Ramping down a clinical 3 T scanner: a journey into MRI and MRS at 0.75 T

OBJECT: Lower-field MR is reemerging as a viable, potentially cost-effective alternative to high-field MR, thanks to advances in hardware, sequence design, and reconstruction over the past decades. Evaluation of lower field strengths, however, is limited by the availability of lower-field systems on...

Descripción completa

Detalles Bibliográficos
Autores principales: Guenthner, Christian, Peereboom, Sophie Marie, Dillinger, Hannes, McGrath, Charles, Albannay, Mohammed Masoud, Vishnevskiy, Valery, Fuetterer, Max, Luechinger, Roger, Jenneskens, Theo, Sturzenegger, Urs, Overweg, Johan, Koken, Peter, Börnert, Peter, Kozerke, Sebastian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10386956/
https://www.ncbi.nlm.nih.gov/pubmed/37171689
http://dx.doi.org/10.1007/s10334-023-01089-9
Descripción
Sumario:OBJECT: Lower-field MR is reemerging as a viable, potentially cost-effective alternative to high-field MR, thanks to advances in hardware, sequence design, and reconstruction over the past decades. Evaluation of lower field strengths, however, is limited by the availability of lower-field systems on the market and their considerable procurement costs. In this work, we demonstrate a low-cost, temporary alternative to purchasing a dedicated lower-field MR system. MATERIALS AND METHODS: By ramping down an existing clinical 3 T MRI system to 0.75 T, proton signals can be acquired using repurposed (13)C transmit/receive hardware and the multi-nuclei spectrometer interface. We describe the ramp-down procedure and necessary software and hardware changes to the system. RESULTS: Apart from presenting system characterization results, we show in vivo examples of cardiac cine imaging, abdominal two- and three-point Dixon-type water/fat separation, water/fat-separated MR Fingerprinting, and point-resolved spectroscopy. In addition, the ramp-down approach allows unique comparisons of, e.g., gradient fidelity of the same MR system operated at different field strengths using the same receive chain, gradient coils, and amplifiers. DISCUSSION: Ramping down an existing MR system may be seen as a viable alternative for lower-field MR research in groups that already own multi-nuclei hardware and can also serve as a testing platform for custom-made multi-nuclei transmit/receive coils. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s10334-023-01089-9.