The measurement of Central Corneal Thickness
We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, ref...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Romanian Society of Ophthalmology
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10385715/ https://www.ncbi.nlm.nih.gov/pubmed/37522018 http://dx.doi.org/10.22336/rjo.2023.29 |
Sumario: | We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. Abbreviations: OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer |
---|