Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness

PURPOSE: To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT). METHODS: This was a prospective, cross-sectional, o...

Descripción completa

Detalles Bibliográficos
Autores principales: Porwal, Amit C, Shrishrimal, Meghna, Punamia, Ravin P, Mathew, Bennet Chacko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10391371/
https://www.ncbi.nlm.nih.gov/pubmed/37203058
http://dx.doi.org/10.4103/ijo.IJO_1982_22
_version_ 1785082690203025408
author Porwal, Amit C
Shrishrimal, Meghna
Punamia, Ravin P
Mathew, Bennet Chacko
author_facet Porwal, Amit C
Shrishrimal, Meghna
Punamia, Ravin P
Mathew, Bennet Chacko
author_sort Porwal, Amit C
collection PubMed
description PURPOSE: To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT). METHODS: This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted. Informed consent of the patients was taken. The IOP readings taken via the three methods were compared and correlated with CCT. Paired t test was used to compare the two devices. Simple and multivariate linear regression analyses were used to study the relationship between factors. A P value less than 0.05 was considered significant. Correlation was determined using the Pearson correlation coefficient, and a Bland–Altman graph was plotted. RESULTS: Mean IOP measured by the NCT was 15.65 ± 2.80 mmHg, by the RBT was 14.23 ± 3.05 mmHg, and by the GAT was 14.69 ± 2.97 mmHg. The mean CCT was 510.61 ± 33.83 microns. The difference between mean IOP recorded by the NCT and that by the RBT was 1.41 ± 2.39 mmHg, between the NCT and GAT was 0.95 ± 2.03 mmHg, and between the GAT and RBT was 0.45 ± 2.22 mmHg. The difference between the IOP values was statistically significant (P < 0.005). All tonometers showed a statistically significant correlation with CCT, but it was observed that the NCT had a stronger correlation (0.4037). CONCLUSION: The IOP readings taken by all the three methods were comparable; however, RBT values were closer to GAT values. CCT did influence the IOP values, and this should be kept in mind while evaluating.
format Online
Article
Text
id pubmed-10391371
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-103913712023-08-02 Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness Porwal, Amit C Shrishrimal, Meghna Punamia, Ravin P Mathew, Bennet Chacko Indian J Ophthalmol Original Article PURPOSE: To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT). METHODS: This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted. Informed consent of the patients was taken. The IOP readings taken via the three methods were compared and correlated with CCT. Paired t test was used to compare the two devices. Simple and multivariate linear regression analyses were used to study the relationship between factors. A P value less than 0.05 was considered significant. Correlation was determined using the Pearson correlation coefficient, and a Bland–Altman graph was plotted. RESULTS: Mean IOP measured by the NCT was 15.65 ± 2.80 mmHg, by the RBT was 14.23 ± 3.05 mmHg, and by the GAT was 14.69 ± 2.97 mmHg. The mean CCT was 510.61 ± 33.83 microns. The difference between mean IOP recorded by the NCT and that by the RBT was 1.41 ± 2.39 mmHg, between the NCT and GAT was 0.95 ± 2.03 mmHg, and between the GAT and RBT was 0.45 ± 2.22 mmHg. The difference between the IOP values was statistically significant (P < 0.005). All tonometers showed a statistically significant correlation with CCT, but it was observed that the NCT had a stronger correlation (0.4037). CONCLUSION: The IOP readings taken by all the three methods were comparable; however, RBT values were closer to GAT values. CCT did influence the IOP values, and this should be kept in mind while evaluating. Wolters Kluwer - Medknow 2023-05 2023-05-17 /pmc/articles/PMC10391371/ /pubmed/37203058 http://dx.doi.org/10.4103/ijo.IJO_1982_22 Text en Copyright: © 2023 Indian Journal of Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Porwal, Amit C
Shrishrimal, Meghna
Punamia, Ravin P
Mathew, Bennet Chacko
Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title_full Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title_fullStr Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title_full_unstemmed Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title_short Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
title_sort assessment of intraocular pressure measurement between goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10391371/
https://www.ncbi.nlm.nih.gov/pubmed/37203058
http://dx.doi.org/10.4103/ijo.IJO_1982_22
work_keys_str_mv AT porwalamitc assessmentofintraocularpressuremeasurementbetweengoldmanapplanationtonometerreboundtonometernoncontacttonometeranditscorrelationwithcentralcornealthickness
AT shrishrimalmeghna assessmentofintraocularpressuremeasurementbetweengoldmanapplanationtonometerreboundtonometernoncontacttonometeranditscorrelationwithcentralcornealthickness
AT punamiaravinp assessmentofintraocularpressuremeasurementbetweengoldmanapplanationtonometerreboundtonometernoncontacttonometeranditscorrelationwithcentralcornealthickness
AT mathewbennetchacko assessmentofintraocularpressuremeasurementbetweengoldmanapplanationtonometerreboundtonometernoncontacttonometeranditscorrelationwithcentralcornealthickness