Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer

BACKGROUND: Joint range of motion (ROM) is an important assessment to aid diagnostic and clinical decision-making for persons with a wide variety of neuromusculoskeletal conditions. The current clinical standard for assessing ROM is the standard goniometer (SG). PURPOSE: The purpose of this study wa...

Descripción completa

Detalles Bibliográficos
Autores principales: Hanks, June, Myers, Betsy
Formato: Online Artículo Texto
Lenguaje:English
Publicado: NASMI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10399115/
https://www.ncbi.nlm.nih.gov/pubmed/37547826
http://dx.doi.org/10.26603/001c.83944
_version_ 1785084201570140160
author Hanks, June
Myers, Betsy
author_facet Hanks, June
Myers, Betsy
author_sort Hanks, June
collection PubMed
description BACKGROUND: Joint range of motion (ROM) is an important assessment to aid diagnostic and clinical decision-making for persons with a wide variety of neuromusculoskeletal conditions. The current clinical standard for assessing ROM is the standard goniometer (SG). PURPOSE: The purpose of this study was to investigate the validity, reliability and time required to assess ROM using a standard goniometer (SG), medical inclinometer (MI), and builder’s digital inclinometer (BI). STUDY DESIGN: Cross-sectional study. METHODS: Fifty participants with no current shoulder, elbow, or forearm pain limiting movement were assessed by a single tester. The tester measured three repetitions of passive forearm and shoulder rotation with an SG, MI, and BI. Device order was randomized. Time to complete assessment with each device was measured. RESULTS: BI and MI were significantly faster than the SG (p < 0.001) for all motions. Inclinometer measurements were more reliable (average ICC = 0.933 for MI and 0.919 for BI) than SG measurements (average ICC = 0.822). There was good correlation between MI and BI and mean differences between devices was less than 2°. Correlations between the SG and the inclinometers ranged from poor to fair and mean differences between devices was 4°. CONCLUSION: The BI and MI were reliable for measuring forearm and shoulder rotation. The poor correlation between the SG and inclinometers indicates that clinicians should utilize the same device for testing. Because time can be a barrier to clinician assessment, the greater efficiency and reliability of inclinometers warrants consideration as the new measurement standard. Standard patient and inclinometer positioning is recommended to enhance reliability. LEVEL OF EVIDENCE: 2
format Online
Article
Text
id pubmed-10399115
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher NASMI
record_format MEDLINE/PubMed
spelling pubmed-103991152023-08-04 Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer Hanks, June Myers, Betsy Int J Sports Phys Ther Original Research BACKGROUND: Joint range of motion (ROM) is an important assessment to aid diagnostic and clinical decision-making for persons with a wide variety of neuromusculoskeletal conditions. The current clinical standard for assessing ROM is the standard goniometer (SG). PURPOSE: The purpose of this study was to investigate the validity, reliability and time required to assess ROM using a standard goniometer (SG), medical inclinometer (MI), and builder’s digital inclinometer (BI). STUDY DESIGN: Cross-sectional study. METHODS: Fifty participants with no current shoulder, elbow, or forearm pain limiting movement were assessed by a single tester. The tester measured three repetitions of passive forearm and shoulder rotation with an SG, MI, and BI. Device order was randomized. Time to complete assessment with each device was measured. RESULTS: BI and MI were significantly faster than the SG (p < 0.001) for all motions. Inclinometer measurements were more reliable (average ICC = 0.933 for MI and 0.919 for BI) than SG measurements (average ICC = 0.822). There was good correlation between MI and BI and mean differences between devices was less than 2°. Correlations between the SG and the inclinometers ranged from poor to fair and mean differences between devices was 4°. CONCLUSION: The BI and MI were reliable for measuring forearm and shoulder rotation. The poor correlation between the SG and inclinometers indicates that clinicians should utilize the same device for testing. Because time can be a barrier to clinician assessment, the greater efficiency and reliability of inclinometers warrants consideration as the new measurement standard. Standard patient and inclinometer positioning is recommended to enhance reliability. LEVEL OF EVIDENCE: 2 NASMI 2023-08-01 /pmc/articles/PMC10399115/ /pubmed/37547826 http://dx.doi.org/10.26603/001c.83944 Text en © The Author(s) https://creativecommons.org/licenses/by-nc/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (4.0) (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Original Research
Hanks, June
Myers, Betsy
Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title_full Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title_fullStr Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title_full_unstemmed Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title_short Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer
title_sort validity, reliability, and efficiency of a standard goniometer, medical inclinometer, and builder’s inclinometer
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10399115/
https://www.ncbi.nlm.nih.gov/pubmed/37547826
http://dx.doi.org/10.26603/001c.83944
work_keys_str_mv AT hanksjune validityreliabilityandefficiencyofastandardgoniometermedicalinclinometerandbuildersinclinometer
AT myersbetsy validityreliabilityandefficiencyofastandardgoniometermedicalinclinometerandbuildersinclinometer