P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis
OBJECTIVES: To evaluate the performance of a novel LAMP-based molecular test for the diagnosis of UTIs. The Llusern Scientific UTI test system consists of a panel of six uropathogen tests, positive and negative controls and an analyser, ‘Lodestar DX’ (Figure 1). The analyser will read the reaction d...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10395438/ http://dx.doi.org/10.1093/jacamr/dlad077.036 |
_version_ | 1785083576563269632 |
---|---|
author | van der Goes, Ayako Diggle, Joanna Nieuwland, Jeroen Roula, Ali Hayhurst, Emma |
author_facet | van der Goes, Ayako Diggle, Joanna Nieuwland, Jeroen Roula, Ali Hayhurst, Emma |
author_sort | van der Goes, Ayako |
collection | PubMed |
description | OBJECTIVES: To evaluate the performance of a novel LAMP-based molecular test for the diagnosis of UTIs. The Llusern Scientific UTI test system consists of a panel of six uropathogen tests, positive and negative controls and an analyser, ‘Lodestar DX’ (Figure 1). The analyser will read the reaction data and give the user a simple ‘positive’ or ‘negative’ read out for each uropathogen. The specific aim was to determine the sensitivity and specificity of the test system for the most common uropathogen, Escherichia coli. [Figure: see text] METHODS: In total, 149 urine samples received by Public Health Wales laboratories, Cardiff were randomly selected; 1 μL of each urine sample was added directly to each pre-made E. coli LAMP test. E. coli LAMP tests were placed into the Lodestar DX analyser and run. Results were shown and recorded after 40 minutes as ‘positive’ or ‘negative’ according to the decision of Lodestar's algorithm. LAMP results were compared to standard culture results and sensitivity and specificity calculated. RESULTS: Overall sensitivity of the E. coli LAMP test was 86.2%. Specificity was 88.3% (Table 1). Of the 14 samples positive in the LAMP test for E. coli but not recorded as E. coli using culture, 11 were from ‘mixed growth’ samples (Table 2). [Table: see text] [Table: see text] CONCLUSIONS: The Llusern Scientific UTI test system shows great promise as a point-of-care diagnostic tool. Sensitivity and specificity for E. coli was high and the test system may be better at diagnosing mixed infections which represent a high proportion of all culture results. The test system is easy to use with no sample processing and minimal training requirements. Test results are available in 40 min, compared to 24–72 h for culture results. The performance of the test system for the detection of other uropathogens (Klebsiella, Enterococcus, Pseudomonas, Proteus and S. saprophyticus) is ongoing. Sensitivity can be further improved by refining the test method. |
format | Online Article Text |
id | pubmed-10395438 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-103954382023-08-03 P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis van der Goes, Ayako Diggle, Joanna Nieuwland, Jeroen Roula, Ali Hayhurst, Emma JAC Antimicrob Resist Abstracts OBJECTIVES: To evaluate the performance of a novel LAMP-based molecular test for the diagnosis of UTIs. The Llusern Scientific UTI test system consists of a panel of six uropathogen tests, positive and negative controls and an analyser, ‘Lodestar DX’ (Figure 1). The analyser will read the reaction data and give the user a simple ‘positive’ or ‘negative’ read out for each uropathogen. The specific aim was to determine the sensitivity and specificity of the test system for the most common uropathogen, Escherichia coli. [Figure: see text] METHODS: In total, 149 urine samples received by Public Health Wales laboratories, Cardiff were randomly selected; 1 μL of each urine sample was added directly to each pre-made E. coli LAMP test. E. coli LAMP tests were placed into the Lodestar DX analyser and run. Results were shown and recorded after 40 minutes as ‘positive’ or ‘negative’ according to the decision of Lodestar's algorithm. LAMP results were compared to standard culture results and sensitivity and specificity calculated. RESULTS: Overall sensitivity of the E. coli LAMP test was 86.2%. Specificity was 88.3% (Table 1). Of the 14 samples positive in the LAMP test for E. coli but not recorded as E. coli using culture, 11 were from ‘mixed growth’ samples (Table 2). [Table: see text] [Table: see text] CONCLUSIONS: The Llusern Scientific UTI test system shows great promise as a point-of-care diagnostic tool. Sensitivity and specificity for E. coli was high and the test system may be better at diagnosing mixed infections which represent a high proportion of all culture results. The test system is easy to use with no sample processing and minimal training requirements. Test results are available in 40 min, compared to 24–72 h for culture results. The performance of the test system for the detection of other uropathogens (Klebsiella, Enterococcus, Pseudomonas, Proteus and S. saprophyticus) is ongoing. Sensitivity can be further improved by refining the test method. Oxford University Press 2023-08-02 /pmc/articles/PMC10395438/ http://dx.doi.org/10.1093/jacamr/dlad077.036 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of British Society for Antimicrobial Chemotherapy. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Abstracts van der Goes, Ayako Diggle, Joanna Nieuwland, Jeroen Roula, Ali Hayhurst, Emma P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title | P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title_full | P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title_fullStr | P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title_full_unstemmed | P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title_short | P32 Performance of a novel molecular test designed for point-of-care UTI diagnosis |
title_sort | p32 performance of a novel molecular test designed for point-of-care uti diagnosis |
topic | Abstracts |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10395438/ http://dx.doi.org/10.1093/jacamr/dlad077.036 |
work_keys_str_mv | AT vandergoesayako p32performanceofanovelmoleculartestdesignedforpointofcareutidiagnosis AT digglejoanna p32performanceofanovelmoleculartestdesignedforpointofcareutidiagnosis AT nieuwlandjeroen p32performanceofanovelmoleculartestdesignedforpointofcareutidiagnosis AT roulaali p32performanceofanovelmoleculartestdesignedforpointofcareutidiagnosis AT hayhurstemma p32performanceofanovelmoleculartestdesignedforpointofcareutidiagnosis |