This study aimed to develop and validate a comprehensive, reproducible and automatic 4DCT Quality Assurance (QA) workflow (QAMotion) that evaluates image accuracy across various regular and irregular breathing patterns. Volume and amplitude deviations, CT number accuracy, and spatial integrity were...

Descripción completa

Detalles Bibliográficos
Autores principales: Bakkali Tahiri, Jinane, Kyndt, Martin, Dhont, Jennifer, Gulyban, Akos, Szkitsak, Juliane, Bogaert, Evelien, Reynaert, Nick, Burghelea, Manuela
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10407954/
https://www.ncbi.nlm.nih.gov/pubmed/37560513
http://dx.doi.org/10.1016/j.phro.2023.100475
Descripción
Sumario:This study aimed to develop and validate a comprehensive, reproducible and automatic 4DCT Quality Assurance (QA) workflow (QAMotion) that evaluates image accuracy across various regular and irregular breathing patterns. Volume and amplitude deviations, CT number accuracy, and spatial integrity were used as evaluation metrics. For repeatability tests, tolerances were respected with a mean CT number deviation < 10 HU, volume deviation < 2% and diameter and amplitude deviation < 2 mm except for irregular amplitude curves for which an amplitude deviation up to 6 mm was measured. QAMotion was able to flag image artefacts for our clinical 4DCT system.