Simultaneous photon counting and charge integrating for pulse pile-up correction in paralyzable photon counting detectors
OBJECTIVE. In photon counting detectors (PCDs), electric pulses induced by two or more x-ray photons can pile up and result in count losses when their temporal separation is less than the detector dead time. The correction of pulse pile-up-induced count loss is particularly difficult for paralyzable...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10415089/ https://www.ncbi.nlm.nih.gov/pubmed/37379858 http://dx.doi.org/10.1088/1361-6560/ace2a9 |
Sumario: | OBJECTIVE. In photon counting detectors (PCDs), electric pulses induced by two or more x-ray photons can pile up and result in count losses when their temporal separation is less than the detector dead time. The correction of pulse pile-up-induced count loss is particularly difficult for paralyzable PCDs since a given value of recorded counts can correspond to two different values of true photon interactions. In contrast, charge (energy) integrating detectors work by integrating collected electric charge induced by x-rays over time and do not suffer from pile-up losses. This work introduces an inexpensive readout circuit element to the circuits of PCDs to simultaneously collect time-integrated charge to correct pile-up-induced count losses. APPROACH. Prototype electronics were constructed to collect time-integrated charges simultaneously with photon counts. A splitter was used to feed the electric signal in parallel to both a digital counter and a charge integrator. After recording PCD counts and integrating collected charge, a lookup table can be generated to map raw counts in the total- and high-energy bins and total charge to estimate pile-up-free true counts. Proof-of-concept imaging experiments were performed with a CdTe-based PCD array to test this method. MAIN RESULTS. The proposed electronics successfully recorded photon counts and time-integrated charge simultaneously, and whereas photon counts exhibited paralyzable pulse pile-up, time-integrated charge using the same electric signal as the counts measurement was linear with x-ray flux. With the proposed correction, paralyzable PCD counts became linear with input flux for both total- and high-energy bins. At high flux levels, uncorrected post-log measurements of PMMA objects severely overestimated radiological path lengths for both energy bins. After the proposed correction, the non-monotonic measurements again became linear with flux and accurately represented the true radiological path lengths. No impact on the spatial resolution was observed after the proposed correction in images of a line-pair test pattern. SIGNIFICANCE. Time-integrated charge can be used to correct for pulse pile-up in paralyzable PCDs where analytical solutions may be difficult to use, and integrated charge can be collected simultaneously with counts using inexpensive electronics. |
---|