Cargando…

Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities

The objective of this study is to present a credibility assessment of finite element modelling of self-expanding nickel-titanium (Ni-Ti) stents through verification and validation (VV) activities, as set out in the ASME VV-40 standard. As part of the study, the role of calculation verification, mode...

Descripción completa

Detalles Bibliográficos
Autores principales: Bernini, Martina, Hellmuth, Rudolf, Dunlop, Craig, Ronan, William, Vaughan, Ted J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10411813/
https://www.ncbi.nlm.nih.gov/pubmed/37556457
http://dx.doi.org/10.1371/journal.pone.0283492
_version_ 1785086753608040448
author Bernini, Martina
Hellmuth, Rudolf
Dunlop, Craig
Ronan, William
Vaughan, Ted J.
author_facet Bernini, Martina
Hellmuth, Rudolf
Dunlop, Craig
Ronan, William
Vaughan, Ted J.
author_sort Bernini, Martina
collection PubMed
description The objective of this study is to present a credibility assessment of finite element modelling of self-expanding nickel-titanium (Ni-Ti) stents through verification and validation (VV) activities, as set out in the ASME VV-40 standard. As part of the study, the role of calculation verification, model input sensitivity, and model validation is examined across three different application contexts (radial compression, stent deployment in a vessel, fatigue estimation). A commercially available self-expanding Ni-Ti stent was modelled, and calculation verification activities addressed the effects of mesh density, element integration and stable time increment on different quantities of interests, for each context of use considered. Sensitivity analysis of the geometrical and material input parameters and validation of deployment configuration with in vitro comparators were investigated. Results showed similar trends for global and local outputs across the contexts of use in response to the selection of discretization parameters, although with varying sensitivities. Mesh discretisation showed substantial variability for less than 4 × 4 element density across the strut cross-section in radial compression and deployment cases, while a finer grid was deemed necessary in fatigue estimation for reliable predictions of strain/stress. Element formulation also led to substantial variation depending on the chosen integration options. Furthermore, for explicit analyses, model results were highly sensitive to the chosen target time increment (e.g., mass scaling parameters), irrespective of whether quasistatic conditions were ensured (ratios of kinetic and internal energies below 5%). The higher variability was found for fatigue life simulation, with the estimation of fatigue safety factor varying up to an order of magnitude depending on the selection of discretization parameters. Model input sensitivity analysis highlighted that the predictions of outputs such as radial force and stresses showed relatively low sensitivity to Ni-Ti material parameters, which suggests that the calibration approaches used in the literature to date appear reasonable, but a higher sensitivity to stent geometry, namely strut thickness and width, was found. In contrast, the prediction of vessel diameter following deployment was least sensitive to numerical parameters, and its validation with in vitro comparators offered a simple and accurate (error ~ 1–2%) method when predicting diameter gain, and lumen area, provided that the material of the vessel is appropriately characterized and modelled.
format Online
Article
Text
id pubmed-10411813
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-104118132023-08-10 Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities Bernini, Martina Hellmuth, Rudolf Dunlop, Craig Ronan, William Vaughan, Ted J. PLoS One Research Article The objective of this study is to present a credibility assessment of finite element modelling of self-expanding nickel-titanium (Ni-Ti) stents through verification and validation (VV) activities, as set out in the ASME VV-40 standard. As part of the study, the role of calculation verification, model input sensitivity, and model validation is examined across three different application contexts (radial compression, stent deployment in a vessel, fatigue estimation). A commercially available self-expanding Ni-Ti stent was modelled, and calculation verification activities addressed the effects of mesh density, element integration and stable time increment on different quantities of interests, for each context of use considered. Sensitivity analysis of the geometrical and material input parameters and validation of deployment configuration with in vitro comparators were investigated. Results showed similar trends for global and local outputs across the contexts of use in response to the selection of discretization parameters, although with varying sensitivities. Mesh discretisation showed substantial variability for less than 4 × 4 element density across the strut cross-section in radial compression and deployment cases, while a finer grid was deemed necessary in fatigue estimation for reliable predictions of strain/stress. Element formulation also led to substantial variation depending on the chosen integration options. Furthermore, for explicit analyses, model results were highly sensitive to the chosen target time increment (e.g., mass scaling parameters), irrespective of whether quasistatic conditions were ensured (ratios of kinetic and internal energies below 5%). The higher variability was found for fatigue life simulation, with the estimation of fatigue safety factor varying up to an order of magnitude depending on the selection of discretization parameters. Model input sensitivity analysis highlighted that the predictions of outputs such as radial force and stresses showed relatively low sensitivity to Ni-Ti material parameters, which suggests that the calibration approaches used in the literature to date appear reasonable, but a higher sensitivity to stent geometry, namely strut thickness and width, was found. In contrast, the prediction of vessel diameter following deployment was least sensitive to numerical parameters, and its validation with in vitro comparators offered a simple and accurate (error ~ 1–2%) method when predicting diameter gain, and lumen area, provided that the material of the vessel is appropriately characterized and modelled. Public Library of Science 2023-08-09 /pmc/articles/PMC10411813/ /pubmed/37556457 http://dx.doi.org/10.1371/journal.pone.0283492 Text en © 2023 Bernini et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Bernini, Martina
Hellmuth, Rudolf
Dunlop, Craig
Ronan, William
Vaughan, Ted J.
Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title_full Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title_fullStr Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title_full_unstemmed Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title_short Recommendations for finite element modelling of nickel-titanium stents—Verification and validation activities
title_sort recommendations for finite element modelling of nickel-titanium stents—verification and validation activities
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10411813/
https://www.ncbi.nlm.nih.gov/pubmed/37556457
http://dx.doi.org/10.1371/journal.pone.0283492
work_keys_str_mv AT berninimartina recommendationsforfiniteelementmodellingofnickeltitaniumstentsverificationandvalidationactivities
AT hellmuthrudolf recommendationsforfiniteelementmodellingofnickeltitaniumstentsverificationandvalidationactivities
AT dunlopcraig recommendationsforfiniteelementmodellingofnickeltitaniumstentsverificationandvalidationactivities
AT ronanwilliam recommendationsforfiniteelementmodellingofnickeltitaniumstentsverificationandvalidationactivities
AT vaughantedj recommendationsforfiniteelementmodellingofnickeltitaniumstentsverificationandvalidationactivities