Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement

Neonatal jaundice is an important cause of morbidity and mortality worldwide, and neonates born in low and middle-income countries bear a disproportionate burden. We previously developed a low-cost, point-of-care system to measure total serum bilirubin (TSB) in neonates. This device was effective at...

Descripción completa

Detalles Bibliográficos
Autores principales: Shapiro, Alyssa, Mtenthaonga, Prince, Mjumira, Rowland, Reuben, Margaret, Samuel, Ayodele, Bond, Meaghan, Carns, Jennifer, Schwarz, Richard, Johnston, Ryan, Mangwiro, Lucky, Odedere, Opeyemi, Miros, Robert, McHugh, Sean, Kawaza, Kondwani, Dube, Queen, Ezeaka, Chinyere, Richards-Kortum, Rebecca
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10409260/
https://www.ncbi.nlm.nih.gov/pubmed/37552665
http://dx.doi.org/10.1371/journal.pgph.0002262
_version_ 1785086240161267712
author Shapiro, Alyssa
Mtenthaonga, Prince
Mjumira, Rowland
Reuben, Margaret
Samuel, Ayodele
Bond, Meaghan
Carns, Jennifer
Schwarz, Richard
Johnston, Ryan
Mangwiro, Lucky
Odedere, Opeyemi
Miros, Robert
McHugh, Sean
Kawaza, Kondwani
Dube, Queen
Ezeaka, Chinyere
Richards-Kortum, Rebecca
author_facet Shapiro, Alyssa
Mtenthaonga, Prince
Mjumira, Rowland
Reuben, Margaret
Samuel, Ayodele
Bond, Meaghan
Carns, Jennifer
Schwarz, Richard
Johnston, Ryan
Mangwiro, Lucky
Odedere, Opeyemi
Miros, Robert
McHugh, Sean
Kawaza, Kondwani
Dube, Queen
Ezeaka, Chinyere
Richards-Kortum, Rebecca
author_sort Shapiro, Alyssa
collection PubMed
description Neonatal jaundice is an important cause of morbidity and mortality worldwide, and neonates born in low and middle-income countries bear a disproportionate burden. We previously developed a low-cost, point-of-care system to measure total serum bilirubin (TSB) in neonates. This device was effective at detecting and monitoring jaundice; however, the disposable strips were difficult to produce at scale. Here, we report a new lateral flow cassette design, called BiliDx, that was produced at scale using traditional manufacturing techniques. We evaluated the performance of BiliDx at sites in Nigeria and Malawi. The lateral flow strip consists of plasma separation membranes, nitrocellulose, and a plastic cassette. We evaluated the performance of the strips and reader at two hospitals located in Nigeria and Malawi compared to reference standard TSB. We also assessed performance for samples with high direct bilirubin (DB) and high hematocrit (HCT). We collected 1,144 samples from 758 neonates (TSB ranged from 0.2 to 45.9 mg/dL). The mean bias of BiliDx measurements in the validation set was +0.75 mg/dL, and 95% limits of agreement were -2.57 to 4.07 mg/dL. The mean bias and limits of agreement were comparable for samples with HCT < 60% and HCT ≥ 60%, and for samples with low and intermediate DB levels; the samples with high DB levels had wider 95% limits of agreement (-4.50 to +3.03 mg/dL). Error grid analysis shows that 96.9% of samples measured with BiliDx would have resulted in the same clinical decision as the reference standard. This performance is comparable to previous results that used a handmade two-dimensional strip. Additionally, error grid analysis shows that all 20 samples with high DB levels would have resulted in the same clinical decision as the reference standard. This evaluation supports the use of BiliDx lateral flow cassettes to provide accurate point-of-care measurements in low-resource settings.
format Online
Article
Text
id pubmed-10409260
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-104092602023-08-09 Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement Shapiro, Alyssa Mtenthaonga, Prince Mjumira, Rowland Reuben, Margaret Samuel, Ayodele Bond, Meaghan Carns, Jennifer Schwarz, Richard Johnston, Ryan Mangwiro, Lucky Odedere, Opeyemi Miros, Robert McHugh, Sean Kawaza, Kondwani Dube, Queen Ezeaka, Chinyere Richards-Kortum, Rebecca PLOS Glob Public Health Research Article Neonatal jaundice is an important cause of morbidity and mortality worldwide, and neonates born in low and middle-income countries bear a disproportionate burden. We previously developed a low-cost, point-of-care system to measure total serum bilirubin (TSB) in neonates. This device was effective at detecting and monitoring jaundice; however, the disposable strips were difficult to produce at scale. Here, we report a new lateral flow cassette design, called BiliDx, that was produced at scale using traditional manufacturing techniques. We evaluated the performance of BiliDx at sites in Nigeria and Malawi. The lateral flow strip consists of plasma separation membranes, nitrocellulose, and a plastic cassette. We evaluated the performance of the strips and reader at two hospitals located in Nigeria and Malawi compared to reference standard TSB. We also assessed performance for samples with high direct bilirubin (DB) and high hematocrit (HCT). We collected 1,144 samples from 758 neonates (TSB ranged from 0.2 to 45.9 mg/dL). The mean bias of BiliDx measurements in the validation set was +0.75 mg/dL, and 95% limits of agreement were -2.57 to 4.07 mg/dL. The mean bias and limits of agreement were comparable for samples with HCT < 60% and HCT ≥ 60%, and for samples with low and intermediate DB levels; the samples with high DB levels had wider 95% limits of agreement (-4.50 to +3.03 mg/dL). Error grid analysis shows that 96.9% of samples measured with BiliDx would have resulted in the same clinical decision as the reference standard. This performance is comparable to previous results that used a handmade two-dimensional strip. Additionally, error grid analysis shows that all 20 samples with high DB levels would have resulted in the same clinical decision as the reference standard. This evaluation supports the use of BiliDx lateral flow cassettes to provide accurate point-of-care measurements in low-resource settings. Public Library of Science 2023-08-08 /pmc/articles/PMC10409260/ /pubmed/37552665 http://dx.doi.org/10.1371/journal.pgph.0002262 Text en © 2023 Shapiro et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Shapiro, Alyssa
Mtenthaonga, Prince
Mjumira, Rowland
Reuben, Margaret
Samuel, Ayodele
Bond, Meaghan
Carns, Jennifer
Schwarz, Richard
Johnston, Ryan
Mangwiro, Lucky
Odedere, Opeyemi
Miros, Robert
McHugh, Sean
Kawaza, Kondwani
Dube, Queen
Ezeaka, Chinyere
Richards-Kortum, Rebecca
Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title_full Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title_fullStr Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title_full_unstemmed Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title_short Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
title_sort design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10409260/
https://www.ncbi.nlm.nih.gov/pubmed/37552665
http://dx.doi.org/10.1371/journal.pgph.0002262
work_keys_str_mv AT shapiroalyssa designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT mtenthaongaprince designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT mjumirarowland designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT reubenmargaret designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT samuelayodele designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT bondmeaghan designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT carnsjennifer designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT schwarzrichard designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT johnstonryan designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT mangwirolucky designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT odedereopeyemi designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT mirosrobert designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT mchughsean designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT kawazakondwani designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT dubequeen designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT ezeakachinyere designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement
AT richardskortumrebecca designandfieldevaluationofalateralflowcassettedeviceforpointofcarebilirubinmeasurement