Compatibility of a novel temperature‐controlled, irrigated radiofrequency catheter with ultra‐high‐density mapping

BACKGROUND: Compatibility of DiamondTemp (DT) radiofrequency (RF) catheter with the Rhythmia mapping system has not been manufacturer‐reported nor its tracking accuracy reported. METHODS: Consecutive patients undergoing macroreentrant atrial tachycardia ablation guided by Rhythmia and ablated using...

Descripción completa

Detalles Bibliográficos
Autores principales: Rodriguez Muñoz, Daniel, Marco del Castillo, Álvaro, Ramos Jimenez, Javier, Borrego Bernabe, Luis, Madrid Montoya, Alba, Lorenzo Balboa, Adrián, Arribas Ynsaurriaga, Fernando, Salguero‐Bodes, Rafael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10407174/
https://www.ncbi.nlm.nih.gov/pubmed/37560277
http://dx.doi.org/10.1002/joa3.12896
Descripción
Sumario:BACKGROUND: Compatibility of DiamondTemp (DT) radiofrequency (RF) catheter with the Rhythmia mapping system has not been manufacturer‐reported nor its tracking accuracy reported. METHODS: Consecutive patients undergoing macroreentrant atrial tachycardia ablation guided by Rhythmia and ablated using DT were prospectively enrolled. Following catheter configuration, ablation lines were performed and remapped to measure the RF tag to effective‐ablation‐line‐center (RFT‐ALC) distance. RESULTS: Among 20 consecutive patients (54 maps), 40 ablation lines were evaluated. Overall, the RFT‐ALC distance was 3.88 ± 2.95 mm, and the operator assessment of accuracy was high. No complications occurred. CONCLUSION: The use of DT catheter guided by the Rhythmia mapping system is feasible and accurate.