dx.doi.org/10.3390/medicina59071250 |
Sumario: | Objective: The primary objective of the present retrospective clinical study was to evaluate and compare the clinical performance presented by castable abutments developed for the MT system versus intermediate machined abutments, specifically regarding prosthetic or implant fractures/loss; the secondary objective was to verify the looseness of the abutments and the behavior of the peri-implant soft tissues. Methods: This clinical retrospective study was conducted on patients rehabilitated between 2019 and 2020. Inclusion criteria were patients in good general health, with an implants-supporting single crown; with solid machined abutments (control group) or castable UCLA abutments; with a connection portion (base) machined in cobalt-chrome (test group) over Morse taper DuoCone implants in the posterior mandible area; and at least two years in function. Clinical assessment was carried out by the same professional, considering the following parameters: (A) prosthetic: (i) loosening of the fixation screw, (ii) fracture of the screw and (iii) the number of times the patient had some type of complication after the installation of the prostheses were evaluated; (B) biological: (i) without keratinized mucosa (KM), (ii) 1 mm or less, (iii) between 1 and 2 mm and (iv) greater than 2 mm of KM width; and the presence or absence of mucositis. Furthermore, radiographic evaluation was performed in order to assess the marginal bone loss. These evaluations permitted to compare the groups analyzed and patients enrolled. Data were statistically analyzed, with the level of significance set at α = 0.05. Results: 79 patients with 120 MT implants were evaluated (80 castable UCLA abutments and 40 machined solid abutments). The follow-up was from 2 to 4 years. There was a 100% implant survival rate. Therefore, the control group showed two fractured abutments (5%) and no abutment loosening (95% for prosthetic survival rate), whereas the test group showed no abutment fracture but nine loosening screws (11.3%) (100% for prosthetic survival rate). Keratinized mucosa was considered thin or absent in 19 implants in the control group (47.5%) and 42 in the test group (52.5%). Mucositis was found in 11 implants in the control group (27.5%) and 27 in the test group (33.8%). A positive correlation was observed between the width of keratinized mucosa and mucositis (r = 0.521, p = 0.002). The mean marginal bone loss was 2.3 mm, ranging from 1.1 to 5.8 mm. No correlation was observed when considering marginal bone loss versus the three parameters (implant diameter, implant length and time of the prosthesis in function). Conclusions: The results suggest that UCLA-type abutments are a viable option for rehabilitating implants with Morse taper connections, suggesting lower fracture risk. Further research is necessary to confirm these findings and thoroughly evaluate the clinical performance and long-term outcomes. |
---|